Your browser doesn't support javascript.
loading
Safety and effectiveness of the three-dimensional-printed guide plate-assisted rotation axis positioning of a hinged external fixator for the elbow.
Ma, Pengcheng; Zheng, Jiachun; Chen, Huizhi; Yang, Weijie; Gao, Hongwei.
Afiliación
  • Ma P; Department of Trauma Surgery, Shandong Public Health Clinical Center, Jinan, Shandong, China.
  • Zheng J; Department of Trauma Surgery, Shandong Public Health Clinical Center, Jinan, Shandong, China.
  • Chen H; Department of Trauma Surgery, Shandong Public Health Clinical Center, Jinan, Shandong, China.
  • Yang W; Department of Trauma Surgery, Shandong Public Health Clinical Center, Jinan, Shandong, China.
  • Gao H; Department of Trauma Surgery, Shandong Public Health Clinical Center, Jinan, Shandong, China. gaohongwei202210@163.com.
Int Orthop ; 48(7): 1799-1808, 2024 Jul.
Article en En | MEDLINE | ID: mdl-38451310
ABSTRACT

PURPOSE:

We aimed to evaluate the safety and effectiveness of three-dimensional (3D)-printed guide plates for assisting in the positioning of the rotation axis of an elbow-hinged external fixator.

METHODS:

Terrible triad (TT) patients, who were screened using the predefined inclusion and exclusion criteria, underwent installation of a hinged external fixator on the basis of internal fixation; 3D-printed guide plates, generated from the patient's imaging data, assisted in positioning the rotation axis. All patients received the same peri-operative management and were followed up at six, 12, 24, and 48 weeks postoperatively. The duration of positioning pin placement, the number of fluoroscopies, pin placement success rate, types and incidence of post-operative complications, and the Mayo elbow performance score (MEPS) of the diseased elbow and range of motion (ROM) of both elbows were assessed.

RESULTS:

In 25 patients who completed the follow-up, the average time required for positioning pin placement was 329.32 ± 42.38 s (263-443 s), the average number of fluoroscopies was 2.32 ± 0.48 times (2-3 times), and the pin placement success rate was 100%. At the last follow-up, the mean MEPS of the diseased elbow was 97.50 ± 6.92 (75-100), with an excellent and good rate of 100%, and all patients demonstrated stable concentric reduction. The average range of flexion and extension was 135.08° ± 17.10° (77-146°), while the average range of rotation was 169.21° ± 18.14° (108-180°). No significant difference was observed in the average ROM between the both elbows (P > 0.05). Eight (32%) patients developed post-operative complications, including elbow stiffness due to heterotopic ossification in three (12%) patients, all of whom did not require secondary intervention.

CONCLUSION:

Utilizing 3D-printed guide plates for positioning the rotation axis of an elbow-hinged external fixator significantly reduced intra-operative positioning pin placement time and the number of fluoroscopies with excellent positioning results. Satisfactory results were also obtained in terms of post-operative complications, elbow ROM, and functional scores.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Fijadores Externos / Rango del Movimiento Articular / Articulación del Codo / Impresión Tridimensional Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Año: 2024 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Fijadores Externos / Rango del Movimiento Articular / Articulación del Codo / Impresión Tridimensional Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Año: 2024 Tipo del documento: Article País de afiliación: China