Hollow-fibre liquid-phase microextraction and gas chromatography-mass spectrometric determination of amphetamines in whole blood.
Bombana, Henrique Silva; Dos Santos, Marcelo Filonzi; Muñoz, Daniel Romero; Leyton, Vilma.
Afiliación
  • Bombana HS; Faculty of Medicine, University of Sao Paulo, Av. Dr. Arnaldo, 455 Cerqueira Cesar, Sao Paulo 01246-903, Brazil. Electronic address: hbombana@usp.br.
  • Dos Santos MF; Faculty of Medicine, University of Sao Paulo, Av. Dr. Arnaldo, 455 Cerqueira Cesar, Sao Paulo 01246-903, Brazil.
  • Muñoz DR; Faculty of Medicine, University of Sao Paulo, Av. Dr. Arnaldo, 455 Cerqueira Cesar, Sao Paulo 01246-903, Brazil.
  • Leyton V; Faculty of Medicine, University of Sao Paulo, Av. Dr. Arnaldo, 455 Cerqueira Cesar, Sao Paulo 01246-903, Brazil.
Article en En | MEDLINE | ID: mdl-31962207
ABSTRACT
Here, we present a fully validated method using a hollow-fibre liquid-phase microextraction technique for the determination by gas chromatography-mass spectrometry (GC-MS) of amphetamine (AMP), methamphetamine (MET), fenproporex (FEN), 3,4-methylenedioxymethamphetamine (MDMA), 3,4-methylenedioxyamphetamine (MDA) and 3,4-methylenedioxyethylamphetamine (MDEA) in whole blood. The validation parameters presented successful values within those recommended by the Scientific Working Group for Forensic Toxicology (SWGTox) in the Standard Practices for Method Validation in Forensic Toxicology. The limits of detection ranged from 1 to 3 ng/mL, and the limits of quantification ranged from 2 to 5 ng/mL. The determination coefficients (r2) ranged from 0.990 to 0.997, and the method presented good intraday and interday accuracy (from 90.4% to 97.2%) and satisfactory recovery (from 68% to 110%). No carryover was observed. The heteroscedasticity was tested, and only AMP presented homoscedasticity. Weighting factors were applied to correct the linearity of MET (1/x2), MDA (1/x), FEN (1/x1/2), MDMA (1/x2) and MDEA (1/y). Dilution integrity was tested at ratios of 12, 15 and 110, and all maintained intraday precision (from 94.9% to 99.3%) and interday precision (from 89.4% to 94.9%). The validated method was applied to six real whole blood samples from individuals suspected of consuming ecstasy, and MDMA, MDA and amphetamine were successfully identified and quantified.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Microextracción en Fase Líquida / Anfetaminas / Cromatografía de Gases y Espectrometría de Masas Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2020 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Microextracción en Fase Líquida / Anfetaminas / Cromatografía de Gases y Espectrometría de Masas Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2020 Tipo del documento: Article