Prediction of corneal power vectors after cataract surgery with toric lens implantation-A vector analysis.
Langenbucher, Achim; Szentmáry, Nóra; Cayless, Alan; Wendelstein, Jascha; Hoffmann, Peter.
Afiliación
  • Langenbucher A; Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany.
  • Szentmáry N; Dr. Rolf M. Schwiete Center for Limbal Stem Cell and Aniridia Research, Saarland University, Homburg/Saar, Germany.
  • Cayless A; Department of Ophthalmology, Semmelweis-University, Budapest, Hungary.
  • Wendelstein J; School of Physical Sciences, The Open University, Milton Keynes, United Kingdom.
  • Hoffmann P; Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany.
PLoS One ; 18(9): e0288316, 2023.
Article en En | MEDLINE | ID: mdl-37682881
ABSTRACT

BACKGROUND:

Intraocular lenses are typically calculated based on a pseudophakic eye model, and for toric lenses (tIOL) a good estimate of corneal astigmatism after cataract surgery is required in addition to the equivalent corneal power. The purpose of this study was to investigate the differences between the preoperative IOLMaster (IOLM) and the preoperative and postoperative Casia2 (CASIA) tomographic measurements of corneal power in a cataractous population with tIOL implantation, and to predict total power (TP) from the IOLM and CASIA keratometric measurements.

METHODS:

The analysis was based on a dataset of 88 eyes of 88 patients from 1 clinical centre before and after tIOL implantation. All IOLM and CASIA keratometric and total corneal power measurements were converted to power vector components, and the differences between preoperative IOLM or CASIA and postoperative CASIA measurements were assessed. Feedforward neural network and multivariate linear regression prediction algorithms were implemented to predict the postoperative total corneal power (as a reference for tIOL calculation) from the preoperative IOLM and CASIA keratometric measurements.

RESULTS:

On average, the preoperative IOLM keratometric / total corneal power under- / overestimates the postoperative CASIA keratometric / real corneal power by 0.12 dpt / 0.21 dpt. The prediction of postoperative CASIA real power from preoperative IOLM or CASIA keratometry shows that postoperative total corneal power is systematically (0.18 dpt / 0.27 dpt) shifted towards astigmatism against the rule, which is not reflected by keratometry. The correlation of postoperative CASIA real power to the corresponding preoperative CASIA values is better than those as compared to the preoperative IOLM keratometry. However, there is a large variation from preoperative IOLM or CASIA keratometry to the postoperative CASIA real power of up to 1.1 dpt (95% confidence interval).

CONCLUSION:

One of the challenges of tIOL calculation is the prediction of postoperative total corneal power from preoperative keratometry. Keratometric power restricted to a front surface measurement does not fully reflect the situation of corneal back surface astigmatism, which typically adds some extra against the rule astigmatism.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Asunto principal: Astigmatismo / Catarata / Enfermedades de la Córnea / Fabaceae / Lentes / Lentes Intraoculares Tipo de estudio: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2023 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Asunto principal: Astigmatismo / Catarata / Enfermedades de la Córnea / Fabaceae / Lentes / Lentes Intraoculares Tipo de estudio: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2023 Tipo del documento: Article País de afiliación: Alemania