Energy response of GR-200A thermoluminescence dosemeters to 60Co and to monoenergetic synchrotron radiation in the energy range 28-40 keV.
Emiro, F; Di Lillo, F; Mettivier, G; Fedon, C; Longo, R; Tromba, G; Russo, P.
Afiliación
  • Emiro F; Dipartimento di Fisica, Università di Napoli Federico II, Via Cintia, Napoli I-80126, Italy INFN, Sezione di Napoli, Via Cintia, Napoli I-80126, Italy.
  • Di Lillo F; Dipartimento di Fisica, Università di Napoli Federico II, Via Cintia, Napoli I-80126, Italy INFN, Sezione di Napoli, Via Cintia, Napoli I-80126, Italy.
  • Mettivier G; Dipartimento di Fisica, Università di Napoli Federico II, Via Cintia, Napoli I-80126, Italy INFN, Sezione di Napoli, Via Cintia, Napoli I-80126, Italy mettivier@na.infn.it.
  • Fedon C; Dipartimento di Fisica, Università di Trieste, Via A. Valerio 2, Trieste I-34127, Italy INFN, Sezione di Trieste, Via A. Valerio 2, Trieste I-34127, Italy.
  • Longo R; Dipartimento di Fisica, Università di Trieste, Via A. Valerio 2, Trieste I-34127, Italy INFN, Sezione di Trieste, Via A. Valerio 2, Trieste I-34127, Italy.
  • Tromba G; Sincrotrone Trieste SCpA, Strada Statale S.S. 14 km 163.5, Basovizza, Trieste I-34012, Italy.
  • Russo P; Dipartimento di Fisica, Università di Napoli Federico II, Via Cintia, Napoli I-80126, Italy INFN, Sezione di Napoli, Via Cintia, Napoli I-80126, Italy.
Radiat Prot Dosimetry ; 168(1): 40-5, 2016 Jan.
Article en En | MEDLINE | ID: mdl-25737582
ABSTRACT
The response of LiFMg,Cu,P thermoluminescence dosemeters (type GR-200A) to monoenergetic radiation of energy 28, 35, 38 and 40 keV was evaluated with respect to irradiation with a calibrated (60)Co gamma-ray source. High-precision measurements of the relative air kerma response performed at the SYRMEP beamline of the ELETTRA synchrotron radiation facility (Trieste, Italy) showed a significant deviation of the average response to low-energy X-rays from that to (60)Co, with an over-response from 6 % (at 28 keV) to 22 % (at 40 keV). These data are not consistent with literature data for these dosemeters, where model predictions gave deviation from unity of the relative air kerma response of about 10 %. The authors conclude for the need of additional determinations of the low-energy relative response of GR-200A dosemeters, covering a wider range of monoenergetic energies sampled at a fine energy step, as planned in future experiments by their group at the ELETTRA facility.