Statistical optimal parameters obtained by using clinical human ocular aberrations for high-precision aberration measurement.
Yue, Xian; Yang, Yaliang; Chen, Shen; Dai, Hao.
Afiliación
  • Yue X; Institute of Optics and Electronics, Chinese Academy of Sciences, Chengdu, 610209, China.
  • Yang Y; Key Laboratory of Adaptive Optics, Chinese Academy of Sciences, Chengdu, 610209, China.
  • Chen S; University of Chinese Academy of Sciences, Beijing, 100049, China.
  • Dai H; Institute of Optics and Electronics, Chinese Academy of Sciences, Chengdu, 610209, China. ylyang@ioe.ac.cn.
Int Ophthalmol ; 44(1): 292, 2024 Jun 28.
Article en En | MEDLINE | ID: mdl-38940969
ABSTRACT

PURPOSE:

Compared to Shack-Hartmann wavefront sensor (SHWS), the parameters of virtual SHWS (vSHWS) can be easily adjusted to obtain the optimal performance of aberration measurement. Its current optimal parameters are obtained with only a set of statistical aberrations and not statistically significant. Whether the above parameters are consistent with the statistical results of the optimal parameters corresponding to each set of aberrations, and which performance is better if not? The purpose of this study was to answer these questions.

METHODS:

The optimal parameters to reconstruct 624 sets of clinical ocular aberrations in the highest accuracy, including the numbers of sub-apertures (NSAs) and the expansion ratios (ERs) of electric field zero-padding, were determined sequentially in this work. By using wavefront-reconstruction accuracy as an evaluation index, the statistical optimal parameter configuration was selected from some possible configurations determined by the optimal NSAs and ERs.

RESULTS:

The statistical optimal parameters are consistent for normal and abnormal eyes. They are different from the optimal parameters obtained with a set of statistical aberrations from the same 624 sets of aberrations, and the performance using the former is better than that using the latter. The performance using a fixed set of statistical optimal parameters is even close to that using the respective optimal parameters corresponding to each set of aberrations.

CONCLUSION:

The vSHWS configured with a fixed set of statistical optimal parameters can be used for high-precision aberration measurement of both normal and abnormal eyes. The statistical optimal parameters are more suitable for vSHWS than the parameters obtained with a set of statistical aberrations. These conclusions are significant for the designs of vSHWS and also SHWS.