Comprehensive Neonatal Cardiac, Feed and Wrap, Non-contrast, Non-sedated, Free-breathing Compressed Sensing 4D Flow MRI Assessment.
Panayiotou, Hannah R; Mills, Lily K; Broadbent, David A; Shelley, David; Scheffczik, Jutta; Olaru, Alexandra M; Jin, Ning; Greenwood, John P; Michael, Helen; Plein, Sven; Bissell, Malenka M.
Afiliación
  • Panayiotou HR; Biomedical Imaging Sciences Department, University of Leeds, Leeds, UK.
  • Mills LK; Biomedical Imaging Sciences Department, University of Leeds, Leeds, UK.
  • Broadbent DA; Biomedical Imaging Sciences Department, University of Leeds, Leeds, UK.
  • Shelley D; Department of Medical Physics and Engineering, Leeds Teaching Hospitals NHS Trust, Leeds, UK.
  • Scheffczik J; Biomedical Imaging Sciences Department, University of Leeds, Leeds, UK.
  • Olaru AM; Department of Anaesthesiology, Leeds Teaching Hospitals NHS Trust, Leeds, UK.
  • Jin N; Siemens Healthcare Ltd., Camberley, UK.
  • Greenwood JP; Siemens Medical Solutions USA, Inc., Chicago, Illinois, USA.
  • Michael H; Biomedical Imaging Sciences Department, University of Leeds, Leeds, UK.
  • Plein S; Department of Paediatric Cardiology, Leeds Teaching Hospitals NHS Trust, UK.
  • Bissell MM; Biomedical Imaging Sciences Department, University of Leeds, Leeds, UK.
J Magn Reson Imaging ; 57(3): 789-799, 2023 03.
Article en En | MEDLINE | ID: mdl-35792484
ABSTRACT

BACKGROUND:

Cardiac MRI is an important imaging tool in congenital cardiac disease, but its use has been limited in the neonatal population as general anesthesia has been needed for breath-holding. Technological advances in four-dimensional (4D) flow MRI have now made nonsedated free-breathing acquisition protocols a viable clinical option, but the method requires prospective validation in neonates.

PURPOSE:

To test the feasibility of compressed sensing (CS) 4D flow MRI in the neonatal population and to compare with standard previously validated two-dimensional (2D) phase-contrast (PC) flow MRI. STUDY TYPE Prospective, cohort, image quality. POPULATION A total of 14 healthy neonates (median [range] age 2.5 [0-80] days; 8 male). FIELD STRENGTH AND SEQUENCE Noncontrast 2D cine gradient echo sequence with through-plane velocity encoding (PC) sequence and compressed sensing (CS) three-dimensional (3D), time-resolved, cine phase-contrast MRI with 3D velocity-encoding (4D flow MRI) at 3 T. ASSESSMENT Aortic 2D PC, and aortic, pulmonary trunk and superior vena cava CS 4D flow MRI were acquired using the feed and wrap technique (nonsedated) and quantified using commercially available software. Aortic flow and peak velocity were compared between methods. Internal consistency of 4D flow MRI was determined by comparing mean forward flow of the main pulmonary artery (MPA) vs. the sum of left and right pulmonary artery flows (LPA and RPA) and by comparing mean ascending aorta forward flow (AAo) vs. the sum of superior vena cava (SVC) and descending aorta flows (DAo). STATISTICAL TESTS Flow and peak-velocity comparisons were assessed using paired t-tests, with P < 0.05 considered significant, and Bland-Altman analysis. Interobserver and intraobserver agreement and internal consistency were analyzed by intraclass correlation co-efficient (ICC).

RESULTS:

There was no statistically significant difference between ascending aortic forward flow between 2D PC and CS 4D Flow MRI (P = 0.26) with a bias of 0.11 mL (-0.59 to 0.82 mL) nor peak velocity (P = 0.11), with a bias of -5 cm/sec and (-26 to 16 cm/sec). There was excellent interobserver and intraobserver agreement for each vessel (interobserver ICC AAo 1.00; DAo 0.94, SVC 0.90, MPA 0.99, RPA 0.98, LPA 0.96; intraobserver ICC AAo 1.00; DAo 0.99, SVC 0.98, MPA 1.00, RPA 1.00, LPA 0.99). Internal consistency measures showed excellent agreement for both mean forward flow of main pulmonary artery vs. the sum of left and right pulmonary arteries (ICC 0.95) and mean ascending aorta forward flow vs. the sum of superior vena cava and descending aorta flows (ICC 1.00).

CONCLUSION:

Sedation-free neonatal feed and wrap MRI is well tolerated and feasible. CS 4D flow MRI quantification is similar to validated 2D PC free-breathing imaging with excellent interobserver and intraobserver agreement. EVIDENCE LEVEL 1 TECHNICAL EFFICACY Stage 2.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Vena Cava Superior / Imagen por Resonancia Magnética Tipo de estudio: Guideline Límite: Child, preschool / Humans / Male / Newborn Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Vena Cava Superior / Imagen por Resonancia Magnética Tipo de estudio: Guideline Límite: Child, preschool / Humans / Male / Newborn Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Reino Unido