1.
Med Phys ; 43(7): 4184, 2016 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-27370138

RESUMO

PURPOSE: Endoluminal ultrasound may serve as a minimally invasive option for delivering thermal ablation to pancreatic tumors adjacent to the stomach or duodenum. The objective of this study was to explore the basic feasibility of this treatment strategy through the design, characterization, and evaluation of proof-of-concept endoluminal ultrasound applicators capable of placement in the gastrointestinal (GI) lumen for volumetric pancreas ablation under MR guidance. METHODS: Two variants of the endoluminal applicator, each containing a distinct array of two independently powered transducers (10 × 10 mm 3.2 MHz planar; or 8 × 10 × 20 mm radius of curvature 3.3 MHz curvilinear geometries) at the distal end of a meter long flexible catheter assembly, were designed and fabricated. Transducers and circulatory water flow for acoustic coupling and luminal cooling were contained by a low-profile polyester balloon covering the transducer assembly fixture. Each applicator incorporated miniature spiral MR coils and mechanical features (guiding tips and hinges) to facilitate tracking and insertion through the GI tract under MRI guidance. Acoustic characterization of each device was performed using radiation force balance and hydrophone measurements. Device delivery into the upper GI tract, adjacent to the pancreas, and heating characteristics for treatment of pancreatic tissue were evaluated in MR-guided ex vivo and in vivo porcine experiments. MR guidance was utilized for anatomical target identification, tracking/positioning of the applicator, and MR temperature imaging (MRTI) for PRF-based multislice thermometry, implemented in the real-time RTHawk software environment. RESULTS: Force balance and hydrophone measurements indicated efficiencies of 48.8% and 47.8% and -3 dB intensity beam-widths of 3.2 and 1.2 mm for the planar and curvilinear transducers, respectively. Ex vivo studies on whole-porcine carcasses revealed capabilities of producing ablative temperature rise (ΔT > 15 °C) contours in pancreatic tissue 4-40 mm long and 4-28 mm wide for the planar transducer applicator (1-13 min sonication duration, ∼4 W/cm(2) applied acoustic intensity). Curvilinear transducers produced more selective heating, with a narrower ΔT > 15 °C contour length and width of up to 1-24 mm and 2-7 mm, respectively (1-7 min sonication duration, ∼4 W/cm(2) applied acoustic intensity). Active tracking of the miniature spiral coils was achieved using a Hadamard encoding tracking sequence, enabling real-time determination of each coil's coordinates and automated prescription of imaging planes for thermometry. In vivo MRTI-guided heating trials in three pigs demonstrated capability of ∼20 °C temperature elevation in pancreatic tissue at 2 cm depths from the applicator, with 5-7 W/cm(2) applied intensity and 6-16 min sonication duration. Dimensions of thermal lesions in the pancreas ranged from 12 to 28 mm, 3 to 10 mm, and 5 to 10 mm in length, width, and depth, respectively, as verified through histological analysis of tissue sections. Multiple-baseline reconstruction and respiratory-gated acquisition were demonstrated to be effective strategies in suppressing motion artifacts for clear evolution of temperature profiles during MRTI in the in vivo studies. CONCLUSIONS: This study demonstrates the technical feasibility of generating volumetric ablation in pancreatic tissue using endoluminal ultrasound applicators positioned in the stomach lumen. MR guidance facilitates target identification, device tracking/positioning, and treatment monitoring through real-time multislice PRF-based thermometry.


Assuntos
Ablação por Ultrassom Focalizado de Alta Intensidade/instrumentação , Neoplasias Pancreáticas/cirurgia , Animais , Catéteres , Desenho de Equipamento , Estudos de Viabilidade , Feminino , Trato Gastrointestinal/diagnóstico por imagem , Ablação por Ultrassom Focalizado de Alta Intensidade/métodos , Imagem por Ressonância Magnética Intervencionista/métodos , Neoplasias Pancreáticas/diagnóstico por imagem , Impressão Tridimensional , Software , Sus scrofa , Termografia/métodos
2.
Int J Hyperthermia ; 32(2): 97-111, 2016.
Artigo em Inglês | MEDLINE | ID: mdl-27097663

RESUMO

PURPOSE: The aim of this study is to investigate endoluminal ultrasound applicator configurations for volumetric thermal ablation and hyperthermia of pancreatic tumours using 3D acoustic and biothermal finite element models. MATERIALS AND METHODS: Parametric studies compared endoluminal heating performance for varying applicator transducer configurations (planar, curvilinear-focused, or radial-diverging), frequencies (1-5 MHz), and anatomical conditions. Patient-specific pancreatic head and body tumour models were used to evaluate feasibility of generating hyperthermia and thermal ablation using an applicator positioned in the duodenal or stomach lumen. Temperature and thermal dose were calculated to define ablation (> 240 EM(43 °C)) and moderate hyperthermia (40-45 °C) boundaries, and to assess sparing of sensitive tissues. Proportional-integral control was incorporated to regulate maximum temperature to 70-80 °C for ablation and 45 °C for hyperthermia in target regions. RESULTS: Parametric studies indicated that 1-3 MHz planar transducers are the most suitable for volumetric ablation, producing 5-8 cm(3) lesion volumes for a stationary 5-min sonication. Curvilinear-focused geometries produce more localised ablation to 20-45 mm depth from the GI tract and enhance thermal sparing (T(max) < 42 °C) of the luminal wall. Patient anatomy simulations show feasibility in ablating 60.1-92.9% of head/body tumour volumes (4.3-37.2 cm(3)) with dose < 15 EM(43 °C) in the luminal wall for 18-48 min treatment durations, using 1-3 applicator placements in GI lumen. For hyperthermia, planar and radial-diverging transducers could maintain up to 8 cm(3) and 15 cm(3) of tissue, respectively, between 40-45 °C for a single applicator placement. CONCLUSIONS: Modelling studies indicate the feasibility of endoluminal ultrasound for volumetric thermal ablation or hyperthermia treatment of pancreatic tumour tissue.


Assuntos
Modelos Teóricos , Neoplasias Pancreáticas/terapia , Modelagem Computacional Específica para o Paciente , Terapia por Ultrassom , Humanos , Transdutores
3.
Proc SPIE Int Soc Opt Eng ; 93262015 Feb 07.
Artigo em Inglês | MEDLINE | ID: mdl-26677314

RESUMO

An ultrasound applicator for endoluminal thermal therapy of pancreatic tumors has been introduced and evaluated through acoustic/biothermal simulations and ex vivo experimental investigations. Endoluminal therapeutic ultrasound constitutes a minimally invinvasive conformal therapy and is compatible with ultrasound or MR-based image guidance. The applicator would be placed in the stomach or duodenal lumen, and sonication would be performed through the luminal wall into the tumor, with concurrent water cooling of the wall tissue to prevent its thermal injury. A finite-element (FEM) 3D acoustic and biothermal model was implemented for theoretical analysis of the approach. Parametric studies over transducer geometries and frequencies revealed that operating frequencies within 1-3 MHz maximize penetration depth and lesion volume while sparing damage to the luminal wall. Patient-specific FEM models of pancreatic head tumors were generated and used to assess the feasibility of performing endoluminal ultrasound thermal ablation and hyperthermia of pancreatic tumors. Results indicated over 80% of the volume of small tumors (~2 cm diameter) within 35 mm of the duodenum could be safely ablated in under 30 minutes or elevated to hyperthermic temperatures at steady-state. Approximately 60% of a large tumor (~5 cm diameter) model could be safely ablated by considering multiple positions of the applicator along the length of the duodenum to increase coverage. Prototype applicators containing two 3.2 MHz planar transducers were fabricated and evaluated in ex vivo porcine carcass heating experiments under MR temperature imaging (MRTI) guidance. The applicator was positioned in the stomach adjacent to the pancreas, and sonications were performed for 10 min at 5 W/cm2 applied intensity. MRTI indicated over 40°C temperature rise in pancreatic tissue with heating penetration extending 3 cm from the luminal wall.

4.
Can J Urol ; 20(2): 6672-81, 2013 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-23587506

RESUMO

INTRODUCTION: There is currently a great deal of interest in the possible use of focal therapies for prostate cancer, since such treatments offer the prospect for control or cure of the primary disease with minimal side effects. Many forms of thermal therapy have been proposed for focal ablation of prostate cancer, including laser, high intensity ultrasound and cryotherapy. This review will demonstrate the important roles that magnetic resonance imaging (MRI) guidance can offer to such focal ablation, focusing on the use of high intensity ultrasonic applicators as an example of one promising technique. MATERIALS AND METHODS: Transurethral and interstitial high intensity ultrasonic applicators, designed specifically for ablation of prostate tissue were tested extensively in vivo in a canine model. The roles of MRI in positioning the devices, monitoring prostate ablation, and depicting ablated tissue were assessed using appropriate MRI sequences. RESULTS: MRI guidance provides a very effective tool for the positioning of ablative devices in the prostate, and thermal monitoring successfully predicted ablation of prostate tissue when a threshold of 52 ºC was achieved. Contrast enhanced MRI accurately depicted the distribution of ablated prostate tissue, which is resorbed at 30 days. CONCLUSIONS: Guidance of thermal therapies for focal ablation of prostate cancer will likely prove critically dependent on MRI functioning in four separate roles. Our studies indicate that in three roles: device positioning; thermal monitoring of prostate ablation; and depiction of ablated prostate tissue, MR techniques are highly accurate and likely to be of great benefit in focal prostate cancer ablation. A fourth critical role, identification of cancer within the gland for targeting of thermal therapy, is more problematic at present, but will likely become practical with further technological advances.


Assuntos
Técnicas de Ablação/métodos , Imageamento por Ressonância Magnética/métodos , Neoplasias da Próstata/patologia , Neoplasias da Próstata/cirurgia , Animais , Crioterapia , Modelos Animais de Doenças , Cães , Humanos , Terapia a Laser , Masculino , Terapia por Ultrassom
5.
Invest Radiol ; 48(6): 387-94, 2013 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-23462673

RESUMO

OBJECTIVES: The aims of this study were to evaluate in a canine model applicators designed for ablation of human benign prostatic hyperplasia (BPH) in vivo under magnetic resonance imaging (MRI) guidance, including magnetic resonance thermal imaging (MRTI), determine the ability of MRI techniques to visualize ablative changes in prostate, and evaluate the acute and longer term histologic appearances of prostate tissue ablated during these studies. MATERIALS AND METHODS: An MRI-compatible transurethral device incorporating a tubular transducer array with dual 120° sectors was used to ablate canine prostate tissue in vivo, in zones similar to regions of human BPH (enlarged transition zones). Magnetic resonance thermal imaging was used for monitoring of ablation in a 3-T environment, and postablation MRIs were performed to determine the visibility of ablated regions. Three canine prostates were ablated in acute studies, and 2 animals were rescanned before killing at 31 days postablation. Acute and chronic appearances of ablated prostate tissue were evaluated histologically and were correlated with the MRTI and postablation MRI scans. RESULTS: It was possible to ablate regions similar in size to enlarged transition zone in human BPH in 6 to 18 minutes. Regions of acute ablation showed a central "heat-fixed" region surrounded by a region of more obvious necrosis with complete disruption of tissue architecture. After 31 days, ablated regions demonstrated complete apparent resorption of ablated tissue with formation of cystic regions containing fluid. The inherent cooling of the urethra using the technique resulted in complete urethral preservation in all cases. CONCLUSIONS: Prostatic ablation of zones of size and shape corresponding to human BPH is possible using appropriate transurethral applicators using MRTI, and ablated tissue may be depicted clearly in contrast-enhanced magnetic resonance images. The ability accurately to monitor prostate tissue heating, the apparent resorption of ablated regions over 1 month, and the inherent urethral preservation suggest that the magnetic resonance-guided techniques described are highly promising for the in vivo ablation of symptomatic human BPH.


Assuntos
Ablação por Ultrassom Focalizado de Alta Intensidade/instrumentação , Imageamento por Ressonância Magnética/instrumentação , Prostatectomia/instrumentação , Hiperplasia Prostática/diagnóstico por imagem , Hiperplasia Prostática/patologia , Cirurgia Assistida por Computador/instrumentação , Transdutores , Animais , Cães , Desenho de Equipamento , Análise de Falha de Equipamento , Masculino , Radiografia , Resultado do Tratamento
6.
Proc SPIE Int Soc Opt Eng ; 8584: 85840V, 2013 Feb 26.
Artigo em Inglês | MEDLINE | ID: mdl-24860246

RESUMO

Image-guided thermal interventions have been proposed for potential palliative and curative treatments of pancreatic tumors. Catheter-based ultrasound devices offer the potential for temporal and 3D spatial control of the energy deposition profile. The objective of this study was to apply theoretical and experimental techniques to investigate the feasibility of endogastric, intraluminal and transgastric catheter-based ultrasound for MR guided thermal therapy of pancreatic tumors. The transgastric approach involves insertion of a catheter-based ultrasound applicator (array of 1.5 mm OD x 10 mm transducers, 360° or sectored 180°, ~7 MHz frequency, 13-14G cooling catheter) directly into the pancreas, either endoscopically or via image-guided percutaneous placement. An intraluminal applicator, of a more flexible but similar construct, was considered for endoscopic insertion directly into the pancreatic or biliary duct. An endoluminal approach was devised based on an ultrasound transducer assembly (tubular, planar, curvilinear) enclosed in a cooling balloon which is endoscopically positioned within the stomach or duodenum, adjacent to pancreatic targets from within the GI tract. A 3D acoustic bio-thermal model was implemented to calculate acoustic energy distributions and used a FEM solver to determine the transient temperature and thermal dose profiles in tissue during heating. These models were used to determine transducer parameters and delivery strategies and to study the feasibility of ablating 1-3 cm diameter tumors located 2-10 mm deep in the pancreas, while thermally sparing the stomach wall. Heterogeneous acoustic and thermal properties were incorporated, including approximations for tumor desmoplasia and dynamic changes during heating. A series of anatomic models based on imaging scans of representative patients were used to investigate the three approaches. Proof of concept (POC) endogastric and transgastric applicators were fabricated and experimentally evaluated in tissue mimicking phantoms, ex vivo tissue and in vivo canine model under multi-slice MR thermometry. RF micro-coils were evaluated to enable active catheter-tracking and prescription of thermometry slice positions. Interstitial and intraluminal ultrasound applicators could be used to ablate (t43>240 min) tumors measuring 2.3-3.4 cm in diameter when powered with 20-30 W/cm2 at 7 MHz for 5-10 min. Endoluminal applicators with planar and curvilinear transducers operating at 3-4 MHz could be used to treat tumors up to 20-25 mm deep from the stomach wall within 5 min. POC devices were fabricated and successfully integrated into the MRI environment with catheter tracking, real-time thermometry and closed-loop feedback control.

7.
J Comput Assist Tomogr ; 36(2): 253-6, 2012.
Artigo em Inglês | MEDLINE | ID: mdl-22446369

RESUMO

To more fully outline cervical and vaginal contours and distend the vagina, we have filled the vagina with sterile water-based gel before the magnetic resonance (MR) examination. The technique is similar to that used for defecating MR proctography and other MR examinations, but has not been well described for MR imaging of female pelvic cancer. We present our preliminary clinical experience, including a review of safety imaging characteristics and maintenance of the distention during the examination.


Assuntos
Meios de Contraste , Géis , Neoplasias dos Genitais Femininos/patologia , Imageamento por Ressonância Magnética/métodos , Administração Intravaginal , Meios de Contraste/administração & dosagem , Feminino , Géis/administração & dosagem , Humanos , Imageamento Tridimensional , Estadiamento de Neoplasias
8.
J Magn Reson Imaging ; 31(5): 1132-6, 2010 May.
Artigo em Inglês | MEDLINE | ID: mdl-20432348

RESUMO

PURPOSE: To create a reliable rat model with small renal cortical scars and evaluate the accuracy and sensitivity of dynamic contrast-enhanced MRI in detecting the kinds of lesions that are associated with reflux nephropathy. MATERIALS AND METHODS: In 16 rats, three unilateral renal cortical lesions were created using either electrocautery or pure alcohol with the contralateral kidney serving as control. MRI on a 1.5 Tesla GE Signa was performed 10-14 days after surgery. After bolus injection of 0.2 mM/Kg Gd-DTPA, sequential MRI acquisitions were performed using a 4-inch quadrature birdcage coil. Renal and scar volumes and pathology were compared after scanning and killing. RESULTS: Of the 48 points of injury, 40 (83%) in the 16 rats were detected grossly. Under microscopy, 36 injuries (75%) were detected on mid-kidney cross-sections. The average lesion was 4.2 mm(3) corresponding to 0.5% of the kidney volume. Using pathological findings as the gold standard, the sensitivity and specificity of scar detection using MRI was 69% and 93%, respectively. CONCLUSION: A rat model was created to demonstrate the sensitivity of dynamic contrast-enhanced MRI for detecting renal scars. Alcohol and electrocautery created reliable renal scars that were confirmed pathologically. MRI detected these lesions that averaged 4.2 mm(3) (0.5% total renal volume) with sensitivity and specificity of 69% and 93%, respectively.


Assuntos
Cicatriz/patologia , Modelos Animais de Doenças , Gadolínio DTPA , Aumento da Imagem/métodos , Nefropatias/patologia , Imageamento por Ressonância Magnética/métodos , Animais , Meios de Contraste , Feminino , Humanos , Masculino , Ratos , Ratos Sprague-Dawley , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
9.
Med Phys ; 35(5): 2081-93, 2008 May.
Artigo em Inglês | MEDLINE | ID: mdl-18561684

RESUMO

The purpose of this study was to explore the feasibility and performance of a multi-sectored tubular array transurethral ultrasound applicator for prostate thermal therapy, with potential to provide dynamic angular and length control of heating under MR guidance without mechanical movement of the applicator. Test configurations were fabricated, incorporating a linear array of two multi-sectored tubular transducers (7.8-8.4 MHz, 3 mm OD, 6 mm length), with three 120 degrees independent active sectors per tube. A flexible delivery catheter facilitated water cooling (100 ml min(-1)) within an expandable urethral balloon (35 mm long x 10 mm diameter). An integrated positioning hub allows for rotating and translating the transducer assembly within the urethral balloon for final targeting prior to therapy delivery. Rotational beam plots indicate approximately 90 degrees-100 degrees acoustic output patterns from each 120 degrees transducer sector, negligible coupling between sectors, and acoustic efficiencies between 41% and 53%. Experiments were performed within in vivo canine prostate (n = 3), with real-time MR temperature monitoring in either the axial or coronal planes to facilitate control of the heating profiles and provide thermal dosimetry for performance assessment. Gross inspection of serial sections of treated prostate, exposed to TTC (triphenyl tetrazolium chloride) tissue viability stain, allowed for direct assessment of the extent of thermal coagulation. These devices created large contiguous thermal lesions (defined by 52 degrees C maximum temperature, t43 = 240 min thermal dose contours, and TTC tissue sections) that extended radially from the applicator toward the border of the prostate (approximately15 mm) during a short power application (approximately 8-16 W per active sector, 8-15 min), with approximately 200 degrees or 360 degrees sector coagulation demonstrated depending upon the activation scheme. Analysis of transient temperature profiles indicated progression of lethal temperature and thermal dose contours initially centered on each sector that coalesced within approximately 5 min to produce uniform and contiguous zones of thermal destruction between sectors, with smooth outer boundaries and continued radial propagation in time. The dimension of the coagulation zone along the applicator was well-defined by positioning and active array length. Although not as precise as rotating planar and curvilinear devices currently under development for MR-guided procedures, advantages of these multi-sectored transurethral applicators include a flexible delivery catheter and that mechanical manipulation of the device using rotational motors is not required during therapy. This multi-sectored tubular array transurethral ultrasound technology has demonstrated potential for relatively fast and reasonably conformal targeting of prostate volumes suitable for the minimally invasive treatment of BPH and cancer under MR guidance, with further development warranted.


Assuntos
Imageamento por Ressonância Magnética/métodos , Neoplasias da Próstata/diagnóstico por imagem , Neoplasias da Próstata/terapia , Animais , Cães , Desenho de Equipamento , Humanos , Hipertermia Induzida , Masculino , Próstata/patologia , Hiperplasia Prostática/patologia , Radiometria , Temperatura , Transdutores , Terapia por Ultrassom , Ultrassonografia , Bexiga Urinária/patologia
10.
Artigo em Inglês | MEDLINE | ID: mdl-19163505

RESUMO

Catheter-based ultrasound applicators have been developed for delivering hyperthermia or high-temperature thermal ablation of cancer and benign disease of the prostate. These devices allow for control of heating along the length and angular expanse during therapy delivery. Four types of transurethral applicators were devised for thermal treatment of prostate combined with MR thermal monitoring: sectored tubular transducer devices with directional heating patterns and rotation; planar and curvilinear devices with narrow heating patterns and rotation; and multi-sectored tubular devices capable of dynamic angular control without applicator movement. Interstitial devices (2.4 mm OD) have been developed for percutaneous implantation with directional or dynamic angular control. In vivo experiments in canine prostate under MR temperature imaging were used to evaluate these devices and develop treatment delivery strategies. MR thermal imaging was used to monitor temperature and thermal dose in multiple slices through the target volume. Multi-sectored transurethral applicators can dynamically control the angular heating profile and target large regions of the gland in short treatment times without applicator manipulation. The sectored tubular, planar, and curvilinear transurethral devices produce directional coagulation zones, extending 15-20 mm radial distance to the outer prostate capsule. Sequential rotation under motor control and modulated dwell time can be used to tightly conform thermal ablation to selected regions. Interstitial implants with directional devices can be used to effectively ablate targeted regions of the gland while protecting the rectum. The MR derived 52 degrees C and lethal thermal dose contours (t43=240 min) effectively defined the extent of thermal damage and provided a means for real-time control of the applicators. Catheter-based ultrasound devices, combined with MR thermal monitoring, can produce relatively fast (5-40 min) and precise thermal ablation of prostate.


Assuntos
Ablação por Cateter/instrumentação , Hipertermia Induzida/instrumentação , Próstata/cirurgia , Neoplasias da Próstata/terapia , Terapia por Ultrassom/instrumentação , Animais , Temperatura Corporal , Cateterismo/instrumentação , Cães , Desenho de Equipamento , Temperatura Alta , Humanos , Hipertermia Induzida/métodos , Imageamento por Ressonância Magnética/métodos , Masculino , Neoplasias da Próstata/cirurgia , Temperatura
11.
AJR Am J Roentgenol ; 189(2): 314-22, 2007 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-17646456

RESUMO

OBJECTIVE: Our purpose was to evaluate the utility of CT urography performed using a split contrast bolus that yields synchronous nephrographic and excretory phase enhancement. MATERIALS AND METHODS: Five hundred consecutive patients referred for evaluation of possible urinary tract abnormalities (327 for painless hematuria) underwent CT urography with unenhanced scanning of the abdomen and pelvis and scanning during concurrent nephrographic and excretory phase enhancement produced by administration of a split contrast bolus. The enhanced abdomen scan was obtained with abdominal compression; the enhanced pelvis scan was obtained after release of compression. Findings from axial sections and coronal maximum intensity projections were correlated with clinical follow-up and, as available, with laboratory and other imaging studies including cystoscopy, ureteroscopy, urine cytology, surgery, and pathology. Follow-up management for each patient was determined by the clinical judgment of the referring physician. RESULTS: CT urography identified 100% of pathologically confirmed renal cell carcinomas (n = 10) and uroepithelial malignancies involving the renal collecting system or ureter (n = 8). An additional nine renal masses were identified for which no pathologic proof has yet been obtained, including eight subcentimeter solid renal masses and one multiloculated lesion. Fourteen of 19 confirmed cases of uroepithelial neoplasm involving the bladder were identified. CT urography yielded one false-positive for bladder tumor, two false-positives for ureteral tumor, and one patient with a bladder mass who refused further evaluation. CT urography yielded sensitivity and specificity of 100% and 99% and 74% and 99% and positive predictive value and negative predictive value of 80% and 100% and 93% and 99% for the renal collecting system and ureter and bladder, respectively. CT urography was ineffective in identifying 11 cases of noninfectious cystitis. CT urography also depicted numerous other congenital and acquired abnormalities of the urinary tract. CONCLUSION: Split-bolus MDCT urography detected all proven cases of tumors of the upper urinary tract, yielding high sensitivity and specificity. The split-bolus technique has the potential to reduce both radiation dose and the number of images generated by MDCT urography.


Assuntos
Tomografia Computadorizada por Raios X , Bexiga Urinária/anormalidades , Bexiga Urinária/diagnóstico por imagem , Urografia/métodos , Doenças Urológicas/diagnóstico por imagem , Adulto , Idoso , Idoso de 80 Anos ou mais , Meios de Contraste , Feminino , Humanos , Iohexol , Masculino , Valor Preditivo dos Testes , Interpretação de Imagem Radiográfica Assistida por Computador , Sensibilidade e Especificidade , Neoplasias da Bexiga Urinária/diagnóstico por imagem
12.
Med Phys ; 32(6): 1555-65, 2005 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-16013714

RESUMO

Thermal therapy offers a minimally invasive option for treating benign prostatic hyperplasia (BPH) and localized prostate cancer. In this study we investigated a transurethral ultrasound applicator design utilizing curvilinear, or slightly focused, transducers to heat prostatic tissue rapidly and controllably. The applicator was constructed with two independently powered transducer segments operating at 6.5 MHz and measuring 3.5 mm x 10 mm with a 15 mm radius of curvature across the short axis. The curvilinear applicator was characterized by acoustic efficiency measurements, acoustic beam plots, biothermal simulations of human prostate, ex vivo heating trials in bovine liver, and in vivo heating trials in canine prostate (n=3). Each transducer segment was found to emit a narrow acoustic beam (max width <3 mm), which extended the length of the transducer, with deeper penetration than previously developed planar or sectored tubular transurethral ultrasound applicators. Acoustic and biothermal simulations of human prostate demonstrated three treatment schemes for the curvilinear applicator: single shot (10 W, 60 s) schemes to generate narrow ablation zones (13 x 4 mm, 52 degrees C at the lesion boundary), incremental rotation (10 W, 10 degrees/45 s) to generate larger sector-shaped ablation zones (16 mm x 180 degrees sector), and rotation with variable sonication times (10 W, 10 degrees/15-90 s) to conform the ablation zone to a predefined boundary (9-17 mm x 180 degrees sector, 13 min total treatment time). During in vivo canine prostate experiments, guided by MR temperature imaging, single shot sonications (6 W/transducer, 2-3 min) with the curvilinear applicator ablated 20 degree sections of tissue to the prostate boundary (9-15 mm). Multiple adjacent sonications ("sweeping") ablated large sections of the prostate (180 degrees) by using the MR temperature imaging to adjust the power (4-6.4 W/transducer) and sonication time (30-180 s) at each 10 degrees rotation such that the periphery of the prostate reached 52 degrees C before the next rotation. The conclusion of this study was that the curvilinear applicator produces a narrow and penetrating ultrasound beam that, when combined with image guidance, can provide a precise technique for ablating target regions with a contoured outer boundary, such as the prostate capsule, by rotating in small steps while dynamically adjusting the net applied electrical power and sonication time at each position.


Assuntos
Hipertermia Induzida/métodos , Neoplasias da Próstata/terapia , Terapia por Ultrassom/métodos , Uretra/diagnóstico por imagem , Acústica , Animais , Cães , Temperatura Alta , Humanos , Masculino , Próstata/efeitos da radiação , Neoplasias da Próstata/diagnóstico por imagem , Sonicação , Temperatura , Fatores de Tempo , Ultrassom , Ultrassonografia , Uretra/patologia
13.
Med Phys ; 32(3): 733-43, 2005 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-15839345

RESUMO

The feasibility of MRI-guided interstitial ultrasound thermal therapy of the prostate was evaluated in an in vivo canine prostate model. MRI compatible, multielement interstitial ultrasound applicators were developed using 1.5 mm diameter cylindrical piezoceramic transducers (7 to 8 MHz) sectored to provide 180 degrees of angular directional heating. Two in vivo experiments were performed in canine prostate. The first using two interstitial ultrasound applicators, the second using three ultrasound applicators in conjunction with rectal and urethral cooling. In both experiments, the applicators were inserted transperineally into the prostate with the energy directed ventrally, away from the rectum. Electrical power levels of 5-17 W per element (approximately 1.6-5.4 W acoustic output power) were applied for heating periods of 18 and 48 min. Phase-sensitive gradient-echo MR imaging was used to monitor the thermal treatment in real-time on a 0.5 T interventional MRI system. Contrast-enhanced T1-weighted images and vital-stained serial tissue sections were obtained to assess thermal damage and correlate to real-time thermal contour plots and calculated thermal doses. Results from these studies indicated a large volume of ablated (nonstained) tissue within the prostate, extending 1.2 to 2.0 cm from the applicators to the periphery of the gland, with the dorsal margin of coagulation well-defined by the applicator placement and directionality. The shape of the lesions correlated well to the hypointense regions visible in the contrast-enhanced T1-weighted images, and were also in good agreement with the contours of the 52 degrees C threshold temperature and t43 > 240 min. This study demonstrates the feasibility of using directional interstitial ultrasound in conjunction with MRI thermal imaging to monitor and possibly control thermal coagulation within a targeted tissue volume while potentially protecting surrounding tissue, such as rectum, from thermal damage.


Assuntos
Interpretação de Imagem Assistida por Computador/métodos , Imageamento por Ressonância Magnética/métodos , Neoplasias da Próstata/diagnóstico , Neoplasias da Próstata/terapia , Terapia Assistida por Computador/métodos , Termografia/métodos , Terapia por Ultrassom/métodos , Animais , Cães , Estudos de Viabilidade , Masculino , Resultado do Tratamento , Terapia por Ultrassom/instrumentação
14.
Magn Reson Med ; 51(6): 1223-31, 2004 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-15170843

RESUMO

The proton resonance frequency (PRF) shift provides a means of measuring temperature changes during minimally invasive thermotherapy. However, conventional PRF thermometry relies on the subtraction of baseline images, which makes it sensitive to tissue motion and frequency drift during the course of treatment. In this study, a new method is presented that eliminates these problems by estimating the background phase from each acquired image phase. In this referenceless method, a polynomial is fit to the background phase outside the heated region in a weighted least-squares fit. Extrapolation of the polynomial to the heated region serves as the background phase estimate, which is then subtracted from the actual phase. The referenceless method is demonstrated on a phantom during laser heating, 0 degrees temperature rise images of in vivo human liver, interstitial laser ablation of porcine liver, and transurethral ultrasound ablation of canine prostate. A good correlation between temperature maps reconstructed with the referenceless and subtraction methods was found.


Assuntos
Imageamento por Ressonância Magnética/métodos , Termografia , Animais , Cães , Humanos , Processamento de Imagem Assistida por Computador , Técnicas In Vitro , Terapia a Laser , Fígado/anatomia & histologia , Fígado/cirurgia , Masculino , Imagens de Fantasmas , Próstata/cirurgia , Suínos
15.
Phys Med Biol ; 49(2): 189-204, 2004 Jan 21.
Artigo em Inglês | MEDLINE | ID: mdl-15083666

RESUMO

Transurethral ultrasound applicators with highly directional energy deposition and rotational control were investigated for precise treatment of benign prostatic hyperplasia (BPH) and adenocarcinoma of the prostate (CaP). Two types of catheter-based applicators were fabricated, using either 90 degrees sectored tubular (3.5 mm OD x 10 mm) or planar transducers (3.5 mm x 10 mm). They were constructed to be MRI compatible, minimally invasive and allow for manual rotation of the transducer array within a 10 mm cooling balloon. In vivo evaluations of the applicators were performed in canine prostates (n = 3) using MRI guidance (0.5 T interventional magnet). MR temperature imaging (MRTI) utilizing the proton resonance frequency shift method was used to acquire multiple-slice temperature overlays in real time for monitoring and guiding the thermal treatments. Post-treatment T1-weighted contrast-enhanced imaging and triphenyl tetrazolium chloride stained tissue sections were used to define regions of tissue coagulation. Single sonications with the 90 degrees tubular applicator (9-15 W, 12 min, 8 MHz) produced coagulated zones covering an 80 degrees wedge of the prostate extending from 1-2 mm outside the urethra to the outer boundary of the gland (16 mm radial coagulation). Single sonications with the planar applicator (15-20 W, 10 min, approximately 8 MHz) generated thermal lesions of approximately 30 degrees extending to the prostate boundary. Multiple sequential sonications (sweeping) of a planar applicator (12 W with eight rotations of 30 degrees each) demonstrated controllable coagulation of a 270 degrees contiguous section of the prostate extending to the capsule boundary. The feasibility of using highly directional transurethral ultrasound applicators with rotational capabilities to selectively coagulate regions of the prostate while monitoring and controlling the treatments with MRTI was demonstrated in this study.


Assuntos
Imageamento por Ressonância Magnética/métodos , Neoplasias da Próstata/patologia , Neoplasias da Próstata/radioterapia , Terapia por Ultrassom/instrumentação , Acústica , Adenocarcinoma/radioterapia , Animais , Cateterismo , Cães , Temperatura Alta , Masculino , Transdutores , Terapia por Ultrassom/métodos
16.
Conf Proc IEEE Eng Med Biol Soc ; 2004: 2496-9, 2004.
Artigo em Inglês | MEDLINE | ID: mdl-17270779

RESUMO

Thermal ablation has been investigated as a treatment for a variety of cancers. Heat treatments have not gained large-scale clinical acceptance due to inconsistencies in controlling heat deposition in vivo and the lack of precise temperature measurement. Interstitial ultrasound provide a good method of controlling the radial depth of a thermal lesion and the applicator designs evaluated in this study allow for dynamic angular control of the shape of the lesion. A trisectored internally water-cooled applicator (TriAD) and a rotating catheter water-cooled applicator (RIUS) angularly controlled thermal dose to a target area. Both devices were small in diameter (1.8 mm-2.4 mm), making them clinically feasible for minimally invasive treatment in device size-sensitive tissues. A biothermal model accounting for changes in acoustic attenuation and perfusion as a function of thermal dose was used to evaluate and predict applicator performance. The MR susceptibility artifact of the applicators was examined with MR temperature imaging (MRTI) sequences at 1.5 T and 0.5 T. Ex vivo experiments in turkey and beef muscle with realtime MRTI correlated well with results from the biothermal model. These results display the feasibility of thermally treating tumors with controllable interstitial ultrasound applicators under real-time MRTI and bracket the applicators' predicted performance in vivo.

SELEÇÃO DE REFERÊNCIAS