Targeted ultra performance liquid chromatography tandem mass spectrometry procedures for the diagnosis of inborn errors of metabolism: validation through ERNDIM external quality assessment schemes.
Clin Chem Lab Med
; 62(10): 1991-2000, 2024 Sep 25.
Article
en En
| MEDLINE
| ID: mdl-38456798
ABSTRACT
OBJECTIVES:
Early diagnosis of inborn errors of metabolism (IEM) is crucial to ensure early detection of conditions which are treatable. This study reports on targeted metabolomic procedures for the diagnosis of IEM of amino acids, acylcarnitines, creatine/guanidinoacetate, purines/pyrimidines and oligosaccharides, and describes its validation through external quality assessment schemes (EQA).METHODS:
Analysis was performed on a Waters ACQUITY UPLC H-class system coupled to a Waters Xevo triple-quadrupole (TQD) mass spectrometer, operating in both positive and negative electrospray ionization mode. Chromatographic separation was performed on a CORTECS C18 column (2.1 × 150, 1.6⯵m). Data were collected by multiple reaction monitoring.RESULTS:
The internal and EQA results were generally adequate, with a few exceptions. We calculated the relative measurement error (RME) and only a few metabolites displayed a RME higher than 30â¯% (asparagine and some acylcarnitine species). For oligosaccharides, semi-quantitative analysis of an educational panel clearly identified the 8 different diseases included.CONCLUSIONS:
Overall, we have validated our analytical system through an external quality control assessment. This validation will contribute to harmonization between laboratories, thus improving identification and management of patients with IEM.Palabras clave
Texto completo:
1
Banco de datos:
MEDLINE
Asunto principal:
Espectrometría de Masas en Tándem
/
Errores Innatos del Metabolismo
Límite:
Humans
Idioma:
En
Revista:
Clin Chem Lab Med
Asunto de la revista:
QUIMICA CLINICA
/
TECNICAS E PROCEDIMENTOS DE LABORATORIO
Año:
2024
Tipo del documento:
Article
País de afiliación:
España