RESUMO
BACKGROUND: Data on procedural and early outcomes after transjugular transcatheter tricuspid valve replacement (TTVR) are limited. OBJECTIVES: This study sought to evaluate first-in-man procedural and clinical outcomes after transjugular TTVR with a special focus on patients who received large device sizes in whom TTVR outcomes have been questioned. METHODS: The retrospective registry included patients who underwent TTVR using the LuX-Valve Plus system (Jenscare Biotechnology Co Ltd) for symptomatic tricuspid regurgitation (TR) from January 2022 until February 2024 at 15 international centers in a compassionate use setting. The endpoints were procedural TR reduction, in-hospital death, adverse events, and 1-month survival. We further stratified results according to the size of the implanted device (<55 vs ≥55 mm). RESULTS: The registry included a total of 76 patients at a median age of 78 years (Q1-Q3: 72-83 years, 47.4% women). TR was reduced to ≤2+ and ≤1+ in 94.7% and 90.8% of patients (75.0% of patients received TTVR devices ≥55 mm) with well-sustained results at 1-month follow-up (TR ≤2+ in 95.0% and ≤1+ 86.8%). Residual TR was paravalvular in all cases. In-hospital death occurred in 4 patients (5.3%). Four patients (5.3%) underwent cardiac surgery during index hospitalization. Major in-hospital bleeding events occurred in 5 patients (6.6%). New in-hospital pacemaker implantation was required in 3.9% of patients in the overall cohort (5.7% in "pacemaker-naive" individuals). No cases of valve thrombosis, stroke, myocardial infarction, or pulmonary embolism were observed. At 1-month follow-up, survival was 94.4%, and NYHA functional class significantly improved. One further patient received a pacemaker, 1 further bleeding event occurred, and 2 patients underwent reintervention or surgery within the first 30 days after TTVR. No differences in procedural outcomes or adverse events were observed after stratification for valve size. CONCLUSIONS: Transjugular TTVR appears to be a safe and effective treatment option for patients with severe TR with comparable outcomes in very large tricuspid anatomies.
Assuntos
Cateterismo Cardíaco , Ensaios de Uso Compassivo , Implante de Prótese de Valva Cardíaca , Próteses Valvulares Cardíacas , Desenho de Prótese , Recuperação de Função Fisiológica , Sistema de Registros , Insuficiência da Valva Tricúspide , Valva Tricúspide , Humanos , Feminino , Masculino , Estudos Retrospectivos , Idoso , Resultado do Tratamento , Fatores de Tempo , Idoso de 80 Anos ou mais , Valva Tricúspide/cirurgia , Valva Tricúspide/fisiopatologia , Valva Tricúspide/diagnóstico por imagem , Insuficiência da Valva Tricúspide/cirurgia , Insuficiência da Valva Tricúspide/fisiopatologia , Insuficiência da Valva Tricúspide/mortalidade , Implante de Prótese de Valva Cardíaca/instrumentação , Implante de Prótese de Valva Cardíaca/efeitos adversos , Implante de Prótese de Valva Cardíaca/mortalidade , Cateterismo Cardíaco/efeitos adversos , Cateterismo Cardíaco/instrumentação , Cateterismo Cardíaco/mortalidade , Fatores de Risco , Mortalidade Hospitalar , Complicações Pós-Operatórias/etiologia , Veias JugularesRESUMO
BACKGROUND: Left atrial appendage (LAA) occluder embolization is an infrequent but serious complication. OBJECTIVES: We aim to describe timing, management and clinical outcomes of device embolization in a multi-center registry. METHODS: Patient characteristics, imaging findings and procedure and follow-up data were collected retrospectively. Device embolizations were categorized according to 1) timing 2) management and 3) clinical outcomes. RESULTS: Sixty-seven centers contributed data. Device embolization occurred in 108 patients. In 70.4 % of cases, it happened within the first 24 h of the procedure. The device was purposefully left in the LA and the aorta in two (1.9 %) patients, an initial percutaneous retrieval was attempted in 81 (75.0 %) and surgery without prior percutaneous retrieval attempt was performed in 23 (21.3 %) patients. Two patients died before a retrieval attempt could be made. In 28/81 (34.6 %) patients with an initial percutaneous retrieval attempt a second, additional attempt was performed, which was associated with a high mortality (death in patients with one attempt: 2.9 % vs. second attempt: 21.4 %, p < 0.001). The primary outcome (bailout surgery, cardiogenic shock, stroke, TIA, and/or death) occurred in 47 (43.5 %) patients. Other major complications related to device embolization occurred in 21 (19.4 %) patients. CONCLUSIONS: The majority of device embolizations after LAA closure occurs early. A percutaneous approach is often the preferred method for a first rescue attempt. Major adverse event rates, including death, are high particularly if the first retrieval attempt was unsuccessful. CONDENSED ABSTRACT: This dedicated multicenter registry examined timing, management, and clinical outcome of device embolization. Early embolization (70.4 %) was most frequent. As a first rescue attempt, percutaneous retrieval was preferred in 75.0 %, followed by surgical removal (21.3 %). In patients with a second retrieval attempt a higher mortality (death first attempt: 2.9 % vs. death second attempt: 24.1 %, p < 0.001) was observed. Mortality (10.2 %) and the major complication rate after device embolization were high.
Assuntos
Apêndice Atrial , Fibrilação Atrial , Cateterismo Cardíaco , Remoção de Dispositivo , Sistema de Registros , Humanos , Masculino , Apêndice Atrial/diagnóstico por imagem , Apêndice Atrial/fisiopatologia , Feminino , Idoso , Estudos Retrospectivos , Resultado do Tratamento , Fatores de Tempo , Idoso de 80 Anos ou mais , Fatores de Risco , Cateterismo Cardíaco/efeitos adversos , Cateterismo Cardíaco/instrumentação , Cateterismo Cardíaco/mortalidade , Fibrilação Atrial/terapia , Fibrilação Atrial/mortalidade , Remoção de Dispositivo/efeitos adversos , Embolia/etiologia , Embolia/mortalidade , Pessoa de Meia-Idade , Dispositivo para Oclusão Septal , Oclusão do Apêndice Atrial EsquerdoRESUMO
BACKGROUND: Several orthotopic transcatheter strategies have been developed to treat severe tricuspid regurgitation (TR); however, many patients are deemed unsuitable. Caval valve implantation with the TricValve system addresses this unmet need. OBJECTIVES: This study sought to determine the impact of TricValve on systemic congestion and quality of life (QOL) at 1 year. METHODS: The TRICUS (Safety and Efficacy of the TricValve® Transcatheter Bicaval Valves System in the Superior and Inferior Vena Cava in Patients With Severe Tricuspid Regurgitation) and TRICUS EURO studies were prospective, nonblinded, nonrandomized, single-arm trials representing the early-in-man experience of the TricValve system in NYHA functional class III or IV severe TR patients, optimally medicated and ineligible for open heart surgery, with significant caval backflow. The primary endpoint was QOL metrics and functional status. The 1-year results of the combined cohort are described here. RESULTS: Forty-four patients were included. Mean age was 76.2 ± 7.5 years, 81.0% were women, and the TRISCORE (risk score model for isolated tricuspid valve surgery) was 5.3 ± 1.3. Clinical improvement at 1 year was achieved in 42 (95.5%) patients, measured by (at least 1 of) an increase in ≥15 points from baseline in 12-item Kansas City Cardiomyopathy Questionnaire score, improvement to NYHA functional class to I or II, or an increase ≥40 m in the 6-minute walk test. There were 3 (6.8%) deaths at 1-year follow-up (1 cardiovascular), and the heart failure rehospitalization rate was 29.5%. Stent fracture, conduction system disturbances, or clinically significant leaflet thrombosis were not detected. Abolished hepatic vein backflow was achieved and persisted in 63.8% of the patients, contributing towards a reduction in congestive symptoms, N-terminal pro-B-type natriuretic peptide levels (P = 0.032), and diuretic treatment. CONCLUSIONS: Caval valve implantation with the TricValve system associated with meaningful 1-year clinical improvements in terms of QOL along with relatively low mortality rates. (TRICUS Study - Safety and Efficacy of the TricValve® Device; NCT03723239).
Assuntos
Implante de Prótese de Valva Cardíaca , Insuficiência da Valva Tricúspide , Idoso , Idoso de 80 Anos ou mais , Feminino , Humanos , Masculino , Cateterismo Cardíaco , Seguimentos , Estudos Prospectivos , Qualidade de Vida , Índice de Gravidade de Doença , Resultado do Tratamento , Valva Tricúspide/diagnóstico por imagem , Valva Tricúspide/cirurgia , Insuficiência da Valva Tricúspide/diagnóstico por imagem , Insuficiência da Valva Tricúspide/cirurgiaRESUMO
BACKGROUND: Myocardial revascularization failure (MRF) and Secondary revascularization (SR) are contemporary interventional cardiology challenges. AIM: To investigate the characteristics, management, and prognosis of patients with myocardial revascularization failure (MRF) and need for secondary revascularization (SR) in contemporary practice. METHODS: The REVASEC study is a prospective registry (NCT03349385), which recruited patients with prior revascularization referred for coronary angiography at 19 centers. The primary endpoint is a patient-oriented composite (POCE) at 1 year, including death, myocardial infarction, or repeat revascularization. RESULTS: A total of 869 patients previously revascularized by percutaneous intervention (83%) or surgery (17%) were recruited. MRF was found in 83.7% (41.1% stent/graft failure, 32.1% progression of coronary disease, and 10.5% residual disease). SR was performed in 70.1%, preferably by percutaneous intervention (95%). The POCE rate at 1 year was 14% in the overall cohort, with 6.4% all-cause death. In the multivariate analysis, lower POCE rates were found in the groups without MRF (9.4%) and with disease progression (11%) compared with graft/stent failure (17%) and residual disease (18%), hazard ratio 0.67 (95% confidence interval: 0.45-0.99), p = 0.043. At 1 year, the SR group had less chronic persistent angina (19% vs. 34%, p < 0.001), but a higher rate of repeat revascularization (9% vs. 2.9%, p < 0.001). CONCLUSION: MRF was found in 84% of patients with prior revascularization referred for coronary angiography. Stent/graft failure and residual coronary disease were associated with a worse prognosis. SR provided better symptom control at the expense of a higher rate of new revascularization.
RESUMO
INTRODUCTION AND OBJECTIVES: Transthyretin cardiac amyloidosis (ATTR-CA) patients often have atrial fibrillation and increased bleeding/thrombogenic risks. We aimed to evaluate outcomes of left atrial appendage closure (LAAC) compared with patients without a known diagnosis of CA. METHODS: Comparison at long-term of patients diagnosed with ATTR-CA who underwent LAAC between 2009 and 2020 and those without a known diagnosis of CA. RESULTS: We studied a total of 1159 patients. Forty patients (3.5%) were diagnosed with ATTR-CA; these patients were older and had more comorbidities, higher HAS-BLED and CHA2DS2-VASc scores, and lower left ventricular function. Successful LAAC was achieved in 1137 patients (98.1%) with no differences between groups. Regarding in-hospital and follow-up complications, there were no differences between the groups in ischemic stroke (5% vs 2.5% in those without a known diagnosis of CA; P=.283), hemorrhagic stroke (2.5% and 0.8% in the control group; P=.284), major or minor bleeding. At the 2-year follow-up, there were no significant differences in mortality (ATTR-CA: 20% vs those without known CA: 13.6%, 0.248); however, the at 5-year follow-up, ATTR-CA patients had higher mortality (40% vs 19.2%; P <.001) but this difference was unrelated to hemorrhagic complications or ischemic stroke. CONCLUSIONS: LAAC could reduce the risk of bleeding complications and ischemic cerebrovascular events without increasing the rate of early or mid-term complications. Although long-term survival was impaired in ATTR-CA patients, it was comparable to that of patients without a known diagnosis of CA at the 2-year follow-up, suggesting that LAAC for patients with ATTR-CA might not be futile.
Assuntos
Amiloidose , Apêndice Atrial , Fibrilação Atrial , AVC Isquêmico , Acidente Vascular Cerebral , Humanos , Acidente Vascular Cerebral/epidemiologia , Acidente Vascular Cerebral/etiologia , Acidente Vascular Cerebral/prevenção & controle , Apêndice Atrial/diagnóstico por imagem , Apêndice Atrial/cirurgia , Hemorragia/complicações , Fibrilação Atrial/complicações , Fibrilação Atrial/cirurgia , Fibrilação Atrial/diagnóstico , Amiloidose/complicações , Amiloidose/diagnóstico , Resultado do TratamentoRESUMO
BACKGROUND: Left atrial appendage closure (LAAC) aims to prevent ischemic events in patients with atrial fibrillation. As a preventive procedure, early death after LAAC could render the procedure futile. OBJECTIVES: The authors sought to evaluate the incidence and factors associated with early death in LAAC recipients. METHODS: This was a multicenter study including consecutive patients undergoing LAAC in a 10-year period (2009-2019). Death was considered early when occurring in the first year after LAAC. RESULTS: A total of 807 patients (mean age 76 ± 8 years, mean CHA2DS2-VASc score 4.5 ± 1.5) were included. Early death occurred in 125 patients (15.5%). In the multivariable analysis, factors associated with early death after LAAC were older age HR: 1.03; 95% CI: 1.01-1.06 per year; P = 0.01), lower body mass index (HR: 0.92; 95% CI: 0.88-0.97 per 1 kg/m2 increase; P < 0.001), diabetes (HR: 1.71; 95% CI: 1.19-2.47; P = 0.002), prior heart failure (HR: 1.74; 95% CI: 1.20-2.53; P = 0.001), and lower estimated glomerular filtration rate (HR: 1.09; 95% CI: 1.05-1.13 per 5 mL/min/1.73 m2 decrease; P < 0.001). There was a stepwise increase in risk of early death within the first year of LAAC with the combination of different risk factors (up to 48.9% in the presence of >3 risk factors). CONCLUSIONS: In this multicenter international registry, close to 1 in 6 patients died within the first year of LAAC. Older age, low body mass index, impaired estimated glomerular filtration rate, prior diabetes, and prior heart failure are independently associated with an increased risk. The risk of early death appeared to be prohibitive (â¼50%) in the presence of >3 of these risk factors.
Assuntos
Apêndice Atrial , Fibrilação Atrial , Insuficiência Cardíaca , Acidente Vascular Cerebral , Idoso , Idoso de 80 Anos ou mais , Apêndice Atrial/cirurgia , Fibrilação Atrial/complicações , Fibrilação Atrial/epidemiologia , Fibrilação Atrial/cirurgia , Humanos , Incidência , Acidente Vascular Cerebral/epidemiologiaRESUMO
BACKGROUND: Paravalvular leak occurs in 5-17% of patients following surgical valve replacement, more often in mitral position. The prognosis without treatment is poor. Percutaneous device closure represents an alternative to repeat surgery. The objective of this work is to evaluate the medium and long-term results in the percutaneous closure of PVL in mitral prosthesis. METHODS: This observational study is based on a retrospective registry including consecutive mitral PVL cases undergoing percutaneous closure at a single tertiary-care center from April 2010 to December 2020. The safety and efficacy results of the procedure, at 90 days and in the long term, were analyzed. Also, predictors of procedure failure and long-term events were identified. RESULTS: A total of 128 consecutive mitral paravalvular leak closure procedures were included. Technical success was achieved in 115 (89.8%) procedures. The presence of multiple PVLs was the sole factor that independently predicted procedural failure. Median follow-up of our sample was 41.8 months (mean 47.7 ± 35.7 months). Underlying hemolytic anemia as the indication for PVL closure, a recent admission for decompensated HF, and lack of improvement in functional class emerged as consistent predictors of MACE and death during long-term follow-up, while lack of procedural success during the first PVL procedure and chronic kidney disease were also associated with MACE during follow-up. CONCLUSIONS: Percutaneous mitral PVL closure displayed high technical and procedural success rates, with an acceptable safety profile, in a high-risk population. Percutaneous mitral PVL closure achieved an improvement in short- and long-term functional class and a reduction of hemolysis in the vast majority of patients. In addition, long-term survival in our study was good, in particular for patients undergoing successful PVL closure procedures.
RESUMO
BACKGROUND: Severe tricuspid regurgitation (TR) is frequently associated with significant morbidity and mortality; such patients are often deemed to be at high surgical risk. Heterotopic bicaval stenting is an emerging, attractive transcatheter solution for these patients. OBJECTIVES: The aim of this study was to evaluate the 30-day safety and 6-month efficacy outcomes of specifically designed bioprosthetic valves for the superior and inferior vena cava. METHODS: TRICUS EURO (Safety and Efficacy of the TricValve® Transcatheter Bicaval Valves System in the Superior and Inferior Vena Cava in Patients With Severe Tricuspid Regurgitation) is a nonblinded, nonrandomized, single-arm, multicenter, prospective trial that enrolled patients from 12 European centers between December 2019 and February 2021. High-risk individuals with severe symptomatic TR despite optimal medical therapy were included. The primary endpoint was quality-of-life (QOL) improvement measured by Kansas City Cardiomyopathy Questionnaire score and New York Heart Association (NYHA) functional class improvement at 6-month follow-up. RESULTS: Thirty-five patients (mean age 76 ± 6.8 years, 83% women) were treated using the TricValve system. All patients at baseline were in NYHA functional class III or IV. At 30 days, procedural success was 94%, with no procedural deaths or conversions to surgery. A significant increase in QOL at 6 months follow-up was observed (baseline and 6-month Kansas City Cardiomyopathy Questionnaire scores 42.01 ± 22.3 and 59.7 ± 23.6, respectively; P = 0.004), correlating with a significant improvement in NYHA functional class, with 79.4% of patients noted to be in functional class I or II at 6 months (P = 0.0006). The rates of 6-month all-cause mortality and heart failure hospitalization were 8.5% and 20%, respectively. CONCLUSIONS: The dedicated bicaval system for treating severe symptomatic TR was associated with a high procedural success rate and significant improvements in both QOL and functional classification at 6 months follow-up.
Assuntos
Cardiomiopatias , Implante de Prótese de Valva Cardíaca , Insuficiência da Valva Tricúspide , Idoso , Idoso de 80 Anos ou mais , Cateterismo Cardíaco/efeitos adversos , Feminino , Implante de Prótese de Valva Cardíaca/efeitos adversos , Humanos , Masculino , Estudos Prospectivos , Qualidade de Vida , Índice de Gravidade de Doença , Resultado do Tratamento , Valva Tricúspide/diagnóstico por imagem , Valva Tricúspide/cirurgia , Insuficiência da Valva Tricúspide/diagnóstico por imagem , Insuficiência da Valva Tricúspide/cirurgiaRESUMO
BACKGROUND: Transcatheter Aortic Valve Implantation (TAVI) is beneficial in patients with symptomatic severe Aortic Stenosis (AS). There is no consensus about the best anticoagulation strategy for patients with a recent TAVI and with atrial fibrillation (AF). Direct oral anticoagulants (DOACs) are effective to prevent embolic events with a significant lower incidence of bleeding. There is scarce evidence about the use of these drugs in patients undergoing TAVI. AIM: To assess the management of anticoagulation at the moment of discharge of patients with AF and TAVI. Material and Methods: A four question survey was sent to cardiologists involved in TAVI programs in different international centers. Results: The survey was answered by 72 interventional cardiologists. Even with the lack of randomized evidence, in most of the scenarios DOACs are prescribed at discharge in patients with indication for anticoagulation. Also, in patients with high bleeding risk, most cardiologists would perform a left atrial appendage closure. In patients with concomitant coronary artery disease, if a stent was recently implanted, prescription of the combination of a DOAC and one antiplatelet drug was the most common answer. In patients with a former coronary angioplasty, DOAC or Warfarin was the therapy of choice. CONCLUSIONS: In the absence of randomized data, interventional cardiologists prescribe DOACs at discharge to patients with AF and TAVI, without following current guidelines in most cases.
Assuntos
Humanos , Estenose da Valva Aórtica/cirurgia , Estenose da Valva Aórtica/complicações , Fibrilação Atrial/cirurgia , Fibrilação Atrial/complicações , Fibrilação Atrial/tratamento farmacológico , Substituição da Valva Aórtica Transcateter/efeitos adversos , Varfarina/efeitos adversos , Resultado do Tratamento , Hemorragia/induzido quimicamente , Anticoagulantes/uso terapêuticoRESUMO
BACKGROUND: Patients with severe tricuspid regurgitation present late and are often ineligible for surgery or transcatheter repair systems. Transfemoral venous implantation of a bicaval valved stent has been proposed as therapeutic option in selected patients. The aim of this study was to summarize the early procedural and clinical outcomes of the novel TRICENTO system for the treatment of patients with symptomatic severe tricuspid regurgitation. METHODS: All consecutive patients treated with the custom-made TRICENTO implant at the participating centers were included in this retrospective multicentre registry. RESULTS: A total of 21 high-risk patients (mean age 76±7 years; 67% female) with severe or higher grade tricuspid regurgitation were analyzed. The majority of the patients were in New York Heart Association class III/IV (95%), had peripheral edema (95%), and previous hospitalization for right heart failure (67%). Technical success was 100%, and there was no case of in-hospital mortality. During follow-up (median 61 days), symptomatic improvement was observed (65% in New York Heart Association class I/II; P<0.001). Computed tomography revealed asymptomatic fractures of the TRICENTO prosthesis in 3 patients. Cardiac magnetic resonance imaging obtained in 7 patients showed a significant decrease (252±65 mm3 at baseline versus 216±58 mm3 at follow-up, P=0.006) of right ventricular end-diastolic volume. The overall-survival rate was 76% at 1 year. CONCLUSIONS: The present data indicate the feasibility of transfemoral bicaval valved stent implantation for the treatment of severe tricuspid regurgitation. Functional improvement and signs of right ventricular reverse remodeling were observed. Stent fractures did not impair valve function, but require refinement of prosthesis design and careful assessment of eligibility criteria.
Assuntos
Implante de Prótese de Valva Cardíaca , Insuficiência da Valva Tricúspide , Idoso , Idoso de 80 Anos ou mais , Cateterismo Cardíaco , Feminino , Humanos , Masculino , Sistema de Registros , Estudos Retrospectivos , Índice de Gravidade de Doença , Stents , Fatores de Tempo , Resultado do Tratamento , Valva Tricúspide/diagnóstico por imagem , Valva Tricúspide/cirurgia , Insuficiência da Valva Tricúspide/diagnóstico por imagem , Insuficiência da Valva Tricúspide/cirurgiaRESUMO
INTRODUCTION AND OBJECTIVES: Transcatheter edge-to-edge repair (TEER) should be considered in patients with heart failure and secondary mitral regurgitation (MR). Angiotensin receptor-neprilysin inhibitors (ARNIs) have been demonstrated to improve prognosis in heart failure. We aimed to evaluate the impact ARNIs on patient selection and outcomes. METHODS: The population of the Spanish TEER prospective registry (March 2012 to January 2021) was divided into 2 groups: a) TEER before the ARNI era (n=450) and b) TEER after the recommendation of ARNIs by European Guidelines (n=639), with further analysis according to intake (n=52) or not (n=587) of ARNIs. RESULTS: A total of 1089 consecutive patients underwent TEER for secondary MR. In the ARNI era, there was a reduction in left ventricle dilation (82mL vs 100mL, P=.025), and better function (35% vs 38%, P=.011). At 2 years of follow-up, mortality (10.6% vs 17.3%, P <.001) and heart failure readmissions (16.6% vs 27.8%, P <.001) were lower in the ARNI era, but not recurrent MR. In the ARNI era, 1- and 2-year mortality were similar irrespective of ARNI intake but patients on ARNIs had a lower risk of readmission+mortality at 2 years (OR, 0.369; 95%CI, 0.137-0.992; P=.048), better NYHA class, and lower recurrence of MR III-IV (1.9% vs 14.3%, P=.011). CONCLUSIONS: Better patient selection for TEER has been achieved in the last few years with a parallel improvement in outcomes. The use of ARNIs was associated with a significant reduction in overall events, better NYHA class, and lower MR recurrence.
Assuntos
Insuficiência Cardíaca , Insuficiência da Valva Mitral , Neprilisina , Humanos , Insuficiência Cardíaca/tratamento farmacológico , Valva Mitral/diagnóstico por imagem , Valva Mitral/cirurgia , Insuficiência da Valva Mitral/tratamento farmacológico , Insuficiência da Valva Mitral/cirurgia , Neprilisina/antagonistas & inibidores , Receptores de Angiotensina , Resultado do TratamentoRESUMO
AIM: To investigate key aspects of the problem of myocardial revascularization failure (MRF) and repeat or secondary myocardial revascularization (SR) in contemporary practice. METHODS: The registry of secondary revascularization (REVASEC) is an investigator-initiated, multicenter, prospective registry enhanced with data monitoring and independent event adjudication (ClinicalTrials.govNCT03349385). It includes patients with prior revascularization referred to coronary angiography for suspected MRF with broad inclusion criteria. The main objectives are to describe the characteristics of patients with prior revascularization referred for repeat angiography, to describe and the rate and mechanisms of MRF (stent or graft failure, coronary artery disease progression or residual coronary artery disease); to evaluate the management including medical treatment and SR of these patients; and to assess the prognosis according to the outlined causative mechanisms. The registry has one year follow up for the primary endpoint (Patient-oriented composite endpoint including all-cause death, any myocardial infarction or any new unplanned revascularization according to subsets of MRF), but extended follow-up will be carried out up to 5 years. CONCLUSION: The REVASEC Registry will provide updated data on the characteristics, patterns of treatment, and 1-year outcomes of patients with MRF and SR in contemporary clinical practice.
Assuntos
Doença da Artéria Coronariana , Insuficiência Cardíaca , Intervenção Coronária Percutânea , Ponte de Artéria Coronária/efeitos adversos , Doença da Artéria Coronariana/diagnóstico por imagem , Doença da Artéria Coronariana/cirurgia , Insuficiência Cardíaca/etiologia , Humanos , Revascularização Miocárdica/efeitos adversos , Intervenção Coronária Percutânea/efeitos adversos , Sistema de Registros , Fatores de Risco , Fatores de Tempo , Resultado do TratamentoRESUMO
AIMS: The SYNTAX II study evaluated the impact of advances in percutaneous coronary intervention (PCI), integrated into a single revascularization strategy, on outcomes of patients with de novo three-vessel disease. The study employed decision-making utilizing the SYNTAX score II, use of coronary physiology, thin-strut biodegradable polymer drug-eluting stents, intravascular ultrasound, enhanced treatments of chronic total occlusions, and optimized medical therapy. Patients treated with this approach were compared with predefined patients from the SYNTAX I trial. METHODS AND RESULTS: SYNTAX II was a multicentre, single-arm, open-label study of patients requiring revascularization who demonstrated clinical equipoise for treatment with either coronary artery bypass grafting (CABG) or PCI, predicted by the SYNTAX score II. The primary endpoint was major adverse cardiac and cerebrovascular events (MACCE), which included any revascularization. The comparators were a matched PCI cohort trial and a matched CABG cohort, both from the SYNTAX I trial. At 5 years, MACCE rate in SYNTAX II was significantly lower than in the SYNTAX I PCI cohort (21.5% vs. 36.4%, P < 0.001). This reflected lower rates of revascularization (13.8% vs. 23.8%, P < 0.001), and myocardial infarction (MI) (2.7% vs. 10.4%, P < 0.001), consisting of both procedural MI (0.2% vs. 3.8%, P < 0.001) and spontaneous MI (2.3% vs. 6.9%, P = 0.004). All-cause mortality was lower in SYNTAX II (8.1% vs. 13.8%, P = 0.013) reflecting a lower rate of cardiac death (2.8% vs. 8.4%, P < 0.001). Major adverse cardiac and cerebrovascular events' outcomes at 5 years among patients in SYNTAX II and predefined patients in the SYNTAX I CABG cohort were similar (21.5% vs. 24.6%, P = 0.35). CONCLUSIONS: Use of the SYNTAX II PCI strategy in patients with de novo three-vessel disease led to improved and durable clinical results when compared to predefined patients treated with PCI in the original SYNTAX I trial. A predefined exploratory analysis found no significant difference in MACCE between SYNTAX II PCI and matched SYNTAX I CABG patients at 5-year follow-up.
Assuntos
Doença da Artéria Coronariana , Stents Farmacológicos , Intervenção Coronária Percutânea , Ponte de Artéria Coronária/métodos , Doença da Artéria Coronariana/terapia , Humanos , Intervenção Coronária Percutânea/métodos , Resultado do TratamentoRESUMO
Contrast-induced nephropathy (CIN) is a complication associated with the administration of contrast media (CM). The CIN diagnosis is based on creatinine, a biomarker late and insensitive. The objective proposed was to evaluate the ability of novel biomarkers to detect patients susceptible to suffering CIN before CM administration. The study was carried out with patients undergoing cardiac catheterization involving CM. Patients were divided into two groups: (1) CIN, patients who developed this pathology; (2) control, patients who did not suffer CIN. Prior to the administration of CM, urine samples were collected to measure proteinuria, N-acetyl-ß-d-glucosaminidase, neutrophil gelatinase-associated lipocalin and kidney injury molecule-1, albumin, transferrin, t-gelsolin and GM2 ganglioside activator protein (GM2AP). The risk factors advanced age, low body mass index and low estimated glomerular filtration rate; and the urinary biomarkers albumin, transferrin and GM2AP showed significant predictive capacity. Of all of them, albuminuria demonstrated the highest diagnostic power. When a cutoff point was established for albuminuria at values still considered subclinical (10-30 µg/mg Cru), it was found that there was a high incidence of CIN (40-75%). Therefore, albuminuria could be applied as a new diagnostic tool to prevent and predict CIN with P4 medicine criteria, independently of risk factors and comorbidities.
RESUMO
BACKGROUND: Gastrointestinal bleeding (GIB) is associated with a high recurrence rate and a prior GIB episode is common in real-world left atrial appendage closure (LAAC) recipients. The present study sought to evaluate the clinical characteristics and outcomes of patients with prior GIB undergoing LAAC, and to determine the factors associated with and clinical impact of GIB recurrence. METHODS: Multicenter study including 277 consecutive patients who underwent percutaneous LAAC and had prior GIB. All-cause death, all bleeding, GIB recurrence, and clinical ischemic stroke were recorded. RESULTS: After a median follow-up of 17 (interquartile range: 6-37) months post-LAAC, the rates of death, bleeding, GIB recurrence, and ischemic stroke were 14.0 per 100 person-year (PY), 29.3 per 100 PY, 17.7 per 100 PY, and 1.1 per 100 PY, respectively. GIB recurrence occurred within 3 months post-LAAC in 55.8% of patients. A previous lower GIB (vs. upper or unclassified) (HR: 1.76; 95% CI: 1.09-2.82; p = .020) and eGFR < 45 mL/min (HR: 1.70; 95% CI:1.04-2.67; p = .033) determined an increased risk of GIB recurrence. By multivariable analysis, eGFR < 45 mL/min (HR: 2.72; 95% CI: 1.70-4.34; p < .001), GIB recurrence following LAAC (HR: 2.15; 95% CI: 1.33-3.46; p = .002), diabetes mellitus (HR: 1.77; 95% CI: 1.10-2.84; p = .018), and age (HR: 1.06; 95% CI: 1.03-1.10; p < .001) were associated with an increased mortality. CONCLUSIONS: Patients with prior GIB undergoing LAAC exhibited a relatively low rate of GIB recurrence, and prior lower GIB and moderate-to-severe chronic kidney disease determined an increased risk. GIB recurrence was associated with an increased mortality.
Assuntos
Apêndice Atrial/cirurgia , Fibrilação Atrial/complicações , Fibrilação Atrial/cirurgia , Hemorragia Gastrointestinal/complicações , Hemorragia Gastrointestinal/epidemiologia , Idoso , Idoso de 80 Anos ou mais , Feminino , Humanos , Incidência , Masculino , Recidiva , Resultado do TratamentoRESUMO
BACKGROUND: Among patients with acute coronary syndrome following transcatheter aortic valve replacement (TAVR), those presenting with ST-segment elevation myocardial infarction (STEMI) are at highest risk. OBJECTIVES: The goal of this study was to determine the clinical characteristics, management, and outcomes of STEMI after TAVR. METHODS: This was a multicenter study including 118 patients presenting with STEMI at a median of 255 days (interquartile range: 9 to 680 days) after TAVR. Procedural features of STEMI after TAVR managed with primary percutaneous coronary intervention (PCI) were compared with all-comer STEMI: 439 non-TAVR patients who had primary PCI within the 2 weeks before and after each post-TAVR STEMI case in 5 participating centers from different countries. RESULTS: Median door-to-balloon time was higher in TAVR patients (40 min [interquartile range: 25 to 57 min] vs. 30 min [interquartile range: 25 to 35 min]; p = 0.003). Procedural time, fluoroscopy time, dose-area product, and contrast volume were also higher in TAVR patients (p < 0.01 for all). PCI failure occurred more frequently in patients with previous TAVR (16.5% vs. 3.9%; p < 0.001), including 5 patients in whom the culprit lesion was not revascularized owing to coronary ostia cannulation failure. In-hospital and late (median of 7 months [interquartile range: 1 to 21 months]) mortality rates were 25.4% and 42.4%, respectively (20.6% and 38.2% in primary PCI patients), and estimated glomerular filtration rate <60 ml/min (hazard ratio [HR]: 3.02; 95% confidence interval [CI]: 1.42 to 6.43; p = 0.004), Killip class ≥2 (HR: 2.74; 95% CI: 1.37 to 5.49; p = 0.004), and PCI failure (HR: 3.23; 95% CI: 1.42 to 7.31; p = 0.005) determined an increased risk. CONCLUSIONS: STEMI after TAVR was associated with very high in-hospital and mid-term mortality. Longer door-to-balloon times and a higher PCI failure rate were observed in TAVR patients, partially due to coronary access issues specific to the TAVR population, and this was associated with poorer outcomes.
Assuntos
Estenose da Valva Aórtica/cirurgia , Complicações Pós-Operatórias/etiologia , Medição de Risco/métodos , Infarto do Miocárdio com Supradesnível do Segmento ST/etiologia , Substituição da Valva Aórtica Transcateter/efeitos adversos , Idoso de 80 Anos ou mais , Angiografia Coronária , Feminino , Seguimentos , Saúde Global , Mortalidade Hospitalar/tendências , Humanos , Incidência , Masculino , Intervenção Coronária Percutânea , Complicações Pós-Operatórias/epidemiologia , Fatores de Risco , Infarto do Miocárdio com Supradesnível do Segmento ST/diagnóstico , Infarto do Miocárdio com Supradesnível do Segmento ST/epidemiologia , Fatores de TempoRESUMO
BACKGROUND AND AIM: Recent randomized data comparing percutaneous mitral valve repair (PMVR) versus optimal medical treatment in patients with functional MR (FMR) seemed to highlight the importance of the learning curve not only for procedural outcomes but also for patient selection. The aim of the study was to compare a contemporary series of patients undergoing PMVR using a second-generation Mitraclip device (Mitraclip NT) with previous cohorts treated with a first-generation system. METHODS: This multicenter study collected individual data from 18 centers between 2012 and 2017. The cohort was divided into three groups according to the use of the first-generation Mitraclip during the first (control-1) or second half (control-2) or the Mitraclip NT system. RESULTS: A total of 545 consecutive patients were included in the study. Among all, 182 (33.3%), 183 (33.3%), and 180 (33.3%) patients underwent mitral repair in the control-1, control-2, and NT cohorts, respectively. Procedural success was achieved in 93.3% of patients without differences between groups. Major adverse events did not statistically differ among groups, but there was a higher rate of pericardial effusion in the control-1 group (4.3%, 0.6%, and 2.6%, respectively; p = 0.025). The composite endpoint of death, surgery, and admission for congestive heart failure (CHF) at 12 months was lower in the NT group (23.5% in control-1, 22.5% in control-2, and 8.3% in the NT group; p = 0.032). CONCLUSIONS: The present paper shows that contemporary clinical outcomes of patients undergoing PMVR with the Mitraclip system have improved over time.
RESUMO
BACKGROUND: In catheter-based procedures, acute kidney injury (AKI) is a frequent, serious complication ranging from 10% to 30%. In MitraClip (Abbott Vascular, Santa Clara, California), a usually contrast-free procedure, there is scarce data about its real incidence and impact. OBJECTIVES: This study aimed to evaluate incidence, predictive factors, and midterm outcomes of AKI in patients with significant mitral regurgitation (MR) undergoing transcatheter valve repair with MitraClip. METHODS: A total of 721 patients undergoing MitraClip were included. AKI was defined as an absolute or a relative increase in serum creatinine of >0.3 mg/dl or ≥50%, respectively, or the need for hemodialysis during index hospitalization. RESULTS: The mean age of the patients was 72 ± 11 years (28.3% women). Median estimated glomerular filtration rate (eGFR) was 43.7 ml/min/1.73 m2 (interquartile range: 30.9 to 60.1 ml/min/1.73 m2), and was <60 ml/min/1.73 m2 in 74.9% of the patients. AKI after MitraClip occurred in 106 patients (14.7%). Baseline hemoglobin (<11 g/dl) (odds ratio [OR]: 1.97; p = 0.003), urgent procedure (OR: 3.44; p = 0.003), and absence of device success (OR: 3.37; p < 0.001) were independent predictors of AKI. Patients with AKI had worse outcomes compared to those without AKI, including a higher proportion of in-hospital bleeding events (3.8% vs. 0.8%; p = 0.011), 2-year all-cause mortality (40.5% vs. 18.7%; p <0.001), and major adverse cardiac events (63.6% vs. 23.5%; p <0.001). Combination of AKI with significant residual MR after the procedure conferred even worst outcomes (2-year all-cause mortality 50.0% vs. 19.6%; p = 0.001, and major adverse cardiac events 70.0% vs. 18.9%; p < 0.001). CONCLUSIONS: Despite being a "zero-contrast" procedure, one-sixth of patients undergoing transcatheter mitral valve repair had AKI, linked to device failure or other severe conditions. The occurrence of AKI was associated with worse outcomes, highlighting the importance to detect and reduce this complication in high-risk population.