A Continuous Noninvasive Method to Assess Mixed Venous Oxygen Saturation: A Proof-of-Concept Study in Pigs.
Karlsson, Jacob; Lönnqvist, Per-Arne; Wallin, Mats; Hallbäck, Magnus.
Afiliación
  • Karlsson J; From the Department of Physiology and Pharmacology (Fysiologi och Farmakologi [FYFA]), Karolinska Institute, Stockholm, Sweden.
  • Lönnqvist PA; Department of Pediatric Perioperative Medicine and Intensive Care, Karolinska University Hospital, Stockholm, Sweden.
  • Wallin M; From the Department of Physiology and Pharmacology (Fysiologi och Farmakologi [FYFA]), Karolinska Institute, Stockholm, Sweden.
  • Hallbäck M; Department of Pediatric Perioperative Medicine and Intensive Care, Karolinska University Hospital, Stockholm, Sweden.
Anesth Analg ; 132(6): 1768-1776, 2021 06 01.
Article en En | MEDLINE | ID: mdl-33009138
ABSTRACT

BACKGROUND:

Mixed venous oxygen saturation (Svo2) is important when evaluating the balance between oxygen delivery and whole-body oxygen consumption. Monitoring Svo2 has so far required blood samples from a pulmonary artery catheter. By combining volumetric capnography, for measurement of effective pulmonary blood flow, with the Fick principle for oxygen consumption, we have developed a continuous noninvasive method, capnodynamic Svo2, for assessment of Svo2. The objective of this study was to validate this new technique against the gold standard cardiac output (CO)-oximetry Svo2 measurement of blood samples obtained from a pulmonary artery catheter and to assess the potential influence of intrapulmonary shunting.

METHODS:

Eight anesthetized mechanically ventilated domestic-breed piglets of both sexes (median weight 23.9 kg) were exposed to a series of interventions intended to reduce as well as increase Svo2. Simultaneous recordings of capnodynamic and CO-oximetry Svo2 as well as shunt fraction, using the Berggren formula, were performed throughout the protocol. Agreement of absolute values for capnodynamic and CO-oximetry Svo2 and the ability for capnodynamic Svo2 to detect change were assessed using Bland-Altman plot and concordance analysis.

RESULTS:

Overall bias for capnodynamic versus CO-oximetry Svo2 was -1 percentage point (limits of agreement -13 to +11 percentage points), a mean percentage error of 22%, and a concordance rate of 100%. Shunt fraction varied between 13% at baseline and 22% at the end of the study and was associated with only minor alterations in agreement between the tested methods.

CONCLUSIONS:

In the current experimental setting, capnodynamic assessment of Svo2 generates absolute values very close to the reference method CO-oximetry and is associated with 100% trending ability.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Consumo de Oxígeno / Oximetría / Gasto Cardíaco / Prueba de Estudio Conceptual Límite: Animals Idioma: En Revista: Anesth Analg Año: 2021 Tipo del documento: Article País de afiliación: Suecia

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Consumo de Oxígeno / Oximetría / Gasto Cardíaco / Prueba de Estudio Conceptual Límite: Animals Idioma: En Revista: Anesth Analg Año: 2021 Tipo del documento: Article País de afiliación: Suecia