Dosimetric verification of the anisotropic analytical algorithm in lung equivalent heterogeneities with and without bone equivalent heterogeneities.
Ono, Kaoru; Endo, Satoru; Tanaka, Kenichi; Hoshi, Masaharu; Hirokawa, Yutaka.
Afiliação
  • Ono K; Department of Radiation Physics, Hiroshima Heiwa Clinic, Japan. koukun@ms4.megaegg.ne.jp
Med Phys ; 37(8): 4456-63, 2010 Aug.
Article em En | MEDLINE | ID: mdl-20879604
ABSTRACT

PURPOSE:

In this study, the authors evaluated the accuracy of dose calculations performed by the convolution/superposition based anisotropic analytical algorithm (AAA) in lung equivalent heterogeneities with and without bone equivalent heterogeneities.

METHODS:

Calculations of PDDs using the AAA and Monte Carlo simulations (MCNP4C) were compared to ionization chamber measurements with a heterogeneous phantom consisting of lung equivalent and bone equivalent materials. Both 6 and 10 MV photon beams of 4 x 4 and 10 x 10 cm(2) field sizes were used for the simulations. Furthermore, changes of energy spectrum with depth for the heterogeneous phantom using MCNP were calculated.

RESULTS:

The ionization chamber measurements and MCNP calculations in a lung equivalent phantom were in good agreement, having an average deviation of only 0.64 +/- 0.45%. For both 6 and 10 MV beams, the average deviation was less than 2% for the 4 x 4 and 10 x 10 cm(2) fields in the water-lung equivalent phantom and the 4 x 4 cm(2) field in the water-lung-bone equivalent phantom. Maximum deviations for the 10 x 10 cm(2) field in the lung equivalent phantom before and after the bone slab were 5.0% and 4.1%, respectively. The Monte Carlo simulation demonstrated an increase of the low-energy photon component in these regions, more for the 10 X 10 cm(2) field compared to the 4 x 4 cm(2) field.

CONCLUSIONS:

The low-energy photon by Monte Carlo simulation component increases sharply in larger fields when there is a significant presence of bone equivalent heterogeneities. This leads to great changes in the build-up and build-down at the interfaces of different density materials. The AAA calculation modeling of the effect is not deemed to be sufficiently accurate.
Assuntos

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Radiometria / Osso e Ossos / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Pulmão / Modelos Biológicos Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Radiometria / Osso e Ossos / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Pulmão / Modelos Biológicos Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Japão