RESUMEN
BACKGROUND: Isavuconazole is a triazole antifungal agent for treatment of invasive aspergillosis and mucormycosis. At present, it is unclear whether a therapeutic drug monitoring (TDM) of isavuconazole would be necessary. The aim of the investigation was to validate a high-performance liquid chromatography (HPLC) assay for routine applications. METHODS: An HPLC assay for routine determination of isavuconazole in plasma has been adapted and validated. The assay used the reagents and HPLC column provided by the ChromSystems HPLC Kit for TDM of itraconazole, posaconazole, and voriconazole. Isocratic flow rate was set at 1.0 mL/min. Detection was performed using a fluorescence detector with excitation wavelength set at 261 nm and emission wavelength set at 366 nm. RESULTS: The assay was linear between 0.15 and 10 mg/L with intraday and interday imprecision and accuracy <10% (<20% at lower limit of quantification). The method was applied to routine TDM of 7 patients after hematopoietic stem cell transplantation (n = 31 samples). In these patients, trough levels ranged from 0.45 to 3.06 mg/L (median 1.44 mg/L). CONCLUSIONS: A robust and simple HPLC assay of isavuconazole in plasma for routine TDM applications is presented.