Nihon Hoshasen Gijutsu Gakkai Zasshi ; 76(10): 1009-1016, 2020.
Artigo em Japonês | MEDLINE | ID: mdl-33087646

RESUMO

PURPOSE: The purpose of this paper was to determine the optimal imaging conditions for four-dimensional cone-beam computed tomography (4D-CBCT) using an X-ray tube and a flat-panel detector mounted on a radiotherapy device. METHODS: The optimal imaging conditions were examined by changing the gantry speed (GS) parameter that affected the exposure time. Exposed dose during imaging and image quality of moving phantom were compared between examined conditions. RESULTS: The weighted computed tomography dose index (CTDIW) decreased linearly with increasing GS. However, when GS was 180°/min or faster, the image quality degraded, and errors of 1 mm or more were observed regarding the size of mock tumor in the moving phantom. The accuracy of automatic image matching was within 0.1 mm when GS of 120°/min or slower was chosen. CONCLUSION: From the results of this study, we concluded that GS of 120°/min is the optimum imaging condition. Under this imaging condition, the exposure time and CTDIW can be reduced by about 50% without compromising the accuracy of image registration, compared to the conventional GS of 70°/min. In addition, it has been clarified that there is an event that image reconstruction is not performed correctly due to the influence of phantom artifacts without depending on GS.


Assuntos
Tomografia Computadorizada de Feixe Cônico , Tomografia Computadorizada Quadridimensional , Artefatos , Processamento de Imagem Assistida por Computador , Imagens de Fantasmas
2.
Artigo em Japonês | MEDLINE | ID: mdl-23358333

RESUMO

Recently electronic portal image devices (EPIDs) have been widely used for quality assurance and dose verification. However there are no reports describing EPID dosimetry for Elekta volumetric modulated arc therapy (VMAT). We have investigated EPID dosimetry during VMAT delivery using a commercial software EPIDose with an Elekta Synergy linac. Dose rate dependence and the linac system sag during gantry rotation were measured. Gamma indices were calculated between measured doses using an EPID and calculation made by a treatment planning system for prostate VMAT test plans. The results were also compared to gamma indices using films and a two-dimensional detector array, MapCHECK2. The pass rates of the gamma analysis with a criterion of 3% and 2 mm for the three methods were over 96% with good consistency. Our results have showed that EPID dosimetry is feasible for Elekta VMAT.


Assuntos
Radiometria/métodos , Radioterapia de Intensidade Modulada , Garantia da Qualidade dos Cuidados de Saúde , Radiometria/instrumentação , Radioterapia de Intensidade Modulada/métodos
3.
Artigo em Japonês | MEDLINE | ID: mdl-23171770

RESUMO

The tolerance of the Backup diaphragm (Backup JAW) setting in Elekta linac was specified as 2 mm according to the AAPM TG-142 report. However, the tolerance and the quality assurance procedure for volumetric modulated arc therapy (VMAT) was not provided. This paper describes positional accuracy and quality assurance procedure of the Backup JAWs required for VMAT. It was found that a gap-width error of the Backup JAW by a sliding window test needed to be less than 1.5 mm for prostate VMAT delivery. It was also confirmed that the gap-widths had been maintained with an error of 0.2 mm during the past one year.


Assuntos
Neoplasias da Próstata/radioterapia , Planejamento da Radioterapia Assistida por Computador/instrumentação , Planejamento da Radioterapia Assistida por Computador/métodos , Erros de Configuração em Radioterapia/prevenção & controle , Radioterapia de Intensidade Modulada/instrumentação , Radioterapia de Intensidade Modulada/métodos , Humanos , Masculino , Aceleradores de Partículas , Controle de Qualidade , Dosagem Radioterapêutica
4.
Artigo em Japonês | MEDLINE | ID: mdl-23089835

RESUMO

Stereotactic radiotherapy requires a quality assurance (QA) program that ensures the mechanical accuracy of a radiation field center. We have proposed a QA method for achieving the above requirement by conducting the Winston Lutz test using an electronic portal image device (EPID). An action limit was defined as three times the standard deviation. Then, the action limits for mean deviations of the radiation field center during collimator rotation, gantry rotation, and couch rotation in clockwise and counterclockwise resulted in 0.11 mm, 0.52 mm, 0.37 mm, and 0.41 mm respectively. Two years after the QA program was launched, the mean deviation of the radiation field center during gantry rotation exceeded the above action limit. Consequently, a mechanical adjustment for the gantry was performed, thereby restoring the accuracy of the radiation field center. A field center shift of 0.5 mm was also observed after a micro multi-leaf collimator was unmounted.


Assuntos
Garantia da Qualidade dos Cuidados de Saúde/métodos , Radiocirurgia/normas , Radiocirurgia/métodos