Quartz-enhanced photoacoustic spectroscopy sensing using trapezoidal- and round-head quartz tuning forks.
Opt Lett
; 49(3): 770-773, 2024 Feb 01.
Article
en En
| MEDLINE
| ID: mdl-38300111
ABSTRACT
In this Letter, two novel, to the best of our knowledge, quartz tuning forks (QTFs) with trapezoidal-head and round-head were designed and adopted for quartz-enhanced photoacoustic spectroscopy (QEPAS) sensing. Based on finite element analysis, a theoretical simulation model was established to optimize the design of QTF. For performance comparison, a reported T-head QTF and a commercial QTF were also investigated. The designed QTFs have decreased resonant frequency (f0) and increased gap between the two prongs of QTF. The experimentally determined f0 of the T-head QTF, trapezoidal-head QTF, and round-head QTF were 8690.69â
Hz, 9471.67â
Hz, and 9499.28â
Hz, respectively. The corresponding quality (Q) factors were measured as 11,142, 11,411, and 11,874. Compared to the commercial QTF, the resonance frequencies of these QTFs have reduced by 73.45%, 71.07%, and 70.99% while maintaining a comparable Q factor to the commercially mature QTF. Methane (CH4) was chosen as the analyte to verify the QTFs' performance. Compared with the commercial QTF, the signal-to-noise ratio (SNR) of the CH4-QEPAS system based on the T-head QTF, trapezoidal-head QTF, and round-head QTF has been improved by 1.75 times, 2.96 times, and 3.26 times, respectively. The performance of the CH4-QEPAS sensor based on the QTF with the best performance of the round-head QTF was investigated in detail. The results indicated that the CH4-QEPAS sensor based on the round-head QTF exhibited an excellent linear concentration response. Furthermore, a minimum detection limit (MDL) of 0.87â
ppm can be achieved when the system's average time was 1200â
s.
Texto completo:
1
Banco de datos:
MEDLINE
Tipo de estudio:
Prognostic_studies
Idioma:
En
Revista:
Opt Lett
Año:
2024
Tipo del documento:
Article