Your browser doesn't support javascript.
loading
Abutment mechanical complications of a Morse taper connection implant system: A 1- to 9-year retrospective study.
Yang, Fan; Ruan, Yan; Liu, Yunqing; Chen, Jianping; Chen, Yunfang; Zhang, Wentao; Ding, Yude; Wang, Linhong.
Afiliación
  • Yang F; Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital, Hangzhou Medical College), Hangzhou, Zhejiang, China.
  • Ruan Y; Department of Stomatology, Bengbu Medical College, Bengbu, Anhui, China.
  • Liu Y; Department of Restorative Dentistry, Division of Oral Health Science, Graduate School of Dental Medicine, Hokkaido University, Sapporo, Japan.
  • Chen J; Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital, Hangzhou Medical College), Hangzhou, Zhejiang, China.
  • Chen Y; Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital, Hangzhou Medical College), Hangzhou, Zhejiang, China.
  • Zhang W; Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital, Hangzhou Medical College), Hangzhou, Zhejiang, China.
  • Ding Y; Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital, Hangzhou Medical College), Hangzhou, Zhejiang, China.
  • Wang L; Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital, Hangzhou Medical College), Hangzhou, Zhejiang, China.
Clin Implant Dent Relat Res ; 24(5): 683-695, 2022 Oct.
Article en En | MEDLINE | ID: mdl-35791805
ABSTRACT

BACKGROUND:

The fracture of a Morse tapered abutment connection in an osseointegrated implant is one of the most serious mechanical complications, and it is extremely hard to deal with this complication in clinical practice.

PURPOSE:

The aim of this study was to explore the cumulative mechanical complications focus on abutment of a platform switching Morse taper connection implant system after loading, and to perform a retrospective, approximately 1- to 9-year follow-up study to identify the predisposing factors. MATERIALS AND

METHODS:

A total of 495 patients with 945 fitted implants were enrolled in this study with a follow-up from January 2012 to January 2020. The data of mechanical complications of the abutment, including abutment fracture (AF) and abutment screw loosening (ASL), and possible causative factors were extracted and evaluated statistically.

RESULTS:

A total of 25 out of 945 (2.65%) cumulative abutment mechanical complications occurred. AF was the most common complication (n = 13, 1.38%), followed by ASL (n = 12, 1.27%). For AF, gender, type of prosthesis, abutment design, and implant diameter were identified as the causative factors. AF was mostly observed in the single crown of males in molar areas, while ASL was more likely to occur on an angled abutment than on a non-angled abutment. Moreover, the abutment with the positioning index (/X) had a higher incidence of fracture than the abutment without the positioning index (C/).

CONCLUSIONS:

This study shows that the Morse taper connection is a safe abutment connection. AF occurs more frequently within single crowns in molar area of males, especially with the positioning index (/X), while ASL is more likely to occur in an angled abutment.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Implantes Dentales Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans / Male Idioma: En Revista: Clin Implant Dent Relat Res Asunto de la revista: ODONTOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Implantes Dentales Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans / Male Idioma: En Revista: Clin Implant Dent Relat Res Asunto de la revista: ODONTOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: China