Rotational Stability, Tilt and Decentration of a New IOL with a 7.0 mm Optic.
Wendelstein, Jascha; Laubichler, Peter; Fischinger, Isaak; Mariacher, Siegfried; Beka, Sophie; Mursch-Edlmayr, Anna; Siska, René; Langenbucher, Achim; Bolz, Matthias.
Afiliación
  • Wendelstein J; Department of Ophthalmology and Optometry, Kepler University Hospital GmbH, Johannes Kepler University Linz, Linz, Austria.
  • Laubichler P; Department of Ophthalmology and Optometry, Kepler University Hospital GmbH, Johannes Kepler University Linz, Linz, Austria.
  • Fischinger I; Department of Ophthalmology and Optometry, Kepler University Hospital GmbH, Johannes Kepler University Linz, Linz, Austria.
  • Mariacher S; Berlin Eye Research Institute and Eye Clinic Spreebogen, Berlin, Germany.
  • Beka S; Department of Ophthalmology and Optometry, Kepler University Hospital GmbH, Johannes Kepler University Linz, Linz, Austria.
  • Mursch-Edlmayr A; Department of Ophthalmology and Optometry, Kepler University Hospital GmbH, Johannes Kepler University Linz, Linz, Austria.
  • Siska R; Department of Ophthalmology and Optometry, Kepler University Hospital GmbH, Johannes Kepler University Linz, Linz, Austria.
  • Langenbucher A; Department of Ophthalmology and Optometry, Kepler University Hospital GmbH, Johannes Kepler University Linz, Linz, Austria.
  • Bolz M; Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany.
Curr Eye Res ; 46(11): 1673-1680, 2021 11.
Article en En | MEDLINE | ID: mdl-34121551
ABSTRACT

Purpose:

To evaluate rotational stability, tilt and decentration of a new monofocal IOL with a 7.0 mm optic and frame haptics.

Methods:

Prospective post-market clinical follow-up study at the Kepler University Hospital Linz, Austria. An Aspira-aXA (HumanOptics, Germany) was implanted in 74 eyes of 42 cataract patients. The lens was manufactured with toric markings. IOL rotational stability was evaluated by comparing its position at the end of surgery (EoS) versus 1 day, 1 week, 1 month and 4 months postoperatively. IOL tilt and decentration were measured using a Scheimpflug camera at 1 week, 1 month and 4 months.

Results:

Median absolute IOL rotation was 1.42 degrees (n = 52; mean = 2.18 ± 2.23°) within 1 day after surgery and was significantly higher compared to all later intervals (median <1.0 degree; P = .001). At the 4 months follow-up, IOL rotation was within 5.0 degrees in 85% of the eyes (n = 40) and within 10.0 degrees in 98% (n = 46) of the eyes. The only eye with an IOL rotation of ≥ 10.ty0 degrees (EoS vs. 1 day) had an AL of 26.45 mm. At the last follow-up, the IOL vertical and horizontal tilt referenced to the pupillary axis was in average less than 1.5 degrees in both eyes (n = 54; maximum 5.85°). Decentration in both meridians was on average less than 0.10 mm in both eyes (maximum 0.30 mm).

Conclusion:

The one-piece Aspira-aXA IOL showed good and stable positioning within the capsular bag over a 4 months period.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Asunto principal: Rotación / Facoemulsificación / Seudofaquia / Implantación de Lentes Intraoculares / Migracion de Implante de Lente Artificial / Lentes Intraoculares Tipo de estudio: Observational_studies / Prognostic_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Curr Eye Res Año: 2021 Tipo del documento: Article País de afiliación: Austria

Texto completo: 1 Colección: 01-internacional Asunto principal: Rotación / Facoemulsificación / Seudofaquia / Implantación de Lentes Intraoculares / Migracion de Implante de Lente Artificial / Lentes Intraoculares Tipo de estudio: Observational_studies / Prognostic_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Curr Eye Res Año: 2021 Tipo del documento: Article País de afiliación: Austria