Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Electrophoresis ; 2024 Mar 15.
Artigo em Inglês | MEDLINE | ID: mdl-38488707

RESUMO

Recombinant adeno-associated virus (rAAV) is the leading platform of gene delivery for its long-lasting gene transformation and low immunogenicity. Characterization of the integrity and purity of the rAAV genome is critical to ensure clinical potency and safety. However, current rAAV genome characterization methods that can provide size assessment are either time-consuming or not easily accessible to general labs. Additionally, there is a lack of right reference standard for analyzing long single-stranded DNA (ssDNA) fragments. Here, we have developed an ssDNA assay on a microfluidic capillary electrophoresis platform using ssDNA reference standard. This assay provides size calling for ssDNA fragment, a detection sensitivity at ∼89 pg/µL (3 × 1010  GC/mL AAV) for 5.1 kb ssDNA fragment, and a turnaround time at ∼100 s per sample with a high throughput sample analyzing capability. Moreover, we have observed that the annealing of AAV ssDNA subsequent to its release from the capsid might introduce an additional double-stranded DNA (dsDNA) peak. This phenomenon is dependent on the sample processing workflow. To avoid the risk of mischaracterization, we recommend the use of dual-reference standards in combination with other orthogonal methods to have a comprehensive understanding of the rAAV genome size and integrity.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA