RESUMO
PURPOSE: To evaluate the reproducibility in the measurement of ocular biometric parameters using a new swept-source optical coherence tomographer and its comparability with an optical low coherence reflectometry biometer. DESIGN: An observational, descriptive, cross-sectional study. METHODS: 45 right eyes of 45 patients diagnosed with cataract were included. Three successive biometric measurements with Anterion and one with Lenstar LS900 were performed on each patient. The following variables were collected: axial length (AXL), anterior chamber depth (ACD), flat K (K1), steep K (K2), central corneal thickness (CCT), lens thickness (LT) and white-to-white distance (WTW). The intrasubject standard deviation (Sw) and the coefficient of Pearson "R" were calculated in order to assess the repeatability. The intraclass correlation coefficient (ICC) and the concordance correlation coefficient (CCC) were obtained to evaluate the comparability between devices. A Bland-Altman plot was performed for each variable. RESULTS: The coefficient of Pearson was excellent and statistically significant in the evaluation of the repeatability in all the variables. The highest values were 0.987 (AXL), 0.983 (CCT) and 0.942 (ACD). There were no statically significant differences between repeated measurements with Anterion in all the parameters. The ICC and CCC were excellent in the evaluation of AXL, CCT and ACD, and were also good in regard to K1, K2, LT and WTW. CONCLUSIONS: Performing biometry with the SS-OCT Anterion is a reliable and reproducible procedure, and it is comparable with the Lenstar LS900.