ABSTRACT

BACKGROUND:

Simultaneous (201)Tl/(99m)Tc-sestamibi dual-isotope myocardial perfusion SPECT imaging can reduce imaging time and produce perfectly registered rest/stress images. However, crosstalk from (99m)Tc into (201)Tl images can significantly reduce (201)Tl image quality. We have developed a model-based compensation (MBC) method to compensate for this crosstalk. The method has previously been validated with phantom and simulation studies. In this study, we evaluated the MBC method using a canine model.

METHODS:

Left anterior descending or left circumflex coronary artery stenoses were created in 50 adult mongrel dogs weighing 20-30 kg. The dogs were injected with 111 MBq (3 mCi) of (201)Tl at rest, and a SPECT study acquired. Stress was induced by administering adenosine to the dog, followed by injection of 740 MBq (20 mCi) of (99m)Tc-sestamibi at peak stress. A second SPECT study was performed with data acquired in both (201)Tl and (99m)Tc energy windows to provide simultaneous dual-isotope projection data. The images were reconstructed using the ordered-subsets expectation-maximization reconstruction algorithm with compensation for attenuation, scatter, and detector response. For simultaneously acquired (201)Tl data, we also applied the MBC method to compensate for crosstalk contamination from (99m)Tc.

RESULTS:

Without compensation, (99m)Tc crosstalk increased the estimated (201)Tl activity concentration in the rest images and reduced defect contrast. After MBC, the (201)Tl images were in good agreement with the registered single-isotope images and ex vivo count data. The ischemic (IS) to non-ischemic (NIS) region (201)Tl activity concentration ratios were computed for single-isotope and dual-isotope studies. The correlation with ex vivo IS-NIS ratios was 0.815 after MBC, compared to the 0.495 from data without compensation. In addition, the regression line for the IS-NIS ratios with MBC was almost parallel to the line of identity with a slope of 0.93, compared to a slope of 0.45 without compensation.

CONCLUSIONS:

These results demonstrate that model-based crosstalk compensation can provide substantial reduction of crosstalk effects in simultaneously acquired myocardial perfusion SPECT images in living biological systems.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Asunto principal: Radioisótopos de Talio / Aumento de la Imagen / Tomografía Computarizada de Emisión de Fotón Único / Artefactos / Tecnecio Tc 99m Sestamibi / Estenosis Coronaria / Modelos Cardiovasculares Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2014 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Asunto principal: Radioisótopos de Talio / Aumento de la Imagen / Tomografía Computarizada de Emisión de Fotón Único / Artefactos / Tecnecio Tc 99m Sestamibi / Estenosis Coronaria / Modelos Cardiovasculares Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2014 Tipo del documento: Article