1.
Am J Gastroenterol ; 96(6): 1803-6, 2001 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-11419833

RESUMO

OBJECTIVES: The effectiveness of manual cleaning and disinfection of gastroendoscopes with 3% glutaraldehyde in decreasing the risk of transmission of hepatitis C virus (HCV) was examined. METHODS: Gastroendoscopes used for treatment of endoscopic esophageal variceal ligation in patients with HCV infection were manually cleaned and disinfected with 3% glutaraldehyde (n = 25), 2% glutaraldehyde (n = 17), or 0.1% benzethonium chloride (n = 25). Samples were obtained by pouring 20 ml of 0.9% NaCl solution into the biopsy-suction channel of the scope before and after cleaning and disinfection. HCV was detected with the polymerase chain reaction. RESULTS: Of the 25 scopes in 3% glutaraldehyde group, nine (36%) were positive for HCV before cleaning and disinfection, but all became negative after cleaning and disinfection; the difference was statistically significant (p < 0.01). In contrast, in 2% glutaraldehyde group and in the routine cleaning group, there were no significant differences in the number of positive samples between before and after cleaning and disinfection. CONCLUSION: Manual cleaning and disinfection of gastroscopes with 3% glutaraldehyde is useful for decreasing the risk of transmission of HCV among patients.


Assuntos
Desinfetantes/farmacologia , Gastroscópios/virologia , Glutaral/farmacologia , Hepatite C/transmissão , Benzetônio/farmacologia , Varizes Esofágicas e Gástricas/virologia , Feminino , Humanos , Masculino , Pessoa de Meia-Idade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA