Adaptation of controlled attenuation parameter (CAP) measurement depth in morbidly obese patients addressed for bariatric surgery.
Somda, Sosthene; Lebrun, Amandine; Tranchart, Hadrien; Lamouri, Karima; Prevot, Sophie; Njike-Nakseu, Micheline; Gaillard, Martin; Lainas, Panagiotis; Balian, Axel; Dagher, Ibrahim; Perlemuter, Gabriel; Naveau, Sylvie; Voican, Cosmin Sebastian.
Afiliación
  • Somda S; Faculté de Médecine Paris-Sud, Univ Paris-Sud, Université Paris-Saclay, Le Kremlin-Bicêtre, France.
  • Lebrun A; Service d'Hépato-Gastroentérologie et Nutrition, Hôpital Antoine-Béclère, Hôpitaux Universitaires Paris-Sud, Assistance Publique-Hôpitaux de Paris, Clamart, France.
  • Tranchart H; Service d'Hépato-Gastroentérologie et Nutrition, Hôpital Antoine-Béclère, Hôpitaux Universitaires Paris-Sud, Assistance Publique-Hôpitaux de Paris, Clamart, France.
  • Lamouri K; INSERM U996, DHU Hepatinov, Labex LERMIT, Clamart, France.
  • Prevot S; Faculté de Médecine Paris-Sud, Univ Paris-Sud, Université Paris-Saclay, Le Kremlin-Bicêtre, France.
  • Njike-Nakseu M; Service de Chirurgie Digestive Minimale Invasive, Hôpital Antoine-Béclère, Hôpitaux Universitaires Paris-Sud, Assistance Publique-Hôpitaux de Paris, Clamart, France.
  • Gaillard M; Service d'Hépato-Gastroentérologie et Nutrition, Hôpital Antoine-Béclère, Hôpitaux Universitaires Paris-Sud, Assistance Publique-Hôpitaux de Paris, Clamart, France.
  • Lainas P; Faculté de Médecine Paris-Sud, Univ Paris-Sud, Université Paris-Saclay, Le Kremlin-Bicêtre, France.
  • Balian A; INSERM U996, DHU Hepatinov, Labex LERMIT, Clamart, France.
  • Dagher I; Service d'Anatomie pathologique, Hôpital Antoine-Béclère, Hôpitaux Universitaires Paris-Sud, Assistance Publique-Hôpitaux de Paris, Clamart, France.
  • Perlemuter G; Service d'Hépato-Gastroentérologie et Nutrition, Hôpital Antoine-Béclère, Hôpitaux Universitaires Paris-Sud, Assistance Publique-Hôpitaux de Paris, Clamart, France.
  • Naveau S; Faculté de Médecine Paris-Sud, Univ Paris-Sud, Université Paris-Saclay, Le Kremlin-Bicêtre, France.
  • Voican CS; Service de Chirurgie Digestive Minimale Invasive, Hôpital Antoine-Béclère, Hôpitaux Universitaires Paris-Sud, Assistance Publique-Hôpitaux de Paris, Clamart, France.
PLoS One ; 14(5): e0217093, 2019.
Article en En | MEDLINE | ID: mdl-31125357
ABSTRACT
BACKGROUND AND

AIM:

The controlled attenuation parameter (CAP) using FibroScan (Echosens, Paris, France) M or XL probe has been developed for liver steatosis assessment. However, CAP performs poorly in patients with high body mass index. The aim of our study was to assess whether CAP is overestimated using the standard XL probe in patients with morbid obesity, and in the case of an overestimation, to reprocess the data at a greater depth to obtain the appropriate CAP (CAPa). PATIENTS AND

METHODS:

We conducted an observational prospective cohort study on a total of 249 severely obese patients admitted to our institution to undergo sleeve gastrectomy. Patients had a liver biopsy performed during the surgery and a CAP measurement during the 15 days preceding biopsy. Patient files were reprocessed retrospectively by an algorithm, blinded to the patients' clinical data. The algorithm automatically assessed the probe-to-capsula distance (PCD) by analysing the echogenicity of ultrasound signals on the time-motion mode. In the case of a distance >35 mm, the algorithm automatically selected a deeper measurement for CAP (CAPa). When PCD was less than 35 mm, the measured CAP was considered as appropriated (CAPa) and no further reprocessing was performed.

RESULTS:

CAP recording was not performed at a sufficient depth in 130 patients. In these patients, the CAPa obtained at the adapted depth was significantly lower than CAP (298±3.9 versus 340±4.2 dB/m; p< 0.0001) measured at the standard depth (35 to 75 mm). Multiple linear regression analysis revealed that both body mass index and hepatic steatosis were independently correlated with CAP values. After reprocessing the CAP in patients with PCD > 35 mm, steatosis stage was the only parameter independently correlated with CAP values. For the diagnosis of steatosis (S≥1), moderate to severe steatosis (S≥2) and severe steatosis (S = 3), the AUROC curves of CAPa (measured CAP in patients with PCD<35 mm and reprocessed CAP in those with PCD>35 mm) were 0.86, 0.83 and 0.79, respectively. The Obuchowski measure for the diagnosis of steatosis was 0.90±0.013.

CONCLUSION:

CAP was overestimated in a half of morbidly obese patients using an XL probe, but CAP can be performed correctly in these patients after adapting the measurement depth.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Asunto principal: Obesidad Mórbida / Cirugía Bariátrica / Diagnóstico por Imagen de Elasticidad / Enfermedad del Hígado Graso no Alcohólico Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2019 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Asunto principal: Obesidad Mórbida / Cirugía Bariátrica / Diagnóstico por Imagen de Elasticidad / Enfermedad del Hígado Graso no Alcohólico Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2019 Tipo del documento: Article País de afiliación: Francia