Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 683
Filtrar
1.
Card Electrophysiol Clin ; 16(2): 117-124, 2024 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-38749629

RESUMO

Transvenous laser lead extraction poses a risk of major complications (0.19%-1.8%), notably injury to the superior vena cava (SVC) in 0.19% to 0.96% of cases. Various factors contribute to SVC injury, which can be categorized as patient-related (such as female gender, low body mass index, diabetes, renal problems, anemia, and reduced ejection fraction), device-related (including the number, dwell time, and type of leads), or procedural-related (such as reason for extraction, venous obstructions, and bilateral lead placements).


Assuntos
Remoção de Dispositivo , Terapia a Laser , Veia Cava Superior , Humanos , Veia Cava Superior/lesões , Veia Cava Superior/cirurgia , Feminino , Remoção de Dispositivo/efeitos adversos , Masculino , Terapia a Laser/efeitos adversos , Pessoa de Meia-Idade , Idoso , Desfibriladores Implantáveis/efeitos adversos , Marca-Passo Artificial/efeitos adversos
2.
Med J Aust ; 220(10): 510-516, 2024 Jun 03.
Artigo em Inglês | MEDLINE | ID: mdl-38711337

RESUMO

OBJECTIVES: To quantify the rate of cardiac implantable electronic device (CIED)-related infections and to identify risk factors for such infections. DESIGN: Retrospective cohort study; analysis of linked hospital admissions and mortality data. SETTING, PARTICIPANTS: All adults who underwent CIED procedures in New South Wales between 1 January 2016 and 30 June 2021 (public hospitals) or 30 June 2020 (private hospitals). MAIN OUTCOME MEASURES: Proportions of patients hospitalised with CIED-related infections (identified by hospital record diagnosis codes); risk of CIED-related infection by patient, device, and procedural factors. RESULTS: Of 37 675 CIED procedures (23 194 men, 63.5%), 500 were followed by CIED-related infections (median follow-up, 24.9 months; interquartile range, 11.2-40.8 months), including 397 people (1.1%) within twelve months of their procedures, and 186 of 10 540 people (2.5%) at high risk of such infections (replacement or upgrade procedures; new cardiac resynchronisation therapy with defibrillator, CRT-D). The overall infection rate was 0.50 (95% confidence interval [CI], 0.45-0.54) per 1000 person-months; it was highest during the first month after the procedure (5.60 [95% CI, 4.89-6.42] per 1000 person-months). The risk of CIED-related infection was greater for people under 65 years of age than for those aged 65-74 years (adjusted hazard ratio [aHR], 1.71; 95% CI, 1.32-2.23), for people with CRT-D devices than for those with permanent pacemakers (aHR, 1.46; 95% CI, 1.02-2.08), for people who had previously undergone CIED procedures (two or more v none: aHR, 1.51; 95% CI, 1.02-2.25) or had CIED-related infections (aHR, 11.4; 95% CI, 8.34-15.7), or had undergone concomitant cardiac surgery (aHR, 1.62; 95% CI, 1.10-2.39), and for people with atrial fibrillation (aHR, 1.33; 95% CI, 1.11-1.60), chronic kidney disease (aHR, 1.54; 95% CI, 1.27-1.87), chronic obstructive pulmonary disease (aHR, 1.37; 95% CI, 1.10-1.69), or cardiomyopathy (aHR 1.60; 95% CI, 1.25-2.05). CONCLUSIONS: Knowledge of risk factors for CIED-related infections can help clinicians discuss them with their patients, identify people at particular risk, and inform decisions about device type, upgrades and replacements, and prophylactic interventions.


Assuntos
Desfibriladores Implantáveis , Infecções Relacionadas à Prótese , Humanos , Masculino , Estudos Retrospectivos , Feminino , Idoso , New South Wales/epidemiologia , Desfibriladores Implantáveis/efeitos adversos , Desfibriladores Implantáveis/estatística & dados numéricos , Infecções Relacionadas à Prótese/epidemiologia , Infecções Relacionadas à Prótese/etiologia , Pessoa de Meia-Idade , Fatores de Risco , Idoso de 80 Anos ou mais , Marca-Passo Artificial/efeitos adversos , Marca-Passo Artificial/estatística & dados numéricos , Adulto , Hospitalização/estatística & dados numéricos
3.
J Cardiovasc Electrophysiol ; 35(5): 1017-1025, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-38501386

RESUMO

Tricuspid regurgitation (TR) secondary to cardiac implantable electronic devices (CIEDs) has been well documented and is associated with worse cardiovascular outcomes. A variety of mechanisms have been proposed including lead-induced mechanical disruption of the tricuspid valvular or subvalvular apparatus and pacing-induced electrical dyssynchrony. Patient characteristics such as age, sex, baseline atrial fibrillation, and pre-existing TR have not been consistent predictors of CIED-induced TR. While two-dimensional echocardiography is helpful in assessing the severity of TR, three-dimensional echocardiography has significantly improved accuracy in identifying the etiology of TR and whether lead position contributes to TR. Three-dimensional echocardiography may therefore play a future role in optimizing lead positioning during implant to reduce the risk of CIED-induced TR. Optimal lead management strategies in addition to percutaneous interventions and surgery in alleviating TR are very important.


Assuntos
Desfibriladores Implantáveis , Marca-Passo Artificial , Insuficiência da Valva Tricúspide , Humanos , Insuficiência da Valva Tricúspide/fisiopatologia , Insuficiência da Valva Tricúspide/etiologia , Insuficiência da Valva Tricúspide/diagnóstico por imagem , Insuficiência da Valva Tricúspide/diagnóstico , Insuficiência da Valva Tricúspide/cirurgia , Desfibriladores Implantáveis/efeitos adversos , Marca-Passo Artificial/efeitos adversos , Valva Tricúspide/fisiopatologia , Valva Tricúspide/diagnóstico por imagem , Valva Tricúspide/cirurgia , Fatores de Risco , Resultado do Tratamento , Ecocardiografia Tridimensional
4.
Pacing Clin Electrophysiol ; 47(4): 496-502, 2024 04.
Artigo em Inglês | MEDLINE | ID: mdl-38462721

RESUMO

BACKGROUND: Inappropriate shock (IAS) caused by subcutaneous air entrapment (AE) in an early period after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation has been reported, however, no detailed data on air volume are available. We evaluated the subcutaneous air volume after implantation and its absorption rate one week after implantation. METHODS: Patients who underwent S-ICD implantation in our hospital received chest CT scans immediately after implantation and followed up 1 week later. The total subcutaneous air volume, air around the generator, the distal electrode, and the proximal electrode within 3 cm were calculated using a three-dimensional workstation. Fat areas at the level of the lower edge of the generator were also analyzed. RESULT: Fifteen patients received CT immediately after implantation. The mean age was 45.6 ± 17.9 (66.7% of men), and the mean body mass index was 24.3 ± 3.3. The three-incision technique was applied in seven patients and two-incision technique was in the latter eight patients. The mean total subcutaneous air volume was 18.54 ± 7.50 mL. Air volume around the generator, the distal electrode, and the proximal electrode were 11.05 ± 5.12, 0.72 ± 0.72, and 0.88 ± 0.87 mL, respectively. Twelve patients received a follow-up CT 1 week later. The mean total subcutaneous air was 0.25 ± 0.45 mL, showing a 98.7% absorption rate. CONCLUSION: Although subcutaneous air was observed in all patients after S-ICD implantation, most of the air was absorbed within 1 week, suggesting a low occurrence of AE-related IAS after a week postoperation.


Assuntos
Desfibriladores Implantáveis , Masculino , Humanos , Adulto , Pessoa de Meia-Idade , Desfibriladores Implantáveis/efeitos adversos , Cardioversão Elétrica , Implantação de Prótese/efeitos adversos , Implantação de Prótese/métodos , Tomografia Computadorizada por Raios X , Tomografia , Resultado do Tratamento
5.
J Cardiovasc Electrophysiol ; 35(4): 747-761, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38361241

RESUMO

INTRODUCTION: The implantation of a cardiac implantable electronic device (CIED) can have esthetic and psychological consequences on patients. We explore a heart team model for care coordination and discuss esthetic approaches for improved cosmetic outcomes in patients undergoing (CIED)-related procedures or de novo implantation. METHODS: Patients undergoing CIED surgery for approved indications between June 2015 and June 2022 were identified. Patients were included when surgical care was provided by a collaborative relationship between the primary electrophysiologist and the plastic surgeon. Patient demographics, details of the surgical procedure, information on breast implants, complications, and outcomes related to cosmesis were recorded. RESULTS: Twenty-two female patients were included in this study. The mean age was 50.2 ± 18.2 years. The mean follow-up duration was 2.2 ± 5.5 months. The top two indications for the procedure included CIED generator change (n = 9, 41%) and implantable cardioverter-defibrillator (ICD) implantation (n = 7, 32%). The most common reasons for involving plastic surgery in the procedure included surgery near breast implants (n = 10, 45%) and device displacement or discomfort (n = 8, 36%). CIED pocket position was prepectoral in 10 cases (45%), subpectoral in 11 patients (50%), and intramuscular in one patient (4.5%). The majority of the patients (20, 91%) had cosmetically acceptable results postprocedure. One patient (4.5%) had breast asymmetry on the CIED side, and another continued to have skin erosion over the CIED and leads. CONCLUSION: A heart team approach incorporating the expertize of cardiac electrophysiology and plastic surgery is essential for providing optimal care for patients with breast implants and patients requesting esthetic appeal.


Assuntos
Implantes de Mama , Procedimentos Cirúrgicos Cardíacos , Desfibriladores Implantáveis , Marca-Passo Artificial , Humanos , Feminino , Adulto , Pessoa de Meia-Idade , Idoso , Marca-Passo Artificial/efeitos adversos , Desfibriladores Implantáveis/efeitos adversos , Implantes de Mama/efeitos adversos , Procedimentos Cirúrgicos Cardíacos/efeitos adversos , Estudos Retrospectivos
6.
Pacing Clin Electrophysiol ; 47(4): 577-582, 2024 04.
Artigo em Inglês | MEDLINE | ID: mdl-38319639

RESUMO

BACKGROUND: The real-world data on the safety profile of transvenous lead extraction (TLE) for infected cardiac implantable electronic devices (CIED) among elderly patients is not well-established. This study aimed to evaluate the hospital outcomes between patients of different age groups who underwent TLE for infected CIED. METHOD: Using the Nationwide Readmissions Database, our study included patients aged ≥18 years who underwent TLE for infected CIED between 2017 and 2020. We divided the patients into four groups: Group A. Young (<50 years), Group B. Young intermediate (50-69 years old), Group C. Older intermediate (70-79 years old), and Group D. Octogenarian (≥80 years old). We then analyzed the in-hospital outcome and 30-day readmission between these age groups. RESULTS: A total of 10,928 patients who were admitted for TLE of infected CIED were included in this study: 982 (9.0%) patients in group A, 4,234 (38.7%) patients in group B, 3,204 (29.3%) patients in group C and 2,508 (23.0%) of patients in group D. Our study demonstrated that the risk of early mortality increased with older age (Group B vs. Group A: OR: 1.92, 95% CI: 1.19-3.09, p < .01; Group C vs. Group A: OR: 2.47, 95% CI: 1.51-4.04, p < .01; Group D vs. Group A: OR: 2.82, 95% CI: 1.69-4.72, p < .01). The risk of non-home discharge also increased in elderly groups (Group B vs. Group A: OR: 1.89; 95% CI: 1.52-2.36; p < .01; Group C vs. Group A: OR: 2.82; 95% CI 2.24-3.56; p < .01; Group D vs. Group A: OR: 4.16; 95% CI: 3.28-5.28; p < .01). There was no significant difference in hospitalization length and 30-day readmission between different age groups. Apart from a higher rate of open heart surgery in group A, the procedural complications were comparable between these age groups. CONCLUSION: Elderly patients had worse in-hospital outcomes in early mortality and non-home discharge following the TLE for infected CIED. There was no significant difference between elderly and non-elderly groups in prolonged hospital stay and 30-day readmission. Elderly patients did not have a higher risk of procedural complications.


Assuntos
Desfibriladores Implantáveis , Marca-Passo Artificial , Idoso , Idoso de 80 Anos ou mais , Humanos , Adolescente , Adulto , Pessoa de Meia-Idade , Desfibriladores Implantáveis/efeitos adversos , Marca-Passo Artificial/efeitos adversos , Remoção de Dispositivo/efeitos adversos , Fatores de Risco , Hospitais , Resultado do Tratamento , Estudos Retrospectivos
7.
J Cardiovasc Electrophysiol ; 35(3): 601-607, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38287171

RESUMO

The subcutaneous implantable cardioverter defibrillator (S-ICD) was developed as an alternative to the traditional transvenous implantable cardioverter defibrillator (TV-ICD), aiming to provide easier implantation, simplified detection algorithm of malignant ventricular arrhythmias and prevention from placing components in the cardiovascular system. The S-ICD is implanted subcutaneously or intramuscularly with the generator placed in the left midaxillary line and the lead tunneled subcutaneously in the left para-sternal region. Preimplant electrocardiogram screening is recommended to prevent implantation in patients at high risk of T wave over-sensing. Currently, the S-ICD is unsuitable for patients requiring pacing or cardiac resynchronization. Since the beginning, the S-ICD underwent extensive preclinical investigation until the first prospective multicentre trial demonstrating high efficacy and safety led to market release. While earlier studies focused on younger patients with higher ejection fraction, more recent studies showed favorable outcomes even in patients with comorbidities similar to those typically observed in patients receiving TV-ICD. The development of second and third generation devices has contributed to reduce inappropriate shocks and overcome previous limitations. The aim of this paper is to review the evidence in the literature over the past decade supporting S-ICD as a valid alternative to TV-ICD in terms of safety and efficacy, highlighting the improvements in technology, as well as outcomes.


Assuntos
Morte Súbita Cardíaca , Desfibriladores Implantáveis , Humanos , Morte Súbita Cardíaca/prevenção & controle , Estudos Prospectivos , Resultado do Tratamento , Arritmias Cardíacas/diagnóstico , Arritmias Cardíacas/terapia , Arritmias Cardíacas/etiologia , Desfibriladores Implantáveis/efeitos adversos , Estudos Multicêntricos como Assunto
8.
Eur Heart J ; 45(7): 538-548, 2024 Feb 16.
Artigo em Inglês | MEDLINE | ID: mdl-38195003

RESUMO

BACKGROUND AND AIMS: Implantable cardioverter-defibrillators (ICDs) are critical for preventing sudden cardiac death (SCD) in arrhythmogenic right ventricular cardiomyopathy (ARVC). This study aims to identify cross-continental differences in utilization of primary prevention ICDs and survival free from sustained ventricular arrhythmia (VA) in ARVC. METHODS: This was a retrospective analysis of ARVC patients without prior VA enrolled in clinical registries from 11 countries throughout Europe and North America. Patients were classified according to whether they received treatment in North America or Europe and were further stratified by baseline predicted VA risk into low- (<10%/5 years), intermediate- (10%-25%/5 years), and high-risk (>25%/5 years) groups. Differences in ICD implantation and survival free from sustained VA events (including appropriate ICD therapy) were assessed. RESULTS: One thousand ninety-eight patients were followed for a median of 5.1 years; 554 (50.5%) received a primary prevention ICD, and 286 (26.0%) experienced a first VA event. After adjusting for baseline risk factors, North Americans were more than three times as likely to receive ICDs {hazard ratio (HR) 3.1 [95% confidence interval (CI) 2.5, 3.8]} but had only mildly increased risk for incident sustained VA [HR 1.4 (95% CI 1.1, 1.8)]. North Americans without ICDs were at higher risk for incident sustained VA [HR 2.1 (95% CI 1.3, 3.4)] than Europeans. CONCLUSIONS: North American ARVC patients were substantially more likely than Europeans to receive primary prevention ICDs across all arrhythmic risk strata. A lower rate of ICD implantation in Europe was not associated with a higher rate of VA events in those without ICDs.


Assuntos
Displasia Arritmogênica Ventricular Direita , Desfibriladores Implantáveis , Humanos , Desfibriladores Implantáveis/efeitos adversos , Displasia Arritmogênica Ventricular Direita/complicações , Displasia Arritmogênica Ventricular Direita/epidemiologia , Displasia Arritmogênica Ventricular Direita/terapia , Estudos Retrospectivos , Arritmias Cardíacas/epidemiologia , Arritmias Cardíacas/terapia , Arritmias Cardíacas/etiologia , Morte Súbita Cardíaca/epidemiologia , Morte Súbita Cardíaca/prevenção & controle , Morte Súbita Cardíaca/etiologia , Fatores de Risco , América do Norte/epidemiologia , Europa (Continente)/epidemiologia
10.
Heart Rhythm ; 21(4): 410-418, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38246594

RESUMO

BACKGROUND: Outcome comparisons among subcutaneous implantable cardioverter-defibrillator (S-ICD) recipients with nonischemic cardiomyopathies are scarce. OBJECTIVE: The aim of this study was to evaluate differences in device-related outcomes among S-ICD recipients with different structural substrates. METHODS: Patients enrolled in the i-SUSI (International SUbcutaneouS Implantable cardioverter defibrillator registry) project were grouped according to the underlying substrate (ischemic vs nonischemic) and subgrouped into dilated cardiomyopathy, hypertrophic cardiomyopathy, Brugada syndrome (BrS), arrhythmogenic right ventricular cardiomyopathy (ARVC). The main outcome of our study was to compare the rates of appropriate and inappropriate shocks and device-related complications. RESULTS: Among 1698 patients, the most common underlying substrate was ischemic (31.7%), followed by dilated cardiomyopathy (20.5%), BrS (10.8%), hypertrophic cardiomyopathy (8.5%), and ARVC (4.4%). S-ICD for primary prevention was more common in the nonischemic cohort (70.9% vs 65.4%; P = .037). Over a median (interquartile range) follow-up of 26.5 (12.6-42.8) months, no differences were observed in appropriate shocks between ischemic and nonischemic patients (4.8%/y vs 3.9%/y; log-rank, P = .282). ARVC (9.0%/y; hazard ratio [HR] 2.492; P = .001) and BrS (1.8%/y; HR 0.396; P = .008) constituted the groups with the highest and lowest rates of appropriate shocks, respectively. Device-related complications did not differ between groups (ischemic: 6.4%/y vs nonischemic: 6.1%/y; log-rank, P = .666), nor among underlying substrates (log-rank, P = .089). Nonischemic patients experienced higher rates of inappropriate shocks than did ischemic S-ICD recipients (4.4%/y vs 3.0%/y; log-rank, P = .043), with patients with ARVC (9.9%/y; P = .001) having the highest risk, even after controlling for confounders (adjusted HR 2.243; confidence interval 1.338-4.267; P = .002). CONCLUSION: Most S-ICD recipients were primary prevention nonischemic cardiomyopathy patients. Among those, patients with ARVC tend to receive the most frequent appropriate and inappropriate shocks and patients with BrS the least frequent appropriate shocks.


Assuntos
Displasia Arritmogênica Ventricular Direita , Cardiomiopatias , Cardiomiopatia Dilatada , Desfibriladores Implantáveis , Humanos , Desfibriladores Implantáveis/efeitos adversos , Morte Súbita Cardíaca/epidemiologia , Morte Súbita Cardíaca/etiologia , Morte Súbita Cardíaca/prevenção & controle , Cardioversão Elétrica/efeitos adversos , Displasia Arritmogênica Ventricular Direita/complicações , Cardiomiopatia Dilatada/complicações , Cardiomiopatia Dilatada/terapia , Sistema de Registros , Resultado do Tratamento
12.
Pacing Clin Electrophysiol ; 47(2): 292-299, 2024 02.
Artigo em Inglês | MEDLINE | ID: mdl-38078545

RESUMO

BACKGROUND: Cardiac implantable electronic device (CIED) procedures are prone to complications. In our study, we investigated the effect of body mass index (BMI) on CIED-related complications. METHODS: 1676 patients who had undergone CIED surgery (de novo implantation, system upgrade, generator change, pocket revision or lead replacement) at two heart centers in Turkey and met the study criteria were included in our study. For analysis of primary and secondary endpoints, patients were classified as non-obese (BMI < 25 kg/m2 ), overweight (25 ≤ BMI < 30 kg/m2 ), and obese (BMI ≥ 30 kg/m2 ). The primary endpoint was accepted as cumulative events, including the composite of clinically significant hematoma (CSH), pericardial effusion or tamponade, pneumothorax, and infection related to the device system. Secondary outcomes included each component of cumulative events. RESULTS: The rate of cumulative events, defined as primary outcome, was higher in the obese patient group, and we found a significant difference between the groups (3.0%, 4.3%, 8.9%, p = .001). CSH and pneumothorax rates were significantly higher in the obese patient group (0.3%, 0.9%, 1.9%, p = .04; 1.0%, 1.4%, 3.3%, p = .04, respectively). According to our multivariate model analysis; gender (OR:1.882, 95%CI:1.156-3.064, p = .01), hypertension (OR:4.768, 95%CI:2.470-9.204, p < .001), BMI (OR:1.069, 95%CI:1.012-1.129, p = .01) were independent predictors of cumulative events rates. CONCLUSIONS: Periprocedural complications associated with CIED (especially hematoma and pneumothorax) are more common in the group with high BMI.


Assuntos
Desfibriladores Implantáveis , Marca-Passo Artificial , Derrame Pericárdico , Pneumotórax , Humanos , Desfibriladores Implantáveis/efeitos adversos , Índice de Massa Corporal , Pneumotórax/epidemiologia , Pneumotórax/etiologia , Obesidade/complicações , Hematoma/etiologia , Marca-Passo Artificial/efeitos adversos , Estudos Retrospectivos
13.
J Cardiovasc Transl Res ; 17(1): 91-101, 2024 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-37556036

RESUMO

Implantable cardioverter defibrillators (ICDs) reduce sudden cardiac death (SCD) when patients experience life-threatening ventricular arrhythmias (LTVA). However, current strategies determining ICD patient selection and risk stratification are inefficient. We used metabolomics to assess whether dysregulated metabolites are associated with LTVA and identify potential biomarkers. Baseline plasma samples were collected from 72 patients receiving ICDs. Over a median follow-up of 524.0 days (range 239.0-705.5), LTVA occurred in 23 (31.9%) patients (22 effective ICD treatments and 1 SCD). After confounding risk factors adjustment for age, smoking, secondary prevention, and creatine kinase MB, 23 metabolites were significantly associated with LTVA. Pathway analysis revealed LTVA associations with disrupted metabolism of glycine, serine, threonine, and branched chain amino acids. Pathway enrichment analysis identified a panel of 6 metabolites that potentially predicted LTVA, with an area under the receiver operating characteristic curve of 0.8. Future studies are necessary on biological mechanisms and potential clinical use.


Assuntos
Desfibriladores Implantáveis , Humanos , Desfibriladores Implantáveis/efeitos adversos , Arritmias Cardíacas/diagnóstico , Arritmias Cardíacas/terapia , Arritmias Cardíacas/complicações , Morte Súbita Cardíaca/etiologia , Morte Súbita Cardíaca/prevenção & controle , Resultado do Tratamento , Fatores de Risco
14.
Heart ; 110(3): 156-162, 2024 Jan 10.
Artigo em Inglês | MEDLINE | ID: mdl-37433658

RESUMO

Arrhythmogenic right ventricular cardiomyopathy (ARVC) is a disease characterised by fibrofatty replacement of the ventricular myocardium due to specific mutations, leading to ventricular arrhythmias and sudden cardiac death. Treating this condition can be challenging due to progressive fibrosis, phenotypic variations and small patient cohorts limiting the feasibility of conducting meaningful clinical trials. Although widely used, the evidence base for anti-arrhythmic drugs is limited. Beta-blockers are theoretically sound, yet their efficacy in reducing arrhythmic risk is not robust. Additionally, the impact of sotalol and amiodarone is inconsistent with studies reporting contradictory results. Emerging evidence suggests that combining flecainide and bisoprolol may be efficacious.Radiofrequency ablation has shown some potential in disrupting ventricular tachycardia circuits, with combined endo and epicardial ablation yielding better results which could be considered at the index procedure. In addition, stereotactic radiotherapy may be a future option that can decrease arrhythmias beyond simple scar formation by altering levels of Nav1.5 channels, Connexin 43 and Wnt signalling, potentially modifying myocardial fibrosis.Future therapies, such as adenoviruses and GSk3b modulation, are still in early-stage research. While implantable cardioverter-defibrillator implantation is a key intervention for reducing arrhythmic death, the risks of inappropriate shocks and device complications must be carefully considered.


Assuntos
Displasia Arritmogênica Ventricular Direita , Desfibriladores Implantáveis , Taquicardia Ventricular , Humanos , Displasia Arritmogênica Ventricular Direita/diagnóstico , Displasia Arritmogênica Ventricular Direita/terapia , Displasia Arritmogênica Ventricular Direita/complicações , Arritmias Cardíacas/complicações , Antiarrítmicos/uso terapêutico , Morte Súbita Cardíaca/etiologia , Morte Súbita Cardíaca/prevenção & controle , Sotalol , Taquicardia Ventricular/diagnóstico , Taquicardia Ventricular/etiologia , Taquicardia Ventricular/terapia , Desfibriladores Implantáveis/efeitos adversos
15.
Heart Rhythm ; 21(4): 419-426, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38142831

RESUMO

BACKGROUND: Left bundle branch pacing (LBBP) has been suggested as an alternative modality for biventricular pacing in cardiac resynchronization therapy (CRT)-eligible patients. As it provides stable R-wave sensing, LBBP has been recently used to provide sensing of ventricular arrhythmia in patients receiving implantable cardioverter-defibrillator (ICD) with CRT. OBJECTIVE: The aim of this study was to analyze the long-term safety and efficacy of the LBBP lead for appropriate detection of ventricular arrhythmia and delivery of antitachycardia pacing (ATP) in patients requiring defibrillator therapy with CRT. METHODS: CRT-eligible patients who underwent successful LBBP-optimized ICD and LBBP-optimized CRT with defibrillator were enrolled. The LBBP lead was connected to the right ventricular-P/S port after capping the IS-1 connector plug of the DF-1-ICD lead. LBBP-optimized ICD or LBBP-optimized CRT with defibrillator was decided on the basis of correction of conduction system disease. Documented arrhythmic episodes and therapy delivered were analyzed. RESULTS: Thirty patients were enrolled. The mean age was 59.7 ± 10.5 years. LBBP resulted in an increase in left ventricular ejection fraction from 29.9% ± 4.6% to 43.9% ± 11.2% (P < .0001). During a mean follow-up of 22.9 ± 12.5 months, 254 ventricular arrhythmic events were documented. Appropriate events (n = 225 [89%]) included nonsustained ventricular tachycardia (VT) (n = 212 episodes [94%]), VT (n = 8 [3.5%]), and ventricular fibrillation (n = 5 [2.5%]). ATP efficacy in terminating VT was 75%. Eleven percent of episodes (n = 29) were inappropriately detected because of T-wave oversensing. Inappropriate therapy (ATP) was delivered for 14 episodes (5.5%). Three patients (10%) had worsening of tricuspid regurgitation. CONCLUSION: Sensing from the LBBP lead for arrhythmia detection is safe as ∼90% of the episodes were detected appropriately. Future studies with a dedicated LBBP-defibrillator lead along with algorithms to avoid oversensing can help in combining defibrillation with conduction system pacing.


Assuntos
Terapia de Ressincronização Cardíaca , Desfibriladores Implantáveis , Humanos , Pessoa de Meia-Idade , Idoso , Projetos Piloto , Desfibriladores Implantáveis/efeitos adversos , Volume Sistólico , Função Ventricular Esquerda , Arritmias Cardíacas/diagnóstico , Arritmias Cardíacas/terapia , Doença do Sistema de Condução Cardíaco , Terapia de Ressincronização Cardíaca/efeitos adversos , Terapia de Ressincronização Cardíaca/métodos , Trifosfato de Adenosina , Resultado do Tratamento
16.
Intern Med J ; 53(11): 1956-1962, 2023 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-37929818

RESUMO

Secondary prevention implantable cardioverter defibrillators (ICDs) are indicated in young patients presenting with aborted sudden cardiac death (SCD) because of ventricular arrhythmias. Transvenous-ICDs (TV-ICDs) are effective, established therapies supported by evidence. The significant morbidity associated with transvenous leads led to the development of the newer subcutaneous-ICD (S-ICD). This review discusses the clinical considerations when selecting an ICD for the young patient presenting with out-of-hospital cardiac arrest. The major benefits of TV-ICDs are their ability to pace (antitachycardia pacing [ATP], bradycardia support and cardiac resynchronisation therapy [CRT]) and the robust evidence base supporting their use. Other benefits include a longer battery life. Significant complications associated with transvenous leads include pneumothorax and tamponade during insertion and infection and lead failure in the long term. Comparatively, S-ICDs, by virtue of having no intravascular leads, prevent these complications. S-ICDs have been associated with a higher incidence of inappropriate shocks. Patients with an indication for bradycardia pacing, CRT or ATP (documented ventricular tachycardia) are seen as unsuitable for a S-ICD. If venous access is unsuitable or undesirable, S-ICDs should be considered given the patient is appropriately screened. There is a need for further randomised controlled trials to directly compare the two devices. TV-ICDs are an effective therapy for preventing SCD limited by significant lead-related complications. S-ICDs are an important development hindered largely by an inability to pace. Young patients stand to gain the most from a S-ICD as the cumulative risk of lead-related complications is high. A clinical framework to aid decision-making is presented.


Assuntos
Desfibriladores Implantáveis , Parada Cardíaca , Humanos , Desfibriladores Implantáveis/efeitos adversos , Bradicardia , Arritmias Cardíacas , Parada Cardíaca/terapia , Trifosfato de Adenosina , Morte Súbita Cardíaca/prevenção & controle , Morte Súbita Cardíaca/epidemiologia , Resultado do Tratamento
17.
Herzschrittmacherther Elektrophysiol ; 34(4): 339-350, 2023 Dec.
Artigo em Alemão | MEDLINE | ID: mdl-37917360

RESUMO

Lead extraction due to infection or lead dysfunction has become more important in recent years. Patients with high risk of severe and life-threatening complications should only undergo surgery in experienced centers where appropriate personnel and equipment are available. In this review, different techniques and methods to safely and successfully perform transvenous lead extraction are summarized.


Assuntos
Desfibriladores Implantáveis , Marca-Passo Artificial , Humanos , Desfibriladores Implantáveis/efeitos adversos , Marca-Passo Artificial/efeitos adversos , Resultado do Tratamento , Remoção de Dispositivo/efeitos adversos , Remoção de Dispositivo/métodos , Estudos Retrospectivos
18.
Europace ; 25(11)2023 11 02.
Artigo em Inglês | MEDLINE | ID: mdl-37935403

RESUMO

Arrhythmogenic right ventricular cardiomyopathy (ARVC) is a heritable cardiomyopathy characterized by a predominantly arrhythmic presentation. It represents the leading cause of sudden cardiac death (SCD) among athletes and poses a significant morbidity threat in the general population. As a causative treatment for ARVC is still not available, the placement of an implantable cardioverter defibrillator represents the current cornerstone for SCD prevention in this setting. Thanks to international ARVC-dedicated efforts, significant steps have been achieved in recent years towards an individualized, patient-centred risk stratification approach. A novel risk calculator algorithm estimating the 5-year risk of arrhythmias of patients with ARVC has been introduced in clinical practice and subsequently validated. The purpose of this article is to summarize the body of evidence that has allowed the development of this tool and to discuss the best way to implement its use in the care of an individual patient.


Assuntos
Displasia Arritmogênica Ventricular Direita , Desfibriladores Implantáveis , Humanos , Fatores de Risco , Displasia Arritmogênica Ventricular Direita/complicações , Displasia Arritmogênica Ventricular Direita/diagnóstico , Displasia Arritmogênica Ventricular Direita/terapia , Morte Súbita Cardíaca/etiologia , Morte Súbita Cardíaca/prevenção & controle , Morte Súbita Cardíaca/epidemiologia , Arritmias Cardíacas/diagnóstico , Arritmias Cardíacas/terapia , Arritmias Cardíacas/complicações , Desfibriladores Implantáveis/efeitos adversos , Medição de Risco
19.
Braz J Cardiovasc Surg ; 38(5): e2023144, 2023 10 23.
Artigo em Inglês | MEDLINE | ID: mdl-37871254

RESUMO

OBJECTIVE: To investigate the risk factors for predicting atrial high-rate episodes (AHREs) detected by cardiac implantable electronic devices (CIEDs). METHODS: A total of 140 patients with CIED in our hospital from June 2013 to June 2018 were included and were followed up to observe whether they had AHREs. AHRE are defined as atrial rate ≥ 175 times/minute, lasting > 5 minutes, and reviewed by an experienced electrophysiologist with unclear clinical diagnosis. The patients fasted for 12 hours after implantation, and blood samples were collected for biochemical, lipid, and whole blood count detection. Follow-up was regular after discharge to record follow-up data of each patient and conduct statistical analysis. RESULTS: One hundred and forty patients were implanted with dual-chamber pacemakers, their median age was 70 years old, 44.29% were male, 27 patients had AHRE within one year, and AHRE incidence rate was 19.29%. The microcytic to hypochromic (M/H) ratio was calculated for all AHRE patients and compared with the patients without AHRE; the M/H value of AHRE patients was significantly higher. Throughout the entire follow-up period, a total of 44 patients developed AHRE; when adjusted by multivariate analysis, only M/H ratio ≥ 4.5 vs. < 4.5 had statistical significance, and the adjusted hazard ratio value was 4.313 (1.675-11.105). CONCLUSION: As an indicator, M/H ratio may play an important role in the occurrence and development of atrial fibrillation and can be used as a predictor of AHRE in patients with CIED.


Assuntos
Fibrilação Atrial , Desfibriladores Implantáveis , Marca-Passo Artificial , Acidente Vascular Cerebral , Humanos , Masculino , Idoso , Feminino , Fibrilação Atrial/complicações , Monócitos , Desfibriladores Implantáveis/efeitos adversos , Átrios do Coração , Marca-Passo Artificial/efeitos adversos , Fatores de Risco , Acidente Vascular Cerebral/etiologia
20.
Radiother Oncol ; 189: 109910, 2023 12.
Artigo em Inglês | MEDLINE | ID: mdl-37709052

RESUMO

BACKGROUND: Stereotactic arrhythmia radioablation (STAR) appears to be beneficial in selected patients with therapy-refractory ventricular tachycardia (VT). However, high-dose radiotherapy used for STAR-treatment may affect functioning of the patients' implantable cardioverter defibrillator (ICD) by direct effects of radiation on ICD components or cardiac tissue. Currently, the effect of STAR on ICD functioning remains unknown. METHODS: A retrospective pre-post multicenter study evaluating ICD functioning in the 12-month before and after STAR was performed. Patients with (non)ischemic cardiomyopathies with therapy-refractory VT and ICD who underwent STAR were included and the occurrence of ICD-related adverse events was collected. Evaluated ICD parameters included sensing, capture threshold and impedance. A linear mixed-effects model was used to investigate the association between STAR, radiotherapy dose and changes in lead parameters over time. RESULTS: In total, 43 patients (88% male) were included in this study. All patients had an ICD with an additional right atrial lead in 34 (79%) and a ventricular lead in 17 (40%) patients. Median ICD-generator dose was 0.1 Gy and lead tip dose ranged from 0-32 Gy. In one patient (2%), a reset occurred during treatment, but otherwise, STAR and radiotherapy dose were not associated with clinically relevant alterations in ICD leads parameters. CONCLUSIONS: STAR treatment did not result in major ICD malfunction. Only one radiotherapy related adverse event occurred during the study follow-up without patient harm. No clinically relevant alterations in ICD functioning were observed after STAR in any of the leads. With the reported doses STAR appears to be safe.


Assuntos
Desfibriladores Implantáveis , Isquemia Miocárdica , Taquicardia Ventricular , Humanos , Masculino , Feminino , Desfibriladores Implantáveis/efeitos adversos , Taquicardia Ventricular/etiologia , Taquicardia Ventricular/terapia , Estudos Retrospectivos , Arritmias Cardíacas/etiologia , Isquemia Miocárdica/etiologia , Resultado do Tratamento
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA