Navigated intraoperative ultrasound in pediatric brain tumors.
Klein Gunnewiek, Kevin; van Baarsen, Kirsten M; Graus, Evie H M; Brink, Wyger M; Lequin, Maarten H; Hoving, Eelco W.
Afiliación
  • Klein Gunnewiek K; Department of Neuro-Oncology, Princess Máxima Center for Pediatric Oncology, Utrecht, The Netherlands. k.kleingunnewiek@prinsesmaximacentrum.nl.
  • van Baarsen KM; Magnetic Detection and Imaging Group, TechMed Centre, University of Twente, Enschede, The Netherlands. k.kleingunnewiek@prinsesmaximacentrum.nl.
  • Graus EHM; Department of Neuro-Oncology, Princess Máxima Center for Pediatric Oncology, Utrecht, The Netherlands.
  • Brink WM; Department of Neuro-Oncology, Princess Máxima Center for Pediatric Oncology, Utrecht, The Netherlands.
  • Lequin MH; Magnetic Detection and Imaging Group, TechMed Centre, University of Twente, Enschede, The Netherlands.
  • Hoving EW; Department of Neuro-Oncology, Princess Máxima Center for Pediatric Oncology, Utrecht, The Netherlands.
Childs Nerv Syst ; 40(9): 2697-2705, 2024 Sep.
Article en En | MEDLINE | ID: mdl-38862795
ABSTRACT

PURPOSE:

The aim of this study was to evaluate the diagnostic value and accuracy of navigated intraoperative ultrasound (iUS) in pediatric oncological neurosurgery as compared to intraoperative magnetic resonance imaging (iMRI).

METHODS:

A total of 24 pediatric patients undergoing tumor debulking surgery with iUS, iMRI, and neuronavigation were included in this study. Prospective acquisition of iUS images was done at two time points during the surgical procedure (1) before resection for tumor visualization and (2) after resection for residual tumor assessment. Dice similarity coefficients (DSC), Hausdorff distances 95th percentiles (HD95) and volume differences, sensitivity, and specificity were calculated for iUS segmentations as compared to iMRI.

RESULTS:

A high correlation (R = 0.99) was found for volume estimation as measured on iUS and iMRI before resection. A good spatial accuracy was demonstrated with a median DSC of 0.72 (IQR 0.14) and a median HD95 percentile of 4.98 mm (IQR 2.22 mm). The assessment after resection demonstrated a sensitivity of 100% and a specificity of 84.6% for residual tumor detection with navigated iUS. A moderate accuracy was observed with a median DSC of 0.58 (IQR 0.27) and a median HD95 of 5.84 mm (IQR 4.04 mm) for residual tumor volumes.

CONCLUSION:

We found that iUS measurements of tumor volume before resection correlate well with those obtained from preoperative MRI. The accuracy of residual tumor detection was reliable as compared to iMRI, indicating the suitability of iUS for directing the surgeon's attention to areas suspect for residual tumor. Therefore, iUS is considered as a valuable addition to the neurosurgical armamentarium. TRIAL REGISTRATION NUMBER AND DATE PMCLAB2023.476, February 12th 2024.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Neoplasias Encefálicas / Imagen por Resonancia Magnética / Neuronavegación Límite: Adolescent / Child / Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Childs Nerv Syst Asunto de la revista: NEUROLOGIA / PEDIATRIA Año: 2024 Tipo del documento: Article País de afiliación: Países Bajos

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Neoplasias Encefálicas / Imagen por Resonancia Magnética / Neuronavegación Límite: Adolescent / Child / Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Childs Nerv Syst Asunto de la revista: NEUROLOGIA / PEDIATRIA Año: 2024 Tipo del documento: Article País de afiliación: Países Bajos