Reliability and validity of handheld structured light scanners and a static stereophotogrammetry system in facial three-dimensional surface imaging.
Schipper, J A M; Merema, B J; Hollander, M H J; Spijkervet, F K L; Dijkstra, P U; Jansma, J; Schepers, R H; Kraeima, J.
Afiliación
  • Schipper JAM; Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands. j.a.m.schipper@umcg.nl.
  • Merema BJ; Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Hollander MHJ; Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Spijkervet FKL; Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Dijkstra PU; Department of Rehabilitation Medicine, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Jansma J; Sirindhorn School of Prosthetics and Orthotics, Faculty of Medicine Siriraj Hospital, Mahidol University, 14 Arun Amarin Rd, Bangkok, 10700, Thailand.
  • Schepers RH; Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Kraeima J; Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
Sci Rep ; 14(1): 8172, 2024 04 08.
Article en En | MEDLINE | ID: mdl-38589391
ABSTRACT
Several new systems for three-dimensional (3D) surface imaging of the face have become available to assess changes following orthognathic or facial surgery. Before they can be implemented in practice, their reliability and validity must be established. Our aim, therefore, was to study the intra- and inter-system reliability and validity of 3dMD (stereophotogrammetry), Artec Eva and Artec Space Spider (both structured light scanners). Intra- and inter-system reliability, expressed in root mean square distance, was determined by scanning a mannequin's head and the faces of healthy volunteers multiple times. Validity was determined by comparing the linear measurements of the scans with the known distances of a 3D printed model. Post-processing errors were also calculated. Intra-system reliability after scanning the mannequin's head was best with the Artec Space Spider (0.04 mm Spider; 0.07 mm 3dMD; 0.08 mm Eva). The least difference in inter-system reliability after scanning the mannequin's head was between the Artec Space Spider and Artec Eva. The best intra-system reliability after scanning human subjects was with the Artec Space Spider (0.15 mm Spider; 0.20 mm Eva; 0.23 mm 3dMD). The least difference in inter-system reliability after scanning human subjects was between the Artec Eva and Artec Space Spider. The most accurate linear measurement validity occurred with the Artec Space Spider. The post-processing error was 0.01 mm for all the systems. The Artec Space Spider is the most reliable and valid scanning system.