An external dosimetry audit programme to credential static and rotational IMRT delivery for clinical trials quality assurance.
Eaton, David J; Tyler, Justine; Backshall, Alex; Bernstein, David; Carver, Antony; Gasnier, Anne; Henderson, Julia; Lee, Jonathan; Patel, Rushil; Tsang, Yatman; Yang, Huiqi; Zotova, Rada; Wells, Emma.
Afiliação
  • Eaton DJ; Radiotherapy Trials QA Group (RTTQA), Mount Vernon Hospital, Northwood, UK. Electronic address: davideaton@nhs.net.
  • Tyler J; RTTQA, Royal Marsden Hospital, Fulham Road, London, UK.
  • Backshall A; RTTQA, Royal Marsden Hospital, Fulham Road, London, UK.
  • Bernstein D; RTTQA, Royal Marsden Hospital, Fulham Road, London, UK.
  • Carver A; RTTQA, Queen Elizabeth Hospital, Birmingham, UK.
  • Gasnier A; RTTQA, Royal Marsden Hospital, Fulham Road, London, UK.
  • Henderson J; RTTQA, Royal Marsden Hospital, Fulham Road, London, UK.
  • Lee J; Radiotherapy Trials QA Group (RTTQA), Mount Vernon Hospital, Northwood, UK.
  • Patel R; Radiotherapy Trials QA Group (RTTQA), Mount Vernon Hospital, Northwood, UK.
  • Tsang Y; Radiotherapy Trials QA Group (RTTQA), Mount Vernon Hospital, Northwood, UK.
  • Yang H; Radiotherapy Trials QA Group (RTTQA), Mount Vernon Hospital, Northwood, UK.
  • Zotova R; Radiotherapy Trials QA Group (RTTQA), Mount Vernon Hospital, Northwood, UK.
  • Wells E; RTTQA, Royal Marsden Hospital, Fulham Road, London, UK.
Phys Med ; 35: 25-30, 2017 Mar.
Article em En | MEDLINE | ID: mdl-28236559
ABSTRACT

PURPOSE:

External dosimetry audits give confidence in the safe and accurate delivery of radiotherapy. The RTTQA group have performed an on-site audit programme for trial recruiting centres, who have recently implemented static or rotational IMRT, and those with major changes to planning or delivery systems.

METHODS:

Measurements of reference beam output were performed by the host centre, and by the auditor using independent equipment. Verification of clinical plans was performed using the ArcCheck helical diode array.

RESULTS:

A total of 54 measurement sessions were performed between May 2014 and June 2016 at 28 UK institutions, reflecting the different combinations of planning and delivery systems used at each institution. Average ratio of measured output between auditor and host was 1.002±0.006. Average point dose agreement for clinical plans was -0.3±1.8%. Average (and 95% lower confidence intervals) of gamma pass rates at 2%/2mm, 3%/2mm and 3%/3mm respectively were 92% (80%), 96% (90%) and 98% (94%). Moderately significant differences were seen between fixed gantry angle and rotational IMRT, and between combination of planning systems and linac manufacturer, but not between anatomical treatment site or beam energy.

CONCLUSION:

An external audit programme has been implemented for universal and efficient credentialing of IMRT treatments in clinical trials. Good agreement was found between measured and expected doses, with few outliers, leading to a simple table of optimal and mandatory tolerances for approval of dosimetry audit results. Feedback was given to some centres leading to improved clinical practice.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Ensaios Clínicos como Assunto / Radioterapia de Intensidade Modulada Tipo de estudo: Clinical_trials País/Região como assunto: Europa Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Ensaios Clínicos como Assunto / Radioterapia de Intensidade Modulada Tipo de estudo: Clinical_trials País/Região como assunto: Europa Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2017 Tipo de documento: Article