8-mm versus 10-mm diameter self-expandable metallic stent in bilateral endoscopic stent-in-stent deployment for malignant hilar biliary obstruction.
Naitoh, Itaru; Nakazawa, Takahiro; Ban, Tesshin; Okumura, Fumihiro; Hirano, Atsuyuki; Takada, Hiroki; Togawa, Shozo; Hayashi, Kazuki; Miyabe, Katsuyuki; Shimizu, Shuya; Kondo, Hiromu; Nishi, Yuji; Yoshida, Michihiro; Yamashita, Hiroaki; Umemura, Shuichiro; Hori, Yasuki; Kato, Akihisa; Sano, Hitoshi; Joh, Takashi.
Afiliação
  • Naitoh I; Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan. inaito@med.nagoya-cu.ac.jp.
J Hepatobiliary Pancreat Sci ; 22(5): 396-401, 2015 May.
Article em En | MEDLINE | ID: mdl-25611012
ABSTRACT

BACKGROUND:

We aimed to clarify the clinical benefits of using 8-mm versus 10-mm diameter self-expandable metallic stent (SEMS) in bilateral endoscopic stent-in-stent (SIS) deployment for malignant hilar biliary obstruction (MHBO).

METHODS:

A total of 72 consecutive patients with MHBO treated by endoscopic bilateral SIS deployment were enrolled. The 8-mm and 10-mm diameter SEMSs were used in 37 patients (8-mm group) and 35 patients (10-mm group), respectively. We retrospectively compared technical success, functional success, complications, successful re-intervention, time to recurrent biliary obstruction (RBO) and survival time between two groups.

RESULTS:

There were no significant inter-group differences in terms of technical success (8-mm vs. 10-mm group, both 100%), functional success (96% vs. 100%), early (5% vs. 9%) and late (19% vs. 14%) complications other than RBO, and successful endoscopic re-intervention (94% vs. 100%), respectively. Time to RBO (median 198 days vs. 222 days; log-rank, P = 0.440) and survival time (median 186 days vs. 235 days; log-rank, P = 0.883) did not differ significantly between the two groups. However, the success rate of endoscopic bilateral revisionary stent insertion for RBO was significantly higher in the 10-mm compared to the 8-mm group (68% vs. 31%; P = 0.044).

CONCLUSIONS:

The 10-mm diameter SEMS is more suitable with regard to endoscopic re-intervention.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Temas: Geral / Tipos_de_cancer / Outros_tipos Base de dados: MEDLINE Assunto principal: Neoplasias dos Ductos Biliares / Ductos Biliares Intra-Hepáticos / Colestase / Stents Metálicos Autoexpansíveis Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Hepatobiliary Pancreat Sci Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Temas: Geral / Tipos_de_cancer / Outros_tipos Base de dados: MEDLINE Assunto principal: Neoplasias dos Ductos Biliares / Ductos Biliares Intra-Hepáticos / Colestase / Stents Metálicos Autoexpansíveis Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Hepatobiliary Pancreat Sci Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Japão