Validation of a New Recombinant Antibody Fragment (rFab)-Based Homogeneous Enzyme Immunoassay for the Highly Specific Detection of 6-Acetylmorphine in Urine.
Wang, Guohong; Huynh, Kim; Barhate, Rekha; Liu, Jialin; Tam, Phillip; Rodrigues, Warren; Coulter, Cynthia; Moore, Christine; Soares, James.
Afiliación
  • Wang G; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA gwang@immunalysis.com.
  • Huynh K; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
  • Barhate R; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
  • Liu J; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
  • Tam P; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
  • Rodrigues W; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
  • Coulter C; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
  • Moore C; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
  • Soares J; Immunalysis Corporation, 829 Towne Center Dr., Pomona, CA 91767, USA.
J Anal Toxicol ; 39(9): 726-33, 2015.
Article en En | MEDLINE | ID: mdl-26311850
ABSTRACT
The detection of 6-acetylmorphine (6-AM) in urine by immunoassay methods is challenging due to its short half-life and its similarity in structure to many commonly abused opiates that are often present at very high concentrations in urine. Current 6-AM homogeneous enzyme immunoassays use lyophilized reagents because of the instability of 6-AM in water or lack of the required specificity due to high cross-reactivity with morphine. A new 6-AM rFab-based homogeneous enzyme immunoassay (HEIA) has been developed with highly improved specificity. Using a cutoff concentration of 10 ng/mL, morphine or morphine glucuronides did not produce a positive signal up to 300,000 or 1,000,000 ng/mL, respectively. Assay imprecision (n = 80) was less than 1.5% using four replicates per day for 20 days over the range 0-20 ng/mL. Cross-reactivity with structurally related or non-related compounds was assessed at concentrations up to 1,000,000 ng/mL. Interferences from endogenous compounds at ±25% cutoff were also performed at the concentrations ranging from 100,000 to 500,000 ng/mL. The effect of varied pH values on assay performance at ±25% cutoff was investigated; no false-positive or false-negative results were observed between pH 4 and -11. Based on the analysis of 149 authentic urine samples, the accuracy of the 6-AM HEIA compared with LC-MS-MS was 100%. These results demonstrated that rFab can be suitable for traditional HEIA with desired detection sensitivity and stability.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Asunto principal: Fragmentos de Inmunoglobulinas / Técnicas para Inmunoenzimas / Derivados de la Morfina Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: J Anal Toxicol Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Asunto principal: Fragmentos de Inmunoglobulinas / Técnicas para Inmunoenzimas / Derivados de la Morfina Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: J Anal Toxicol Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos