Breast MR imaging at 3 T with dual-source radiofrequency transmission offers superior B1 homogeneity: an intraindividual comparison with breast MR imaging at 1.5 T.
Trop, Isabelle; Gilbert, Guillaume; Ivancevic, Marko K; Beaudoin, Gilles.
Afiliação
  • Trop I; Radiology Department, Breast Imaging Centre, 3è Le Royer, Centre Hospitalier de l'Université de Montréal, Hôtel-Dieu de Montréal, 3840 Saint-Urbain, Montreal, QC, Canada H2W 1T8. itrop@yahoo.com
Radiology ; 267(2): 602-8, 2013 May.
Article em En | MEDLINE | ID: mdl-23401582
ABSTRACT

PURPOSE:

To evaluate and compare B1 homogeneity for breast magnetic resonance (MR) imaging performed at 3 T with dual-source radiofrequency (RF) transmission to 1.5-T MR imaging and 3-T MR imaging with quadrature transmission. MATERIALS AND

METHODS:

This prospective study received institutional review board approval and patients provided informed consent. Women (n = 25; mean age, 53 years; range, 30-68 years) suspected of having breast lesions underwent breast MR imaging examinations on comparable 1.5-T and 3-T clinical systems between February and May 2012. B1 maps were obtained at 1.5 T and at 3 T with quadrature and dual-source RF transmission. Intrabreast differences and differences in mean B1 values between right and left breasts were investigated by using two-sided multivariate analysis of variance with interaction; t tests were used to compare the differences between measured whole-breast mean B1 values and requested B1 values.

RESULTS:

With quadrature transmission at 1.5 T and 3 T, the mean B1 values showed a statistically significant difference left-breast measured B1 was -8.9% of requested B1 value at 1.5 T and -13.7% at 3 T (P < .001), whereas right-breast measured B1 was +5.4% of requested B1 value at 1.5 T (P < .001) and +2.7% at 3 T (P = .01). With dual-source RF transmission at 3 T, mean B1 values across the breasts were not statistically different, nor were the measured B1 values compared with requested B1 values (left breast, -0.6%; right breast, -0.7%). At 3 T with dual-source transmission, slight intrabreast local variations in B1 were recorded.

CONCLUSION:

MR imaging at 3 T with dual-source RF transmission offered an overall B1 homogeneity for breast imaging that was better than that obtained at 1.5 T and with quadrature transmission.
Assuntos

Texto completo: 1 Coleções: 01-internacional Temas: Geral / Tipos_de_cancer / Outros_tipos Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Imageamento por Ressonância Magnética Tipo de estudo: Observational_studies Limite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: Radiology Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Temas: Geral / Tipos_de_cancer / Outros_tipos Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Imageamento por Ressonância Magnética Tipo de estudo: Observational_studies Limite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: Radiology Ano de publicação: 2013 Tipo de documento: Article