A European multicientre study on the comparison of HIV-1 viral loads between VERIS HIV-1 Assay and Roche COBAS® TAQMAN® HIV-1 test, Abbott RealTime HIV-1 Assay, and Siemens VERSANT HIV-1 Assay.
Braun, Patrick; Delgado, Rafael; Drago, Monica; Fanti, Diana; Fleury, Hervé; Hofmann, Jörg; Izopet, Jacques; Kühn, Sebastian; Lombardi, Alessandra; Mancon, Alessandro; Marcos, Mª Angeles; Mileto, Davide; Sauné, Karine; O'Shea, Siobhan; Pérez-Rivilla, Alfredo; Ramble, John; Trimoulet, Pascale; Vila, Jordi; Whittaker, Duncan; Artus, Alain; Rhodes, Daniel.
Afiliación
  • Braun P; Laboratory Dr. Knechten, Medical Center for HIV and Hepatitis, Aachen, Germany.
  • Delgado R; Microbiology Department, Hospital Universitario 12 de Octubre and Insituto de Investigation, Hospital 12 de Octubre (i+12) Madrid, Spain.
  • Drago M; Clinical Chemistry and Microbiology Laboratories, Niguarda Ca' Granda Hospital, Milan, Italy.
  • Fanti D; Clinical Chemistry and Microbiology Laboratories, Niguarda Ca' Granda Hospital, Milan, Italy.
  • Fleury H; Virology Department, Hôpital Pellegrin, CHU, Bordeaux, France.
  • Hofmann J; Virology Department, Labor Berlin - Charité Vivantes GmbH, Berlin, Germany.
  • Izopet J; Department of Virology, Federative Institute of Biology, CHU, Toulouse, France.
  • Kühn S; Virology Department, Labor Berlin - Charité Vivantes GmbH, Berlin, Germany.
  • Lombardi A; Clinical Microbiology, Virology and Bioemergence Diagnosis, L. Sacco University Hospital, Milan, Italy.
  • Mancon A; Clinical Microbiology, Virology and Bioemergence Diagnosis, L. Sacco University Hospital, Milan, Italy.
  • Marcos MA; Department of Clinical Microbiology, Hospital Clinic, School of Medicine, University of Barcelona, Centre for International Health Research (CRESIB), Barcelona, Spain.
  • Mileto D; Virology Department, Labor Berlin - Charité Vivantes GmbH, Berlin, Germany.
  • Sauné K; Department of Virology, Federative Institute of Biology, CHU, Toulouse, France.
  • O'Shea S; Viapath Analytics, Infection Sciences, Guy's and St. Thomas' NHS Foundation Trust, London, United Kingdom.
  • Pérez-Rivilla A; Microbiology Department, Hospital Universitario 12 de Octubre and Insituto de Investigation, Hospital 12 de Octubre (i+12) Madrid, Spain.
  • Ramble J; Viapath Analytics, Infection Sciences, Guy's and St. Thomas' NHS Foundation Trust, London, United Kingdom.
  • Trimoulet P; Virology Department, Hôpital Pellegrin, CHU, Bordeaux, France.
  • Vila J; Department of Clinical Microbiology, Hospital Clinic, School of Medicine, University of Barcelona, Centre for International Health Research (CRESIB), Barcelona, Spain.
  • Whittaker D; Laboratory Medicine Building, North Lane, Northern General Hospital, Sheffield, United Kingdom.
  • Artus A; Beckman Coulter, Immunotech, 130 Ave Lattre de Tassigny, Marseille 13009, France.
  • Rhodes D; Beckman Coulter, Immunotech, 130 Ave Lattre de Tassigny, Marseille 13009, France. Electronic address: dwrhodes@beckman.com.
J Clin Virol ; 92: 75-82, 2017 07.
Article en En | MEDLINE | ID: mdl-28599228
ABSTRACT

BACKGROUND:

Viral load monitoring is essential for patients under treatment for HIV. Beckman Coulter has developed the VERIS HIV-1 Assay for use on the novel, automated DxN VERIS Molecular Diagnostics System.¥

OBJECTIVES:

Evaluation of the clinical performance of the new quantitative VERIS HIV-1 Assay at multiple EU laboratories. STUDY

DESIGN:

Method comparison with the VERIS HIV-1 Assay was performed with 415 specimens at 5 sites tested with COBAS® AmpliPrep/COBAS® TaqMan® HIV-1 Test, v2.0, 169 specimens at 3 sites tested with RealTime HIV-1 Assay, and 202 specimens from 2 sites tested with VERSANT HIV-1 Assay. Patient monitoring sample results from 4 sites were also compared.

RESULTS:

Bland-Altman analysis showed the average bias between VERIS HIV-1 Assay and COBAS HIV-1 Test, RealTime HIV-1 Assay, and VERSANT HIV-1 Assay to be 0.28, 0.39, and 0.61 log10 cp/mL, respectively. Bias at low end levels below 1000cp/mL showed predicted bias to be <0.3 log10 cp/mL for VERIS HIV-1 Assay versus COBAS HIV-1 Test and RealTime HIV-1 Assay, and <0.5 log10cp/mL versus VERSANT HIV-1 Assay. Analysis on 174 specimens tested with the 0.175mL volume VERIS HIV-1 Assay and COBAS HIV-1 Test showed average bias of 0.39 log10cp/mL. Patient monitoring results using VERIS HIV-1 Assay demonstrated similar viral load trends over time to all comparators.

CONCLUSIONS:

The VERIS HIV-1 Assay for use on the DxN VERIS System demonstrated comparable clinical performance to COBAS® HIV-1 Test, RealTime HIV-1 Assay, and VERSANT HIV-1 Assay.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Asunto principal: ARN Viral / Infecciones por VIH / VIH-1 / Carga Viral / Técnicas de Diagnóstico Molecular Tipo de estudio: Clinical_trials / Diagnostic_studies / Prognostic_studies / Screening_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: J Clin Virol Asunto de la revista: VIROLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Asunto principal: ARN Viral / Infecciones por VIH / VIH-1 / Carga Viral / Técnicas de Diagnóstico Molecular Tipo de estudio: Clinical_trials / Diagnostic_studies / Prognostic_studies / Screening_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: J Clin Virol Asunto de la revista: VIROLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Alemania