RESUMO
We surveyed the actual conditions of mammography with regard to image quality and radiation dose at 44 facilities in Kagoshima prefecture in 1999. In April 2004, guidelines for mammography newly included the standard of digital mammography. From September to October 2005, the survey was conducted at 48 facilities, and the results of the survey were compared with that in 1999. We visited 44 of the 48 facilities, and visually evaluated the image quality of mammograms for RMI156 and clinical mammograms. In addition, we measured average mammary gland dose at each facility. The number of the mammography device that satisfied the specified guideline criterion was larger than that in 1999. Image quality for the RMI156 mammograms improved. However, the results of the present survey revealed several problems. First, the number of facilities that had quality control instruments for mammography are few. Second, radiological technologists, medical doctors, and nurses did not share knowledge or information regarding mammography. Finally, there were differences in devices and image quality for mammography among the facilities. We achieved an understanding of the actual conditions of mammography in Kagoshima prefecture by visiting many facilities, evaluating image quality, and communicating with many staff members. Our results may be useful for the development of mammography examinations.