Your browser doesn't support javascript.
loading
Early outcomes of Jotec inner-branched endografts in complex endovascular aortic aneurysm repair.
Abisi, Said; Zymvragoudakis, Vassilios; Gkoutzios, Panos; Sallam, Morad; Donati, Tommaso; Saha, Prakash; Zayed, Hany.
Afiliación
  • Abisi S; Department of Vascular Surgery, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom. Electronic address: said.abisi@gstt.nhs.uk.
  • Zymvragoudakis V; Department of Vascular Surgery, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
  • Gkoutzios P; Department of Radiology, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
  • Sallam M; Department of Vascular Surgery, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
  • Donati T; Department of Vascular Surgery, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
  • Saha P; Department of Vascular Surgery, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
  • Zayed H; Department of Vascular Surgery, Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
J Vasc Surg ; 74(3): 871-879, 2021 09.
Article en En | MEDLINE | ID: mdl-33647435
ABSTRACT

OBJECTIVE:

Complex aortic endografts have evolved to include inner branches to overcome specific challenges with existing technologies. We have reported the early outcomes of endovascular aortic aneurysm repair (EVAR) using a Jotec inner branched endograft (iBEVAR).

METHODS:

All patients who had undergone complex EVARs using extra-design engineering iBEVAR (Jotec GmbH, Hechingen, Germany) from 2018 to 2020 at a single center were reviewed. The patient demographics, cardiovascular risk factors, anatomic features of the aneurysms, and target vessels were recorded. The reasons for using inner branches instead of fenestrated and standard branched endografts and the procedural details, outcomes, and reintervention during follow-up were examined.

RESULTS:

A total of 110 patients were treated with branched and fenestrated endografts during the study period, of whom 18 patients had had a patient-specific custom-made iBEVAR endograft with downward inner branches. The technical success rate was 100%. A total of 68 target vessels were cannulated, and bridging stent-grafts were placed successfully in all. The reasons for choosing the iBEVAR design included unfavorable target vessel trajectory for fenestrated repair (n = 15), excessive infrarenal aortic angulation and/or adverse iliac access vessels for fenestrated repair (n = 11), the presence of a narrow aortic lumen (n = 14), and/or to reduce aortic coverage compared with that with standard outer branched repair (n = 14). We also used iBEVAR to treat type Ia endoleaks after failed EVAR with a short main body (n = 5). The median contrast volume used was 120 mL (range, 48-200 mL), with a median fluoroscopy screening time of 66 minutes (range, 35-136 minutes) and a median dose-area product of 17,832 dGy∙cm2 (range, 8260-55,070 dGycm2). No 30-day mortality and no major complications occurred. One early intervention was required for a suspected type Ib endoleak from an iliac limb and one late intervention for in-stent stenosis in a renal bridging stent-graft. One patient had died of non-aortic-related causes at 3 months. All other patients continued with follow-up with their aneurysms excluded, patent target vessels, and no type I or III endoleak identified at a median follow-up of 12 months (range, 1-26 months).

CONCLUSIONS:

The use of Jotec extra-design engineering endografts incorporating downward inner branches resulted in satisfactory early outcomes with a low reintervention rate. The technology has the potential to be a useful addition to our armamentarium for treating complex aortic endografts; however, long-term outcomes data are needed.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Prótesis Vascular / Aneurisma de la Aorta Torácica / Aneurisma de la Aorta Abdominal / Implantación de Prótesis Vascular / Procedimientos Endovasculares Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Vasc Surg Asunto de la revista: ANGIOLOGIA Año: 2021 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Prótesis Vascular / Aneurisma de la Aorta Torácica / Aneurisma de la Aorta Abdominal / Implantación de Prótesis Vascular / Procedimientos Endovasculares Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Vasc Surg Asunto de la revista: ANGIOLOGIA Año: 2021 Tipo del documento: Article