PLoS One ; 19(3): e0299597, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-38452034

RESUMO

BACKGROUND: Extrusion of electrodes outside the cochlea and tip fold overs may lead to suboptimal outcomes in cochlear implant (CI) recipients. Intraoperative measures such as Trans-Impedance Matrix (TIM) measurements may enable clinicians to identify electrode malposition and direct surgeons to correctly place the electrode array during surgery. OBJECTIVES: To assess the current literature on the effectiveness of TIM measurements in identifying extracochlear electrodes and tip fold overs. METHODS: A scoping review of studies on TIM-based measurements were carried out using the Databases-Medline/PubMed, AMED, EMBASE, CINAHL and the Cochrane Library following PRISMA guidelines. Eleven full texts articles met the inclusion criteria. Only human studies pertaining to TIM as a tool used in CI were included in the review. Further, patient characteristics, electrode design, and TIM measurement outcomes were reported. RESULTS: TIM measurements were available for 550 implanted ears with the subjects age ranged between 9 months to 89 years. Abnormal TIM measurements were reported for 6.55% (36). Tip fold over was detected in 3.64% (20) of the cases, extracochlear electrodes in 1.45% (8), and 1.45% (8) were reported as buckling. Slim-modiolar electrode array designs were more common (54.71%) than pre-curved (23.34%) or lateral wall (21.95%) electrode array. Abnormal cochlear anatomy was reported for five ears (0.89%), with normal cochlear anatomy for all other patients. CONCLUSION: TIM measurement is a promising tool for the intraoperative detection of electrode malposition. TIM measurement has a potential to replace intraoperative imaging in future. Though, TIM measurement is in its early stages of clinical utility, intuitive normative data sets coupled with standardised criteria for detection of abnormal electrode positioning would enhance its sensitivity.


Assuntos
Implante Coclear , Implantes Cocleares , Humanos , Cóclea/cirurgia , Implante Coclear/métodos , Impedância Elétrica , Eletrodos Implantados , Resultado do Tratamento
2.
Clin Otolaryngol ; 47(6): 641-649, 2022 11.
Artigo em Inglês | MEDLINE | ID: mdl-35833359

RESUMO

OBJECTIVE: To investigate the value of scalp surface potentials to identify and manage partial short circuits to ground in cochlear implant electrodes. DESIGN: A retrospective review of patients with suspected partial short circuits. MAIN OUTCOME MEASURE: Electrical output of individual electrodes was measured using scalp surface potentials for patients reporting a change in hearing function. Electrical output was compared to functional performance and impedance measurements to determine if devices with suspected partial short circuits were experiencing a decrease in performance as a result of reduced electrical output. Electrical output was checked in an artificial cochlea for two implants following explant surgery to confirm scalp surface potential results. RESULTS: All patients with suspected partial short circuits (n = 49) had reduced electrical output, a drop in impedances to approximately ½ of previously stable measurements or to below 2 kΩ, an atypical electrical field measurement (EFI) and a decline in hearing function. Only devices with an atypical EFI showed reduced electrical output. Results of scalp based surface potentials could be replicated in an artificial cochlea following explantation of the device. All explant reports received to date (n = 42) have confirmed partial short circuits, with an additional four devices failing integrity tests. CONCLUSION: Surface potential measurements can detect partial shorts and had 100% correlation with atypical EFI measurements, which are characteristic of a partial short to ground in this device. Surface potentials can help determine the degree to which the electrode array is affected, particularly when behavioural testing is limited or not possible.


Assuntos
Implante Coclear , Implantes Cocleares , Cóclea , Implante Coclear/métodos , Impedância Elétrica , Testes Auditivos , Humanos , Couro Cabeludo/cirurgia
3.
Otol Neurotol ; 43(4): 452-459, 2022 04 01.
Artigo em Inglês | MEDLINE | ID: mdl-35085112

RESUMO

OBJECTIVES: To investigate the combined effect of changing pulse shape and grounding configuration to manage facial nerve stimulation (FNS) in cochlear implant (CI) recipients. PATIENTS: Three adult CI recipients with severe FNS were offered a replacement implant when standard stimulation strategies and programming adjustments did not resolve symptoms. Our hypothesis was that the facial nerve was less likely to be activated when using anodic pulses with "mixed-mode" intra-cochlear and extra-cochlear current return. INTERVENTION: All patients were reimplanted with an implant that uses a pseudo-monophasic anodic pulse shape, with mixed-mode grounding (stimulus mixed-mode anodic)-the Neuro Zti CI (Oticon Medical). This device also allows measurements of neural function and loudness with monopolar, symmetric biphasic pulses (stimulus MB), the clinical standard used by most CIs as a comparison. MAIN OUTCOME MEASURES: The combined effect of pulse shape and grounding configuration on FNS was monitored during surgery. Following CI activation, FNS symptoms and performance with the Neuro Zti implant were compared with outcomes before reimplantation. RESULTS: FNS could only be recorded using stimulus MB for all patients. In clinical use, all patients reported reduced FNS and showed an improvement in Bamford-Kowal-Bench sentences recognition compared with immediately before reimplantation. Bamford-Kowal-Bench scores with a male speaker were lower compared with measurements taken before the onset of severe FNS for patients 1 and 2. CONCLUSIONS: In patients where CI auditory performance was severely limited by FNS, charge-balanced pseudo-monophasic stimulation mode with a mixed-mode grounding configuration limited FNS and improved loudness percept compared with standard biphasic stimulation with monopolar grounding.


Assuntos
Implante Coclear , Implantes Cocleares , Adulto , Cóclea , Implante Coclear/efeitos adversos , Implantes Cocleares/efeitos adversos , Nervo Coclear , Estimulação Elétrica , Nervo Facial/cirurgia , Humanos , Masculino , Reimplante
4.
Otol Neurotol ; 42(7): 1014-1021, 2021 08 01.
Artigo em Inglês | MEDLINE | ID: mdl-33710152

RESUMO

OBJECTIVE: To determine if cochlear implantation without removal of inner ear schwannomas (IES) is an effective treatment option for Neurofibromatosis 2 (NF2) patients. To determine how the presence of an intracochlear schwannoma might impact cochlear implant (CI) outcomes and programming parameters. STUDY DESIGN: Retrospective chart review. SETTING: Tertiary center for cochlear and auditory brainstem implantation. PATIENTS: Of 10 NF2 patients with IES, 8 are reported with no previous tumor removal on the implanted ear. INTERVENTIONS: Cochlear implant without tumor removal. MAIN OUTCOME MEASURES: Performance outcomes with CI at least 1-year post intervention. Programming parameters, including impedances, for patients with IES. RESULTS: All patients had full insertion of the electrode arrays with round window approaches. Performance outcomes ranged from 0 to 100% for Bamford-Kowal-Bench sentences. Impedance measurements for active electrodes for patients with IES were comparable to those measured in patients without vestibular schwannoma (VS). Only patients who had radiation treatment before receiving their implant had elevated threshold requirements for CI programming compared with CI recipients without VS. CONCLUSION: Cochlear implantation without tumor removal is an effective option for treating NF2 patients with IES. The presence of an intracochlear tumor did not have an impact on CI performance or programming requirements compared with patients without VS; however, previous treatment with radiation may be related to elevated current requirements in the CI settings.


Assuntos
Implante Auditivo de Tronco Encefálico , Implante Coclear , Implantes Cocleares , Neurofibromatose 2 , Humanos , Neurofibromatose 2/complicações , Neurofibromatose 2/cirurgia , Estudos Retrospectivos