Your browser doesn't support javascript.
loading
High-definition computed tomography for coronary artery stents: image quality and radiation doses for low voltage (100 kVp) and standard voltage (120 kVp) ECG-triggered scanning.
Lee, Ji Won; Kim, Chang Won; Lee, Han Cheol; Wu, Ming-Ting; Hwangbo, Lee; Choo, Ki Seok; Kim, June Hong; Lee, Ki-Nam; Kim, Jin You; Jeong, Yeon Joo.
Afiliación
  • Lee JW; Department of Radiology, Pusan National University Hospital, Pusan National University School of Medicine and Medical Research Institute, #1-10 Ami-Dong, Seo-Gu, Pusan, 602-739, Republic of Korea.
Int J Cardiovasc Imaging ; 31 Suppl 1: 39-49, 2015 Jun.
Article en En | MEDLINE | ID: mdl-26022439
ABSTRACT
The noninvasive assessment of coronary stents by coronary CT angiography (CCTA) is an attractive method. However, the radiation dose associated with CCTA remains a concern for patients. The purpose of this study is to compare the radiation doses and image qualities of CCTA performed using tube voltages of 100 or 120 kVp for the evaluation of coronary stents. After receiving institutional review board approval, 53 consecutive patients with previously implanted stents (101 stents) underwent 64-slice CCTA. Patients were divided into three different protocol groups, namely, prospective ECG triggering at 100 kVp, prospective ECG triggering at 120 kVp, or retrospective gating at 100 kVp. Two reviewers qualitatively scored the quality of the resulting images for coronary stents and determined levels of artificial lumen narrowing (ALN), stent lumen attenuation increase ratio (SAIR), image noise, and radiation dose parameters. No significant differences were found between the three protocol groups concerning qualitative image quality or SAIR. Coronary lumen attenuation and in-stent attenuation of 100 kVp prospective CCTA (P-CCTA) were higher than in the 120 kVp P-CCTA protocol (all Ps < 0.001). Mean ALN was significantly lower for 100 kVp P-CCTA than for 100 kVp retrospective CCTA (R-CCTA, P = 0.007). The mean effective radiation dose was significantly lower (P < 0.001) for 100 kVp P-CCTA (3.3 ± 0.4 mSv) than for the other two protocols (100 kVp R-CCTA 6.7 ± 1.0 mSv, 120 kVp P-CCTA 4.6 ± 1.2 mSv). We conclude that the use of 100 kVp P-CCTA can reduce radiation doses for patients while maintaining the imaging quality of 100 kVp R-CCTA and 120 kVp P-CCTA for the evaluation of coronary stents.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Asunto principal: Dosis de Radiación / Enfermedad de la Arteria Coronaria / Stents / Angiografía Coronaria / Vasos Coronarios / Reestenosis Coronaria / Electrocardiografía / Técnicas de Imagen Sincronizada Cardíacas / Tomografía Computarizada Multidetector / Intervención Coronaria Percutánea Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Qualitative_research / Risk_factors_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Cardiovasc Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2015 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Asunto principal: Dosis de Radiación / Enfermedad de la Arteria Coronaria / Stents / Angiografía Coronaria / Vasos Coronarios / Reestenosis Coronaria / Electrocardiografía / Técnicas de Imagen Sincronizada Cardíacas / Tomografía Computarizada Multidetector / Intervención Coronaria Percutánea Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Qualitative_research / Risk_factors_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Cardiovasc Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2015 Tipo del documento: Article