Clinical Evaluation of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay.
Chan, Wai-Sing; Wong, Kan-Pui; Yau, Siu-Kei; Wong, Ching-Yan; Chan, Tsz-Ching; Hung, Jeffrey; Lai, Kristi Tsz-Wan; Leung, Chin-Pang; Wang, Candy Ling-Na; Au, Chun-Hang; Wan, Thomas Shek-Kong; Ma, Edmond Shiu-Kwan; Tang, Bone Siu-Fai.
Afiliación
  • Chan WS; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Wong KP; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Yau SK; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Wong CY; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Chan TC; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Hung J; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Lai KT; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Leung CP; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Wang CL; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Au CH; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Wan TS; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Ma ES; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
  • Tang BS; Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
Diagnostics (Basel) ; 14(7)2024 Mar 24.
Article en En | MEDLINE | ID: mdl-38611596
ABSTRACT
The performance of the Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assays were evaluated using 167 specimens, including 158 human respiratory specimens and 9 external quality assessment program (EQAP) samples. For respiratory specimens, CoV-2/Flu/RSV plus exhibited perfect agreement with the standard-of-care (SOC) methods (Cohen's κ 1, 100% agreement). The overall positive and negative percent agreement (PPA and NPA) were 100%, with 95% confidence intervals of 96.50 to 100% and 85.70 to 100%, respectively. On the other hand, Resp-4-Plex revealed an almost perfect agreement with the SOC methods (Cohen's κ 0.92, 97.71% agreement). The overall PPA and NPA were 100% (95.76 to 100%) and 88.46% (70.20 to 96.82%), respectively. For EQAP samples, the results of CoV-2/Flu/RSV plus (9/9) and Resp-4-Plex (4/4) were concordant with the expected results. The experimental limit of detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was the lowest (25 copies/mL for both methods), and that of the respiratory syncytial virus was the highest (400 copies/mL for CoV-2/Flu/RSV plus and 100 copies/mL for Resp-4-Plex). Threshold cycle (Ct) value correlation showed a large positive linear association between CoV-2/Flu/RSV plus and Resp-4-Plex, with R-squared values of 0.92-0.97, and on average, the Ct values of CoV-2/Flu/RSV plus were higher than that of Resp-4-Plex by 1.86-2.78, except for Flu A1 target (-0.66). To conclude, the performance of both assay was comparable to the SOC methods for both upper and lower respiratory specimens. Implementation of these rapid assay may reinforce the diagnostic capacity for the post-pandemic co-circulation of SARS-CoV-2 and other respiratory viruses.