Your browser doesn't support javascript.
loading
Intra-Observer and Inter-Observer Variability of Intraocular Lens Measurements Using an Interferometry Metrology Device.
Stern, Benjamin; Saad, Alain; Flamant, Roxane; Joannes, Luc; Gatinel, Damien.
Afiliación
  • Stern B; Department of Ophthalmology, Rothschild Foundation Hospital, 75019 Paris, France.
  • Saad A; Department of Ophthalmology, Hadassah-Hebrew University Medical Center, Jerusalem 91120, Israel.
  • Flamant R; Department of Ophthalmology, Rothschild Foundation Hospital, 75019 Paris, France.
  • Joannes L; Department of Ophthalmology, Rothschild Foundation Hospital, 75019 Paris, France.
  • Gatinel D; Lambda-X SA, Avenue Robert Schuman, 102, B-1400 Nivelles, Belgium.
Diagnostics (Basel) ; 14(2)2024 Jan 19.
Article en En | MEDLINE | ID: mdl-38275463
ABSTRACT
The NIMO TEMPO (Lambda-X, Nivelles, Belgium) is a novel, user-friendly and compact device designed for in vitro optical analysis of refractive and diffractive intraocular lenses (IOLs). This device analyzes the IOL wavefront and generates a synthetic eye model for numerical computation. The objective of this study was to evaluate the precision of this innovative device. Intra- and inter-observer variability were calculated using a two-way analysis of variance (ANOVA) after conducting ten measurements of eight different IOL models, with each measurement being repeated by three distinct operators (resulting in a total of 30 measurements for each IOL). The device demonstrated satisfactory intra- and inter-observer variability in evaluating IOL power and modulation transfer function (MTF) profiles, with values of 0.066 and 0.078 diopters for IOL power and 0.018 and 0.019 for MTF measurements, respectively. Furthermore, this hybrid optical and numerical in vitro IOL wavefront analyzer appears to have several advantages over conventional optical bench devices. It reduces the need for operator manipulation, and allows for numerical modeling of various optical environments, including cornea models and apertures. In conclusion, this novel metrology device designed for refractive and diffractive IOLs appears to provide a satisfactory precision, making it a promising tool in the field of IOL metrology.
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Diagnostics (Basel) Año: 2024 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Diagnostics (Basel) Año: 2024 Tipo del documento: Article País de afiliación: Francia