1.
Int J Radiat Oncol Biol Phys ; 119(3): 737-749, 2024 Jul 01.
Artigo em Inglês | MEDLINE | ID: mdl-38110104

RESUMO

PURPOSE: The highly heterogeneous dose delivery of spatially fractionated radiation therapy (SFRT) is a profound departure from standard radiation planning and reporting approaches. Early SFRT studies have shown excellent clinical outcomes. However, prospective multi-institutional clinical trials of SFRT are still lacking. This NRG Oncology/American Association of Physicists in Medicine working group consensus aimed to develop recommendations on dosimetric planning, delivery, and SFRT dose reporting to address this current obstacle toward the design of SFRT clinical trials. METHODS AND MATERIALS: Working groups consisting of radiation oncologists, radiobiologists, and medical physicists with expertise in SFRT were formed in NRG Oncology and the American Association of Physicists in Medicine to investigate the needs and barriers in SFRT clinical trials. RESULTS: Upon reviewing the SFRT technologies and methods, this group identified challenges in several areas, including the availability of SFRT, the lack of treatment planning system support for SFRT, the lack of guidance in the physics and dosimetry of SFRT, the approximated radiobiological modeling of SFRT, and the prescription and combination of SFRT with conventional radiation therapy. CONCLUSIONS: Recognizing these challenges, the group further recommended several areas of improvement for the application of SFRT in cancer treatment, including the creation of clinical practice guidance documents, the improvement of treatment planning system support, the generation of treatment planning and dosimetric index reporting templates, and the development of better radiobiological models through preclinical studies and through conducting multi-institution clinical trials.


Assuntos
Ensaios Clínicos como Assunto , Fracionamento da Dose de Radiação , Planejamento da Radioterapia Assistida por Computador , Humanos , Planejamento da Radioterapia Assistida por Computador/métodos , Planejamento da Radioterapia Assistida por Computador/normas , Estudos Prospectivos , Neoplasias/radioterapia , Radioterapia (Especialidade)/normas , Estudos Multicêntricos como Assunto , Radiobiologia , Consenso
2.
Radiother Oncol ; 182: 109577, 2023 05.
Artigo em Inglês | MEDLINE | ID: mdl-36841341

RESUMO

AIM OF THE STUDY: To elucidate the important factors and their interplay that drive performance on IMRT phantoms from the Imaging and Radiation Oncology Core (IROC). METHODS: IROC's IMRT head and neck phantom contains two targets and an organ at risk. Point and 2D dose are measured by TLDs and film, respectively. 1,542 irradiations between 2012-2020 were retrospectively analyzed based on output parameters, complexity metrics, and treatment parameters. Univariate analysis compared parameters based on pass/fail, and random forest modeling was used to predict output parameters and determine the underlying importance of the variables. RESULTS: The average phantom pass rate was 92% and has not significantly improved over time. The step-and-shoot irradiation technique had significantly lower pass rates that significantly affected other treatment parameters' pass rates. The complexity of plans has significantly increased with time, and all aperture-based complexity metrics (except MCS) were associated with the probability of failure. Random forest-based prediction of failure had an accuracy of 98% on held-out test data not used in model training. While complexity metrics were the most important contributors, the specific metric depended on the set of treatment parameters used during the irradiation. CONCLUSION: With the prevalence of errors in radiotherapy, understanding which parameters affect treatment delivery is vital to improve patient treatment. Complexity metrics were strongly predictive of irradiation failure; however, they are dependent on the specific treatment parameters. In addition, the use of one complexity metric is insufficient to monitor all aspects of the treatment plan.


Assuntos
Radioterapia (Especialidade) , Radioterapia de Intensidade Modulada , Humanos , Radioterapia de Intensidade Modulada/métodos , Estudos Retrospectivos , Planejamento da Radioterapia Assistida por Computador/métodos , Imagens de Fantasmas , Dosagem Radioterapêutica , Aprendizado de Máquina
3.
Neurooncol Adv ; 4(1): vdac058, 2022.
Artigo em Inglês | MEDLINE | ID: mdl-35664554

RESUMO

Background: Stereotactic radiosurgery (SRS) is a common treatment for intracranial lesions. This work explores the state of SRS treatment delivery to characterize current treatment accuracy based on treatment parameters. Methods: NCI clinical trials involving SRS rely on an end-to-end treatment delivery on a patient surrogate (credentialing phantom) from the Imaging and Radiation Oncology Core (IROC) to test their treatment accuracy. The results of 1072 SRS phantom irradiations between 2012 and 2020 were retrospectively analyzed. Univariate analysis and random forest models were used to associate irradiation conditions with phantom performance. The following categories were evaluated in terms of how they predicted outcomes: year of irradiation, TPS algorithm, machine model, energy, and delivered field size. Results: Overall, only 84.6% of irradiations have met the IROC/NCI acceptability criteria. Pass rate has remained constant over time, while dose calculation accuracy has slightly improved. Dose calculation algorithm (P < .001), collimator (P = .024), and field size (P < .001) were statistically significant predictors of pass/fail. Specifically, pencil beam algorithms and cone collimators were more likely to be associated with failing phantom results. Random forest modeling identified the size of the field as the most important factor for passing or failing followed by algorithm. Conclusion: Constant throughout this retrospective study, approximately 15% of institutions fail to meet IROC/NCI standards for SRS treatment. In current clinical practice, this is particularly associated with smaller fields that yielded less accurate results. There is ongoing need to improve small field dosimetry, beam modeling, and QA to ensure high treatment quality, patient safety, and optimal clinical trials.

4.
Int J Radiat Oncol Biol Phys ; 111(5): 1155-1164, 2021 12 01.
Artigo em Inglês | MEDLINE | ID: mdl-34352289

RESUMO

PURPOSE: The aim of this study was to examine current practice patterns in pediatric total body irradiation (TBI) techniques among COG member institutions. METHODS AND MATERIALS: Between November 2019 and February 2020, a questionnaire containing 52 questions related to the technical aspects of TBI was sent to medical physicists at 152 COG institutions. The questions were designed to obtain technical information on commonly used TBI treatment techniques. Another set of 9 questions related to the clinical management of patients undergoing TBI was sent to 152 COG member radiation oncologists at the same institutions. RESULTS: Twelve institutions were excluded because TBI was not performed in their institutions. A total of 88 physicists from 88 institutions (63% response rate) and 96 radiation oncologists from 96 institutions (69% response rate) responded. The anterior-posterior/posterior-anterior (AP/PA) technique was the most common technique reported (49 institutions [56%]); 44 institutions (50%) used the lateral technique, and 14 (16%) used volumetric modulated arc therapy or tomotherapy. Midplane dose rates of 6 to 15 cGy/min were most commonly used. The most common specification for lung dose was the midlung dose for both AP/PA techniques (71%) and lateral techniques (63%). Almost all physician responders agreed with the need to refine current TBI techniques, and 79% supported the investigation of new TBI techniques to further lower the lung dose. CONCLUSIONS: There was no consistency in the practice patterns, methods for dose measurement, and reporting of TBI doses among COG institutions. The lack of standardization precludes meaningful correlation between TBI doses and clinical outcomes including disease control and normal tissue toxicity. The COG radiation oncology discipline is currently undertaking several steps to standardize the practice and dose reporting of pediatric TBI using detailed questionnaires and phantom-based credentialing for all COG centers.


Assuntos
Radioterapia (Especialidade) , Radioterapia de Intensidade Modulada , Criança , Humanos , Pulmão , Inquéritos e Questionários , Irradiação Corporal Total
5.
Pediatr Blood Cancer ; 68 Suppl 2: e28609, 2021 05.
Artigo em Inglês | MEDLINE | ID: mdl-33818891

RESUMO

The Children's Oncology Group (COG) has a strong quality assurance (QA) program managed by the Imaging and Radiation Oncology Core (IROC). This program consists of credentialing centers and providing real-time management of each case for protocol compliant target definition and radiation delivery. In the International Society of Pediatric Oncology (SIOP), the lack of an available, reliable online data platform has been a challenge and the European Society for Paediatric Oncology (SIOPE) quality and excellence in radiotherapy and imaging for children and adolescents with cancer across Europe in clinical trials (QUARTET) program currently provides QA review for prospective clinical trials. The COG and SIOP are fully committed to a QA program that ensures uniform execution of protocol treatments and provides validity of the clinical data used for analysis.


Assuntos
Neoplasias/radioterapia , Garantia da Qualidade dos Cuidados de Saúde/normas , Radioterapia (Especialidade)/normas , Planejamento da Radioterapia Assistida por Computador/normas , Adolescente , Criança , Humanos
6.
Med Phys ; 45(5): 2337-2344, 2018 May.
Artigo em Inglês | MEDLINE | ID: mdl-29537634

RESUMO

PURPOSE: Reference dosimetry data can provide an independent second check of acquired values when commissioning or validating a treatment planning system (TPS). The Imaging and Radiation Oncology Core at Houston (IROC-Houston) has measured numerous linear accelerators throughout its existence. The results of those measurements are given here, comparing accelerators and the agreement of measurement versus institutional TPS calculations. METHODS: Data from IROC-Houston on-site reviews from 2000 through 2014 were analyzed for all Elekta accelerators, approximately 50. For each, consistent point dose measurements were conducted for several basic parameters in a water phantom, including percentage depth dose, output factors, small-field output factors, off-axis factors, and wedge factors. The results were compared by accelerator type independently for 6, 10, 15, and 18 MV. Distributions of the measurements for each parameter are given, providing the mean and standard deviation. Each accelerator's measurements were also compared to its corresponding TPS calculation from the institution to determine the level of agreement, as well as determining which dosimetric parameters were most often in error. RESULTS: Accelerators were grouped by head type and reference dosimetric values were compiled. No class of linac had better overall agreement with its TPS, but percentage depth dose and output factors commonly agreed well, while small-field output factors, off-axis factors, and wedge factors often disagreed substantially from their TPS calculations. CONCLUSION: Reference data has been collected and analyzed for numerous Elekta linacs, which provide an independent way for a physicist to double-check their own measurements to prevent gross treatment errors. In addition, treatment planning parameters more often in error have been highlighted, providing practical caution for physicists commissioning treatment planning systems for Elekta linacs.


Assuntos
Modelos Teóricos , Aceleradores de Partículas , Radiometria/instrumentação , Radiometria/normas , Bases de Dados Factuais , Neoplasias/radioterapia , Padrões de Referência
7.
Int J Radiat Oncol Biol Phys ; 100(4): 1057-1066, 2018 03 15.
Artigo em Inglês | MEDLINE | ID: mdl-29485047

RESUMO

A substantial barrier to the single- and multi-institutional aggregation of data to supporting clinical trials, practice quality improvement efforts, and development of big data analytics resource systems is the lack of standardized nomenclatures for expressing dosimetric data. To address this issue, the American Association of Physicists in Medicine (AAPM) Task Group 263 was charged with providing nomenclature guidelines and values in radiation oncology for use in clinical trials, data-pooling initiatives, population-based studies, and routine clinical care by standardizing: (1) structure names across image processing and treatment planning system platforms; (2) nomenclature for dosimetric data (eg, dose-volume histogram [DVH]-based metrics); (3) templates for clinical trial groups and users of an initial subset of software platforms to facilitate adoption of the standards; (4) formalism for nomenclature schema, which can accommodate the addition of other structures defined in the future. A multisociety, multidisciplinary, multinational group of 57 members representing stake holders ranging from large academic centers to community clinics and vendors was assembled, including physicists, physicians, dosimetrists, and vendors. The stakeholder groups represented in the membership included the AAPM, American Society for Radiation Oncology (ASTRO), NRG Oncology, European Society for Radiation Oncology (ESTRO), Radiation Therapy Oncology Group (RTOG), Children's Oncology Group (COG), Integrating Healthcare Enterprise in Radiation Oncology (IHE-RO), and Digital Imaging and Communications in Medicine working group (DICOM WG); A nomenclature system for target and organ at risk volumes and DVH nomenclature was developed and piloted to demonstrate viability across a range of clinics and within the framework of clinical trials. The final report was approved by AAPM in October 2017. The approval process included review by 8 AAPM committees, with additional review by ASTRO, European Society for Radiation Oncology (ESTRO), and American Association of Medical Dosimetrists (AAMD). This Executive Summary of the report highlights the key recommendations for clinical practice, research, and trials.


Assuntos
Radioterapia (Especialidade)/normas , Sociedades Científicas/normas , Terminologia como Assunto , Comitês Consultivos/organização & administração , Comitês Consultivos/normas , Ensaios Clínicos como Assunto , Humanos , Dosagem Radioterapêutica/normas , Planejamento da Radioterapia Assistida por Computador/normas , Padrões de Referência , Software/normas , Estados Unidos
8.
Med Phys ; 45(4): e53-e83, 2018 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-29443390

RESUMO

PURPOSE: Patient-specific IMRT QA measurements are important components of processes designed to identify discrepancies between calculated and delivered radiation doses. Discrepancy tolerance limits are neither well defined nor consistently applied across centers. The AAPM TG-218 report provides a comprehensive review aimed at improving the understanding and consistency of these processes as well as recommendations for methodologies and tolerance limits in patient-specific IMRT QA. METHODS: The performance of the dose difference/distance-to-agreement (DTA) and γ dose distribution comparison metrics are investigated. Measurement methods are reviewed and followed by a discussion of the pros and cons of each. Methodologies for absolute dose verification are discussed and new IMRT QA verification tools are presented. Literature on the expected or achievable agreement between measurements and calculations for different types of planning and delivery systems are reviewed and analyzed. Tests of vendor implementations of the γ verification algorithm employing benchmark cases are presented. RESULTS: Operational shortcomings that can reduce the γ tool accuracy and subsequent effectiveness for IMRT QA are described. Practical considerations including spatial resolution, normalization, dose threshold, and data interpretation are discussed. Published data on IMRT QA and the clinical experience of the group members are used to develop guidelines and recommendations on tolerance and action limits for IMRT QA. Steps to check failed IMRT QA plans are outlined. CONCLUSION: Recommendations on delivery methods, data interpretation, dose normalization, the use of γ analysis routines and choice of tolerance limits for IMRT QA are made with focus on detecting differences between calculated and measured doses via the use of robust analysis methods and an in-depth understanding of IMRT verification metrics. The recommendations are intended to improve the IMRT QA process and establish consistent, and comparable IMRT QA criteria among institutions.


Assuntos
Garantia da Qualidade dos Cuidados de Saúde/métodos , Radioterapia de Intensidade Modulada/efeitos adversos , Radioterapia de Intensidade Modulada/métodos , Segurança , Humanos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
9.
Phys Imaging Radiat Oncol ; 7: 39-44, 2018 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-31872085

RESUMO

BACKGROUND AND PURPOSE: Remote beam output audits, which independently measure an institution's machine calibration, are a common component of independent radiotherapy peer review. This work reviews the results and trends of these audit results across several organisations and geographical regions. MATERIALS AND METHODS: Beam output audit results from the Australian Clinical Dosimetry Services, International Atomic Energy Agency, Imaging and Radiation Oncology Core, and Radiation Dosimetry Services were evaluated from 2010 to the present. The rate of audit results outside a +/-5% tolerance was evaluated for photon and electron beams as a function of the year of irradiation and nominal beam energy. Additionally, examples of confirmed calibration errors were examined to provide guidance to clinical physicists and auditing bodies. RESULTS: Of the 210,167 audit results, 1323 (0.63%) were outside of tolerance. There was a clear trend of improved audit performance for more recent dates, and while all photon energies generally showed uniform rates of results out of tolerance, low (6 MeV) and high (≥18 MeV) energy electron beams showed significantly elevated rates. Twenty nine confirmed calibration errors were explored and attributed to a range of issues, such as equipment failures, errors in setup, and errors in performing the clinical reference calibration. Forty-two percent of these confirmed errors were detected during ongoing periodic monitoring, and not at the time of the first audit of the machine. CONCLUSIONS: Remote beam output audits have identified, and continue to identify, numerous and often substantial beam calibration errors.

10.
Int J Radiat Oncol Biol Phys ; 99(5): 1094-1100, 2017 12 01.
Artigo em Inglês | MEDLINE | ID: mdl-29029890

RESUMO

PURPOSE: To review the dosimetric, mechanical, and programmatic deficiencies most frequently observed during on-site visits of radiation therapy facilities by the Imaging and Radiation Oncology Core Quality Assurance Center in Houston (IROC Houston). METHODS AND MATERIALS: The findings of IROC Houston between 2000 and 2014, including 409 institutions and 1020 linear accelerators (linacs), were compiled. On-site evaluations by IROC Houston include verification of absolute calibration (tolerance of ±3%), relative dosimetric review (tolerances of ±2% between treatment planning system [TPS] calculation and measurement), mechanical evaluation (including multileaf collimator and kilovoltage-megavoltage isocenter evaluation against Task Group [TG]-142 tolerances), and general programmatic review (including institutional quality assurance program vs TG-40 and TG-142). RESULTS: An average of 3.1 deficiencies was identified at each institution visited, a number that has decreased slightly with time. The most common errors are tabulated and include TG-40/TG-142 compliance (82% of institutions were deficient), small field size output factors (59% of institutions had errors ≥3%), and wedge factors (33% of institutions had errors ≥3%). Dosimetric errors of ≥10%, including in beam calibration, were seen at many institutions. CONCLUSIONS: There is substantial room for improvement of both dosimetric and programmatic issues in radiation therapy, which should be a high priority for the medical physics community. Particularly relevant was suboptimal beam modeling in the TPS and a corresponding failure to detect these errors by not including TPS data in the linac quality assurance process.


Assuntos
Institutos de Câncer/normas , Garantia da Qualidade dos Cuidados de Saúde/normas , Radioterapia (Especialidade)/normas , Radioterapia/normas , Calibragem/normas , Institutos de Câncer/estatística & dados numéricos , Humanos , Radioterapia (Especialidade)/instrumentação , Radiometria/normas , Radioterapia/instrumentação , Dosagem Radioterapêutica/normas , Erros de Configuração em Radioterapia/estatística & dados numéricos , Fatores de Tempo
11.
Med Phys ; 43(12): 6491, 2016 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-27908168

RESUMO

PURPOSE: To analyze the most recent results of the Imaging and Radiation Oncology Core Houston Quality Assurance Center's (IROC-H) anthropomorphic head and neck (H&N) phantom to determine the nature of failing irradiations and the feasibility of altering credentialing criteria. METHODS: IROC-H's H&N phantom, used for intensity-modulated radiation therapy credentialing for National Cancer Institute-sponsored clinical trials, requires that an institution's treatment plan agrees within ±7% of measured thermoluminescent dosimeter (TLD) doses; it also requires that ≥85% of pixels pass ±4 mm distance to agreement (7%/4 mm gamma analysis for film). The authors re-evaluated 156 phantom irradiations (November 1, 2014-October 31, 2015) according to the following tighter criteria: (1) 5% TLD and 5%/4 mm, (2) 5% TLD and 5%/3 mm, (3) 4% TLD and 4%/4 mm, and (4) 3% TLD and 3%/3 mm. Failure rates were evaluated with respect to individual film and TLD performance by location in the phantom. Overall poor phantom results were characterized qualitatively as systematic errors (correct shape and position but wrong magnitude of dose), setup errors/positional shifts, global but nonsystematic errors, and errors affecting only a local region. RESULTS: The pass rate for these phantoms using current criteria was 90%. Substituting criteria 1-4 reduced the overall pass rate to 77%, 70%, 63%, and 37%, respectively. Statistical analyses indicated that the probability of noise-induced TLD failure, even at the 5% criterion, was <0.5%. Phantom failures were generally identified by TLD (≥66% failed TLD, whereas ≥55% failed film), with most failures occurring in the primary planning target volume (≥77% of cases). Results failing current criteria or criteria 1 were primarily diagnosed as systematic >58% of the time (11/16 and 21/36 cases, respectively), with a greater extent due to underdosing. Setup/positioning errors were seen in 11%-13% of all failing cases (2/16 and 4/36 cases, respectively). Local errors (8/36 cases) could only be demonstrated at criteria 1. Only three cases of global errors were identified in these analyses. For current criteria and criteria 1, irradiations that failed from film only were overwhelmingly associated with phantom shifts/setup errors (≥80% of cases). CONCLUSIONS: This study highlighted that the majority of phantom failures are the result of systematic dosimetric discrepancies between the treatment planning system and the delivered dose. Further work is necessary to diagnose and resolve such dosimetric inaccuracy. In addition, the authors found that 5% TLD and 5%/4 mm gamma criteria may be both practically and theoretically achievable as an alternative to current criteria.


Assuntos
Credenciamento , Cabeça/anatomia & histologia , Pescoço/anatomia & histologia , Imagens de Fantasmas/normas , Radioterapia de Intensidade Modulada/instrumentação , Estudos de Viabilidade , Humanos , Controle de Qualidade
12.
Int J Radiat Oncol Biol Phys ; 95(5): 1527-1534, 2016 08 01.
Artigo em Inglês | MEDLINE | ID: mdl-27315667

RESUMO

PURPOSE: To compare radiation machine measurement data collected by the Imaging and Radiation Oncology Core at Houston (IROC-H) with institutional treatment planning system (TPS) values, to identify parameters with large differences in agreement; the findings will help institutions focus their efforts to improve the accuracy of their TPS models. METHODS AND MATERIALS: Between 2000 and 2014, IROC-H visited more than 250 institutions and conducted independent measurements of machine dosimetric data points, including percentage depth dose, output factors, off-axis factors, multileaf collimator small fields, and wedge data. We compared these data with the institutional TPS values for the same points by energy, class, and parameter to identify differences and similarities using criteria involving both the medians and standard deviations for Varian linear accelerators. Distributions of differences between machine measurements and institutional TPS values were generated for basic dosimetric parameters. RESULTS: On average, intensity modulated radiation therapy-style and stereotactic body radiation therapy-style output factors and upper physical wedge output factors were the most problematic. Percentage depth dose, jaw output factors, and enhanced dynamic wedge output factors agreed best between the IROC-H measurements and the TPS values. Although small differences were shown between 2 common TPS systems, neither was superior to the other. Parameter agreement was constant over time from 2000 to 2014. CONCLUSIONS: Differences in basic dosimetric parameters between machine measurements and TPS values vary widely depending on the parameter, although agreement does not seem to vary by TPS and has not changed over time. Intensity modulated radiation therapy-style output factors, stereotactic body radiation therapy-style output factors, and upper physical wedge output factors had the largest disagreement and should be carefully modeled to ensure accuracy.


Assuntos
Algoritmos , Neoplasias/radioterapia , Radiometria/métodos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Conformacional/métodos , Radiometria/instrumentação , Planejamento da Radioterapia Assistida por Computador/instrumentação , Radioterapia Conformacional/instrumentação , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Texas
13.
Med Phys ; 43(5): 2374, 2016 May.
Artigo em Inglês | MEDLINE | ID: mdl-27147349

RESUMO

PURPOSE: Accurate data regarding linear accelerator (Linac) radiation characteristics are important for treatment planning system modeling as well as regular quality assurance of the machine. The Imaging and Radiation Oncology Core-Houston (IROC-H) has measured the dosimetric characteristics of numerous machines through their on-site dosimetry review protocols. Photon data are presented and can be used as a secondary check of acquired values, as a means to verify commissioning a new machine, or in preparation for an IROC-H site visit. METHODS: Photon data from IROC-H on-site reviews from 2000 to 2014 were compiled and analyzed. Specifically, data from approximately 500 Varian machines were analyzed. Each dataset consisted of point measurements of several dosimetric parameters at various locations in a water phantom to assess the percentage depth dose, jaw output factors, multileaf collimator small field output factors, off-axis factors, and wedge factors. The data were analyzed by energy and parameter, with similarly performing machine models being assimilated into classes. Common statistical metrics are presented for each machine class. Measurement data were compared against other reference data where applicable. RESULTS: Distributions of the parameter data were shown to be robust and derive from a student's t distribution. Based on statistical and clinical criteria, all machine models were able to be classified into two or three classes for each energy, except for 6 MV for which there were eight classes. Quantitative analysis of the measurements for 6, 10, 15, and 18 MV photon beams is presented for each parameter; supplementary material has also been made available which contains further statistical information. CONCLUSIONS: IROC-H has collected numerous data on Varian Linacs and the results of photon measurements from the past 15 years are presented. The data can be used as a comparison check of a physicist's acquired values. Acquired values that are well outside the expected distribution should be verified by the physicist to identify whether the measurements are valid. Comparison of values to this reference data provides a redundant check to help prevent gross dosimetric treatment errors.


Assuntos
Aceleradores de Partículas/instrumentação , Fótons , Radiometria/instrumentação , Aceleradores de Partículas/estatística & dados numéricos , Radiometria/estatística & dados numéricos , Valores de Referência , Água
14.
Int J Radiat Oncol Biol Phys ; 90(5): 1195-201, 2014 Dec 01.
Artigo em Inglês | MEDLINE | ID: mdl-25442044

RESUMO

PURPOSE: To determine whether in-house patient-specific intensity modulated radiation therapy quality assurance (IMRT QA) results predict Imaging and Radiation Oncology Core (IROC)-Houston phantom results. METHODS AND MATERIALS: IROC Houston's IMRT head and neck phantoms have been irradiated by numerous institutions as part of clinical trial credentialing. We retrospectively compared these phantom results with those of in-house IMRT QA (following the institution's clinical process) for 855 irradiations performed between 2003 and 2013. The sensitivity and specificity of IMRT QA to detect unacceptable or acceptable plans were determined relative to the IROC Houston phantom results. Additional analyses evaluated specific IMRT QA dosimeters and analysis methods. RESULTS: IMRT QA universally showed poor sensitivity relative to the head and neck phantom, that is, poor ability to predict a failing IROC Houston phantom result. Depending on how the IMRT QA results were interpreted, overall sensitivity ranged from 2% to 18%. For different IMRT QA methods, sensitivity ranged from 3% to 54%. Although the observed sensitivity was particularly poor at clinical thresholds (eg 3% dose difference or 90% of pixels passing gamma), receiver operator characteristic analysis indicated that no threshold showed good sensitivity and specificity for the devices evaluated. CONCLUSIONS: IMRT QA is not a reasonable replacement for a credentialing phantom. Moreover, the particularly poor agreement between IMRT QA and the IROC Houston phantoms highlights surprising inconsistency in the QA process.


Assuntos
Neoplasias de Cabeça e Pescoço/radioterapia , Imagens de Fantasmas/normas , Garantia da Qualidade dos Cuidados de Saúde/normas , Planejamento da Radioterapia Assistida por Computador/normas , Radioterapia de Intensidade Modulada/normas , Credenciamento , Curva ROC , Dosagem Radioterapêutica/normas , Planejamento da Radioterapia Assistida por Computador/instrumentação , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos , Estudos Retrospectivos , Sensibilidade e Especificidade
15.
Radiat Oncol ; 9: 123, 2014 May 29.
Artigo em Inglês | MEDLINE | ID: mdl-24885438

RESUMO

BACKGROUND AND PURPOSE: Intensity-modulated radiotherapy (IMRT) credentialing for a EORTC study was performed using an anthropomorphic head phantom from the Radiological Physics Center (RPC; RPC(PH)). Institutions were retrospectively requested to irradiate their institutional phantom (INST(PH)) using the same treatment plan in the framework of a Virtual Phantom Project (VPP) for IMRT credentialing. MATERIALS AND METHODS: CT data set of the institutional phantom and measured 2D dose matrices were requested from centers and sent to a dedicated secure EORTC uploader. Data from the RPC(PH) and INST(PH) were thereafter centrally analyzed and inter-compared by the QA team using commercially available software (RIT; ver.5.2; Colorado Springs, USA). RESULTS: Eighteen institutions participated to the VPP. The measurements of 6 (33%) institutions could not be analyzed centrally. All other centers passed both the VPP and the RPC ±7%/4 mm credentialing criteria. At the 5%/5 mm gamma criteria (90% of pixels passing), 11(92%) as compared to 12 (100%) centers pass the credentialing process with RPC(PH) and INST(PH) (p = 0.29), respectively. The corresponding pass rate for the 3%/3 mm gamma criteria (90% of pixels passing) was 2 (17%) and 9 (75%; p = 0.01), respectively. CONCLUSIONS: IMRT dosimetry gamma evaluations in a single plane for a H&N prospective trial using the INST(PH) measurements showed agreement at the gamma index criteria of ±5%/5 mm (90% of pixels passing) for a small number of VPP measurements. Using more stringent, criteria, the RPC(PH) and INST(PH) comparison showed disagreement. More data is warranted and urgently required within the framework of prospective studies.


Assuntos
Credenciamento , Cabeça/efeitos da radiação , Imagens de Fantasmas , Planejamento da Radioterapia Assistida por Computador , Radioterapia de Intensidade Modulada , Ensaios Clínicos como Assunto , Humanos , Estudos Multicêntricos como Assunto , Estudos Prospectivos , Garantia da Qualidade dos Cuidados de Saúde , Dosagem Radioterapêutica , Ensaios Clínicos Controlados Aleatórios como Assunto , Estudos Retrospectivos , Software
16.
J Appl Clin Med Phys ; 14(4): 4313, 2013 Jul 08.
Artigo em Inglês | MEDLINE | ID: mdl-23835394

RESUMO

An anthropomorphic head phantom, constructed from a water-equivalent plastic shell with only a spherical target, was modified to include a nonspherical target (pituitary) and an adjacent organ at risk (OAR) (optic chiasm), within 2 mm, simulating the anatomy encountered when treating acromegaly. The target and OAR spatial proximity provided a more realistic treatment planning and dose delivery exercise. A separate dosimetry insert contained two TLD for absolute dosimetry and radiochromic film, in the sagittal and coronal planes, for relative dosimetry. The prescription was 25 Gy to 90% of the GTV, with ≤ 10% of the OAR volume receiving ≥ 8 Gy for the phantom trial. The modified phantom was used to test the rigor of the treatment planning process and phantom reproducibility using a Gamma Knife, CyberKnife, and linear accelerator (linac)-based radiosurgery system. Delivery reproducibility was tested by repeating each irradiation three times. TLD results from three irradiations on a CyberKnife and Gamma Knife agreed with the calculated target dose to within ± 4% with a maximum coefficient of variation of ± 2.1%. Gamma analysis in the coronal and sagittal film planes showed an average passing rate of 99.4% and 99.5% using ± 5%/3 mm criteria, respectively. Results from the linac irradiation were within ± 6.2% for TLD with a coefficient of variation of ± 0.1%. Distance to agreement was calculated to be 1.2 mm and 1.3mm along the inferior and superior edges of the target in the sagittal film plane, and 1.2 mm for both superior and inferior edges in the coronal film plane. A modified, anatomically realistic SRS phantom was developed that provided a realistic clinical planning and delivery challenge that can be used to credential institutions wanting to participate in NCI-funded clinical trials.


Assuntos
Imagens de Fantasmas/normas , Radiocirurgia/normas , Acromegalia/cirurgia , Adenoma/cirurgia , Ensaios Clínicos como Assunto , Cabeça , Humanos , Pescoço , Quiasma Óptico/efeitos da radiação , Órgãos em Risco , Neoplasias Hipofisárias/cirurgia , Controle de Qualidade , Planejamento da Radioterapia Assistida por Computador/normas , Reprodutibilidade dos Testes , Dosimetria Termoluminescente
18.
Med Phys ; 40(5): 051710, 2013 May.
Artigo em Inglês | MEDLINE | ID: mdl-23635258

RESUMO

PURPOSE: The novel deterministic radiation transport algorithm, Acuros XB (AXB), has shown great potential for accurate heterogeneous dose calculation. However, the clinical impact between AXB and other currently used algorithms still needs to be elucidated for translation between these algorithms. The purpose of this study was to investigate the impact of AXB for heterogeneous dose calculation in lung cancer for intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT). METHODS: The thorax phantom from the Radiological Physics Center (RPC) was used for this study. IMRT and VMAT plans were created for the phantom in the Eclipse 11.0 treatment planning system. Each plan was delivered to the phantom three times using a Varian Clinac iX linear accelerator to ensure reproducibility. Thermoluminescent dosimeters (TLDs) and Gafchromic EBT2 film were placed inside the phantom to measure delivered doses. The measurements were compared with dose calculations from AXB 11.0.21 and the anisotropic analytical algorithm (AAA) 11.0.21. Two dose reporting modes of AXB, dose-to-medium in medium (Dm,m) and dose-to-water in medium (Dw,m), were studied. Point doses, dose profiles, and gamma analysis were used to quantify the agreement between measurements and calculations from both AXB and AAA. The computation times for AAA and AXB were also evaluated. RESULTS: For the RPC lung phantom, AAA and AXB dose predictions were found in good agreement to TLD and film measurements for both IMRT and VMAT plans. TLD dose predictions were within 0.4%-4.4% to AXB doses (both Dm,m and Dw,m); and within 2.5%-6.4% to AAA doses, respectively. For the film comparisons, the gamma indexes (± 3%∕3 mm criteria) were 94%, 97%, and 98% for AAA, AXB_Dm,m, and AXB_Dw,m, respectively. The differences between AXB and AAA in dose-volume histogram mean doses were within 2% in the planning target volume, lung, heart, and within 5% in the spinal cord. However, differences up to 8% between AXB and AAA were found at lung∕soft tissue interface regions for individual IMRT fields. AAA was found to be 5-6 times faster than AXB for IMRT, while AXB was 4-5 times faster than AAA for VMAT plan. CONCLUSIONS: AXB is satisfactorily accurate for the dose calculation in lung cancer for both IMRT and VMAT plans. The differences between AXB and AAA are generally small except in heterogeneous interface regions. AXB Dw,m and Dm,m calculations are similar inside the soft tissue and lung regions. AXB can benefit lung VMAT plans by both improving accuracy and reducing computation time.


Assuntos
Algoritmos , Neoplasias Pulmonares/radioterapia , Doses de Radiação , Planejamento da Radioterapia Assistida por Computador/métodos , Humanos , Imagens de Fantasmas , Radiometria , Dosagem Radioterapêutica , Radioterapia de Intensidade Modulada , Fatores de Tempo
19.
Med Phys ; 40(2): 022101, 2013 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-23387762

RESUMO

PURPOSE: This study was performed to report and analyze the results of the Radiological Physics Center's head and neck intensity-modulated radiation therapy (IMRT) phantom irradiations done by institutions seeking to be credentialed for participation in clinical trials using intensity modulated radiation therapy. METHODS: The Radiological Physics Center's anthropomorphic head and neck phantom was sent to institutions seeking to participate in multi-institutional clinical trials. The phantom contained two planning target volume (PTV) structures and an organ at risk (OAR). Thermoluminescent dosimeters (TLD) and film dosimeters were imbedded in the PTV. Institutions were asked to image, plan, and treat the phantom as they would treat a patient. The treatment plan should cover at least 95% of the primary PTV with 6.6 Gy and at least 95% of the secondary PTV with 5.4 Gy. The plan should limit the dose to the OAR to less than 4.5 Gy. The passing criteria were ±7% for the TLD in the PTVs and a distance to agreement of 4 mm in the high dose gradient area between the PTV and the OAR. Pass rates for different delivery types, treatment planning systems (TPS), linear accelerators, and linear accelerator-planning system combinations were compared. RESULTS: The phantom was irradiated 1139 times by 763 institutions from 2001 through 2011. 929 (81.6%) of the irradiations passed the criteria. 156 (13.7%) irradiations failed only the TLD criteria, 21 (1.8%) failed only the film criteria, and 33 (2.9%) failed both sets of criteria. Only 69% of the irradiations passed a narrowed TLD criterion of ±5%. Varian-Elipse and TomoTherapy-HiArt combinations had the highest pass rates, ranging from 90% to 93%. Varian-Pinnacle(3), Varian-XiO, Siemens-Pinnacle(3), and Elekta-Pinnacle(3) combinations had pass rates that ranged from 66% to 81%. CONCLUSIONS: The head and neck phantom is a useful credentialing tool for multi-institutional IMRT clinical trials. The most commonly represented linear accelerator-planning system combinations can all pass the phantom, though some combinations had higher passing percentages than others. Tightening the criteria would significantly reduce the number of institutions passing the credentialing criteria. Causes for failures include incorrect data entered into the TPS, inexact beam modeling, and software and hardware failures.


Assuntos
Cabeça/efeitos da radiação , Pescoço/efeitos da radiação , Imagens de Fantasmas , Radioterapia de Intensidade Modulada/instrumentação , Humanos , Órgãos em Risco/efeitos da radiação
20.
Int J Radiat Oncol Biol Phys ; 85(1): e95-100, 2013 Jan 01.
Artigo em Inglês | MEDLINE | ID: mdl-23237006

RESUMO

PURPOSE: To determine the impact of treatment planning algorithm on the accuracy of heterogeneous dose calculations in the Radiological Physics Center (RPC) thorax phantom. METHODS AND MATERIALS: We retrospectively analyzed the results of 304 irradiations of the RPC thorax phantom at 221 different institutions as part of credentialing for Radiation Therapy Oncology Group clinical trials; the irradiations were all done using 6-MV beams. Treatment plans included those for intensity-modulated radiation therapy (IMRT) as well as 3-dimensional conformal therapy (3D-CRT). Heterogeneous plans were developed using Monte Carlo (MC), convolution/superposition (CS), and the anisotropic analytic algorithm (AAA), as well as pencil beam (PB) algorithms. For each plan and delivery, the absolute dose measured in the center of a lung target was compared to the calculated dose, as was the planar dose in 3 orthogonal planes. The difference between measured and calculated dose was examined as a function of planning algorithm as well as use of IMRT. RESULTS: PB algorithms overestimated the dose delivered to the center of the target by 4.9% on average. Surprisingly, CS algorithms and AAA also showed a systematic overestimation of the dose to the center of the target, by 3.7% on average. In contrast, the MC algorithm dose calculations agreed with measurement within 0.6% on average. There was no difference observed between IMRT and 3D CRT calculation accuracy. CONCLUSION: Unexpectedly, advanced treatment planning systems (those using CS and AAA algorithms) overestimated the dose that was delivered to the lung target. This issue requires attention in terms of heterogeneity calculations and potentially in terms of clinical practice.


Assuntos
Algoritmos , Credenciamento , Neoplasias Pulmonares/radioterapia , Imagens de Fantasmas , Radioterapia (Especialidade) , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Institutos de Câncer , Neoplasias Pulmonares/diagnóstico por imagem , Método de Monte Carlo , Radiografia , Radioterapia Conformacional , Radioterapia de Intensidade Modulada , Estudos Retrospectivos , Tórax