J Magn Reson Imaging ; 10(2): 216-9, 1999 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-10441028

RESUMEN

This article presents a miniaturized electron spin resonance (ESR) probe for deducing the position of a surgical instrument on an MR image. The ESR probe constructed was small enough to fit inside a 14-G biopsy needle sheath, and position information of the sheath could be acquired using a simple gradient sequence. The position accuracy was estimated from needle trajectories as inferred from the needle artifact, the actual physical trajectory, and measured coordinates. The probe was able to track the tip of a biopsy needle quickly (10 samples/sec) and precisely with accuracy better than +/-2 mm. J. Magn. Reson. Imaging 1999;10:216-219.


Asunto(s)
Espectroscopía de Resonancia por Spin del Electrón/instrumentación , Imagen por Resonancia Magnética/instrumentación , Artefactos , Biopsia con Aguja/instrumentación , Biopsia con Aguja/métodos , Diseño de Equipo , Estudios de Evaluación como Asunto , Humanos , Imagen por Resonancia Magnética/métodos , Fantasmas de Imagen
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA