RESUMO
ICD implantation is standard therapy for malignant ventricular arrhythmias. The advantage of dual and single coil defibrillator leads in the successful conversion of arrhythmias is unclear. This study compared the effectiveness of dual versus single coil defibrillation leads. The study was a prospective, multicenter, randomized study comparing a dual with a single coil defibrillation system as part of an ICD using an active pectoral electrode. Seventy-six patients (64 men, 12 women; age 61 +/- 11 years) were implanted with a dual (group 1, n = 38) or single coil lead system (group 2,n = 38). The patients represented a typical ICD cohort: 60% presented with ischemic cardiomyopathy as their primary cardiac disease, the mean left ventricular ejection fraction was 0.406 +/- 0.158. The primary tachyarrhythmia was monomorphic ventricular tachyarrhythmia in 52.6% patients and ventricular fibrillation in 38.4%. There was no significant difference in terms of P and R wave amplitudes, pacing thresholds, and lead impedance at implantation and follow-up in the two groups. There was similarly no difference in terms of defibrillation thresholds (DFT) at implantation. Patients in group 1 had an average DFT of 10.2 +/- 5.2 J compared to 10.3 +/- 4.1 J in Group 2, P = NS. This study demonstrates no significant advantage of a dual coil lead system over a single coil system in terms of lead values and defibrillation thresholds. This may have important bearing on the choice of lead systems when implanting ICDs.