ABSTRACT
BACKGROUND AND

AIMS:

This pilot study aimed to evaluate safety and tissue sampling from subepithelial lesions (SEL) in the upper gastrointestinal tract with a novel electric motor driven endoscopic ultrasonography (EUS)-guided 17-gauge (G) size core needle biopsy (CNB) instrument.

METHODS:

An investigator-led prospective open label, performance and safety control study, including seven patients (female n = 4, median 71 y, range 28-75) with a determined SEL (median size 30 mm, range 17-150 mm) in the upper digestive tract (stomach n = 6, duodenum n = 1) were eligible and later followed up 14 days after index procedure. All investigations were completed according to protocol with three FNB 22-G passes with four fanning strokes and two EndoDrill® 17-G passes with three fanning strokes.

RESULTS:

Quality of samples as 'visible pieces' (>5 mm) FNB (n = 5/7) (fragmented/blood imbibed n = 1, poor tissue quantity n = 1) compared with 17-G CNB (n = 7/7). Histological result which led to final diagnosis (leiomyoma n = 2, adenocarcinoma n = 1, schwannoma n = 1, neuroendocrine tumour n = 1, desmoid tumour n = 1 and gastrointestinal stromal tumour (GIST) n = 1) could be obtained with the 17-G CNB instrument in all seven patients. FNB technique reached correct diagnosis in six patients. No serious adverse event were recorded.

CONCLUSIONS:

By using an electric driven 17-G biopsy device, a true cylinder of core tissue can be obtained in one single puncture from the area of interest reducing the need for a second sampling. The absolute benefit of EUS-guided CNB is that the sample can be handled and histologically prepared in the same manner as standard percutaneous core needle sample, e.g., breast and prostate cancer.