A cryogenic radio-frequency ion trap for quantum logic spectroscopy of highly charged ions.
Leopold, T; King, S A; Micke, P; Bautista-Salvador, A; Heip, J C; Ospelkaus, C; Crespo López-Urrutia, J R; Schmidt, P O.
Afiliación
  • Leopold T; Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany.
  • King SA; Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany.
  • Micke P; Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany.
  • Bautista-Salvador A; Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany.
  • Heip JC; Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany.
  • Ospelkaus C; Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany.
  • Crespo López-Urrutia JR; Max-Planck-Institut für Kernphysik, Saupfercheckweg 1, 69117 Heidelberg, Germany.
  • Schmidt PO; Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany.
Rev Sci Instrum ; 90(7): 073201, 2019 Jul.
Article en En | MEDLINE | ID: mdl-31370455
A cryogenic radio-frequency ion trap system designed for quantum logic spectroscopy of highly charged ions (HCI) is presented. It includes a segmented linear Paul trap, an in-vacuum imaging lens, and a helical resonator. We demonstrate ground state cooling of all three modes of motion of a single 9Be+ ion and determine their heating rates as well as excess axial micromotion. The trap shows one of the lowest levels of electric field noise published to date. We investigate the magnetic-field noise suppression in cryogenic shields made from segmented copper, the resulting magnetic field stability at the ion position and the resulting coherence time. Using this trap in conjunction with an electron beam ion trap and a deceleration beamline, we have been able to trap single highly charged Ar13+ (Ar XIV) ions concurrently with single Be+ ions, a key prerequisite for the first quantum logic spectroscopy of a HCI. This major stepping stone allows us to push highly-charged-ion spectroscopic precision from the gigahertz to the hertz level and below.

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2019 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2019 Tipo del documento: Article País de afiliación: Alemania