The purpose of this article is to describe the setting up of 15 blood gas analyzers GEM(®) Premier™ 4000 (IL) at Montpellier hospital. This experience includes analytical characterization (within and between run coefficient of variation) using GSE and GHE IL controls, correlation of 35 samples with a routinely used laboratory blood gas analyzer (Cobas b221, Roche(®)). We shall also develop the training, the habilitation and its follow-up for the user staff (450 people) of the different hospital's units in the aim of the accreditation.


Asunto(s)
Equipo Médico Durable , Gases/análisis , Gases/sangre , Laboratorios de Hospital , Sistemas de Atención de Punto , Automatización de Laboratorios , Análisis de los Gases de la Sangre/instrumentación , Análisis de los Gases de la Sangre/normas , Calibración , Equipo Médico Durable/normas , Estudios de Seguimiento , Francia , Hospitales Universitarios/organización & administración , Hospitales Universitarios/normas , Humanos , Laboratorios de Hospital/organización & administración , Laboratorios de Hospital/normas , Personal de Laboratorio Clínico/educación , Personal de Laboratorio Clínico/normas , Sistemas de Atención de Punto/organización & administración , Sistemas de Atención de Punto/normas , Reproducibilidad de los Resultados , Sensibilidad y Especificidad , Recursos Humanos