A comparison of intraocular pressure measurement using SUOER SW-500 rebound tonometer and conventional reusable Goldmann prisms.
Chaung, Jia Quan; Sangapillai, Thanendthire; Quilat, Karen Kate; Perera, Shamira.
Afiliación
  • Chaung JQ; Singapore National Eye Centre, Singapore, Singapore.
  • Sangapillai T; Singapore Eye Research Institute, Singapore, Singapore.
  • Quilat KK; Singapore National Eye Centre, Singapore, Singapore.
  • Perera S; Singapore National Eye Centre, Singapore, Singapore.
Front Med (Lausanne) ; 11: 1269332, 2024.
Article en En | MEDLINE | ID: mdl-38938385
ABSTRACT

Introduction:

To determine the agreement between intraocular pressure (IOP) measurements using conventional Goldmann applanation tonometry (GA1,2T) and SUOER SW-500 Rebound Tonometer.

Methods:

This was a retrospective observational study where 205 eyes of 106 glaucoma patients had their IOPs measured by 2 fellowship trained ophthalmologists. Data were analyzed using the Bland-Altman method of differences. Correlation was measured using the Pearson coefficient.

Results:

Most of our patients were Chinese (88.7%) and female (51.9%). The average age was 66.9 years. The range of IOPs as measured by GAT was 2 to 58 mm Hg. Using the Bland-Altman method to compare GAT and SUOER SW-500 Rebound Tonometer. The tonometer overestimated the IOP by 0.5 mm Hg in the right eye and underestimated it by 0.1 mm Hg in the left eye. Overall, the tonometer overestimated the IOP by 0.2 mmHg. The Tonometer IOP correlated well with GAT, with a Pearson coefficient of correlation(r) of 0.89 (p < 0.001) for the right eye and 0.86 (p < 0.001) for the left eye, respectively. In patients with GAT IOP ≥ 21 mm Hg (n = 25), the Tonometer underestimated the IOP by 2.96 mm Hg.

Discussion:

The IOP measurements from the SUOER SW-500 Rebound Tonometer correlates well with the conventional GAT in measuring the IOP within normal ranges of IOP. SUOER SW-500 Rebound Tonometer may be of use, especially if the risk of transmission of infection is high considering that the probes are disposable. It is easy to use and its small size and portability makes it useful in situations where the patient is unable to be examined at the slit lamp.