Effect of modulation factor and low dose threshold level on gamma pass rates of single isocenter multi-target SRT treatment plans.
Timakova, Elena; Zavgorodni, Sergei F.
Afiliación
  • Timakova E; University of Victoria, Victoria, British Columbia, Canada.
  • Zavgorodni SF; BC Cancer Agency, Vancouver Island Centre, Victoria, British Columbia, Canada.
J Appl Clin Med Phys ; : e14459, 2024 Jul 25.
Article en En | MEDLINE | ID: mdl-39053489
ABSTRACT

PURPOSE:

SRS MapCHECK (SMC) is a commercially available patient-specific quality assurance (PSQA) tool for stereotactic radiosurgery (SRS) applications. This study investigates the effects of degree of modulation, location off-axis, and low dose threshold (LDT) selection on gamma pass rates (GPRs) between SMC and treatment planning system, Analytical Anisotropic Algorithm (AAA), or Vancouver Island Monte Carlo (VMC++ algorithm) system calculated dose distributions.

METHODS:

Volumetric-modulated arc therapy (VMAT) plans with modulation factors (MFs) ranging from 2.7 to 10.2 MU/cGy were delivered to SMC at isocenter and 6 cm off-axis. SMC measured dose distributions were compared against AAA and VMC++ via gamma analysis (3%/1 mm) with LDT of 10% to 80% using SNC Patient software.

RESULTS:

Comparing on-axis SMC dose against AAA and VMC++ with LDT of 10%, all AAA-calculated plans met the acceptance criteria of GPR ≥ 90%, and only one VMC++ calculated plan was marginally outside the acceptance criteria with pass rate of 89.1%. Using LDT of 80% revealed decreasing GPR with increasing MF. For AAA, GPRs reduced from 100% at MF of 2.7 MU/cGy to 57% at MF of 10.2 MU/cGy, and for VMC++ calculated plans, the GPRs reduced from 89% to 60% in the same MF range. Comparison of SMC dose off-axis against AAA and VMC++ showed more pronounced reduction of GPR with increasing MF. For LDT of 10%, AAA GPRs reduced from 100% to 83% in the MF range of 2.7 to 9.8 MU/cGy, and VMC++ GPR reduced from 100% to 91% in the same range. With 80% LDT, GPRs dropped from 100% to 42% for both algorithms.

CONCLUSIONS:

MF, dose calculation algorithm, and LDT selections are vital in VMAT-based SRT PSQA. LDT of 80% enhances sensitivity of gamma analysis for detecting dose differences compared to 10% LDT. To achieve better agreement between calculated and SMC dose, it is recommended to limit the MF to 4.6 MU/cGy on-axis and 3.6 MU/cGy off-axis.