Reduction of the cardiac pulsation artifact and improvement of lesion conspicuity in flow-compensated diffusion images in the liver-A quantitative evaluation of postprocessing algorithms.
Führes, Tobit; Saake, Marc; Lorenz, Jennifer; Seuss, Hannes; Stemmer, Alto; Benkert, Thomas; Uder, Michael; Laun, Frederik Bernd.
Afiliación
  • Führes T; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Saake M; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Lorenz J; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Seuss H; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Stemmer A; Abteilung für Radiologie, Klinikum Forchheim - Fränkische Schweiz, Forchheim, Germany.
  • Benkert T; MR Application Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany.
  • Uder M; MR Application Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany.
  • Laun FB; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
Magn Reson Med ; 89(1): 423-439, 2023 Jan.
Article en En | MEDLINE | ID: mdl-36089798
ABSTRACT

PURPOSE:

To enhance image quality of flow-compensated diffusion-weighted liver MRI data by increasing the lesion conspicuity and reducing the cardiac pulsation artifact using postprocessing algorithms.

METHODS:

Diffusion-weighted image data of 40 patients with liver lesions had been acquired at 1.5 T. These data were postprocessed with 5 different algorithms (weighted averaging, p-mean, percentile, outlier exclusion, and exception set). Four image properties of the postprocessed data were evaluated for optimizing the algorithm parameters. These properties were the lesion to tissue contrast-to-noise ratio (CNR), the reduction of the cardiac pulsation artifact, the data consistency, and the vessel darkness. They were combined into a total quality score ( Q total , $$ {Q}_{\mathrm{total}}, $$ set to 1 for the trace-weighted reference image), which was used to rate the image quality objectively.

RESULTS:

The weighted averaging algorithm performed best according to the total quality score ( Q total = 1.111 ± 0.067 $$ {Q}_{\mathrm{total}}=1.111\pm 0.067 $$ ). The further ranking was outlier exclusion algorithm ( Q total = 1.086 ± 0.061 $$ {Q}_{\mathrm{total}}=1.086\pm 0.061 $$ ), p-mean algorithm ( Q total = 1.045 ± 0.049 $$ {Q}_{\mathrm{total}}=1.045\pm 0.049 $$ ), percentile algorithm ( Q total = 1.012 ± 0.049 $$ {Q}_{\mathrm{total}}=1.012\pm 0.049 $$ ), and exception set algorithm ( Q total = 0.957 ± 0.027 $$ {Q}_{\mathrm{total}}=0.957\pm 0.027 $$ ). All optimized algorithms except for the exception set algorithm corrected the pulsation artifact and increased the lesion CNR. Changes in Q total $$ {Q}_{\mathrm{total}} $$ were significant for all optimized algorithms except for the percentile algorithm. Liver ADC was significantly reduced (except for the exception set algorithm), particularly in the left lobe.

CONCLUSION:

Postprocessing algorithms should be used for flow-compensated liver DWI. The proposed weighted averaging algorithm seems to be suited best to increase the image quality of artifact-corrupted flow-compensated diffusion-weighted liver data.