Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Phys Med ; 32(1): 266-71, 2016 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-26703446

RESUMO

The demand for greater accuracy of intensity-modulated radiotherapy (IMRT) has driven the development of more advanced verification systems for image-guided radiotherapy (IGRT). The purpose of this study is to investigate setup discrepancies measured between an orthogonal X-ray guidance system (XGS-10) and cone-beam computed tomography (CBCT) of Varian in the IMRT of patients with nasopharyngeal cancer (NPC). The setup errors measured by XGS-10 and CBCT at the treatment unit with respect to the planning CTs were recorded for 30 patients with NPC. The differences in residual setup errors between XGS-10 system and CBCT were computed and quantitatively analyzed. The time of image acquisition and image registration was recorded. The radiation doses delivered by CBCT and XGS-10 were measured using PTW0.6CC ionization chambers and a water phantom. The differences between setup errors measured by the XGS-10 system and CBCT were generally <1.5 mm for translations, indicating a reasonably good agreement between the two systems for patients with NPC in the translation directions of A-P (P = 0.856), L-R (P = 0.856) and S-I (P = 0.765). Moreover, compared with CBCT, XGS-10 took much shorter image acquisition and registration time (P <0.001) and delivered only a small fraction of extra radiation dose to the patients (P <0.001). These results indicate that XGS-10 offers high localization accuracy similar to CBCT and additional benefits including prompt imaging process, low imaging radiation exposure, real time monitoring, which therefore represents a potential attractive alternative to CBCT for clinical use.


Assuntos
Tomografia Computadorizada de Feixe Cônico/métodos , Neoplasias Nasofaríngeas/radioterapia , Radiografia/métodos , Radioterapia Guiada por Imagem/métodos , Algoritmos , Carcinoma , Simulação por Computador , Desenho de Equipamento , Humanos , Processamento de Imagem Assistida por Computador , Carcinoma Nasofaríngeo , Imagens de Fantasmas , Doses de Radiação , Radiometria , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada , Reprodutibilidade dos Testes , Estudos Retrospectivos , Água , Raios X
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA