ABSTRACT
BACKGROUND:
CD4 T-cell counts are widely used to assess treatment eligibility and to follow-up HIV-infected patients. The World Health Organization prequalification of in vitro diagnostics program conducted a performance evaluation of the FACSPresto (BD Biosciences), a new point-of-care instrument to measure absolute CD4-T cell (CD4) counts and percentages in venous and capillary blood samples from HIV-infected patients.METHODS:
Patients were recruited in Belgium (200 patients) and in Tanzania (247 patients). Venous blood samples were analyzed in two nearby reference laboratories. In addition, nurses/technicians collected a capillary blood sample by finger prick directly into a FACSPresto CD4 cartridge. Assay precision was assessed on fresh blood and on external quality control samples. Trueness (bias) was assessed by comparing results from FACSPresto with the reference (single-platform FACSCalibur). Clinical misclassification was measured at 200, 350 and 500 cells/µL thresholds.RESULTS:
Intra-assay precision was < 6%, and inter-assay < 8%. CD4 results from FACSPresto and reference method resulted in regression slopes of 0.99-1.11 using either venous or capillary blood. Correlation was better for venous than for capillary blood (minimum 0.97 vs 0.93 respectively). Capillary blood resulted in a larger bias than venous blood, with 24 and 83 cells/µL for absolute CD4 counts on capillary blood in Antwerp and Dar es Salaam respectively, vs 12 and 41 cells/µL on venous blood. Bias on CD4% was < 1% on both venous and capillary blood, and was proportionally better than for absolute CD4 counts. Clinical misclassification was in line with the average overestimation, showing a very good specificity, but sensitivity around 70-90%. The rejection rate was 11% on first reading, leading to 6% of all samples without final result after a second reading.CONCLUSIONS:
The FACSPresto performed very well on venous blood samples, and well on capillary blood samples.
Texto completo:
1
Coleções:
01-internacional
Temas:
Cuidados_paliativos
/
Geral
Base de dados:
MEDLINE
Assunto principal:
Linfócitos T CD4-Positivos
/
Infecções por HIV
/
HIV
Limite:
Adolescent
/
Adult
/
Aged
/
Female
/
Humans
/
Male
/
Middle aged
País/Região como assunto:
Africa
/
Europa
Idioma:
En
Revista:
PLoS One
Assunto da revista:
CIENCIA
/
MEDICINA
Ano de publicação:
2017
Tipo de documento:
Article
País de afiliação:
Bélgica