Your browser doesn't support javascript.
loading
The pathology of failed McKee-Farrar implants: correlation with modern metal-on-metal-implant failure.
Munemoto, Mitsuru; Grammatopoulos, George; Tanaka, Yasuhito; Gibbons, Max; Athanasou, Nicholas A.
Afiliação
  • Munemoto M; Nuffield Department Orthopaedics Rheumatology & Musculoskeletal Sciences (NDORMS), University of Oxford, Windmill Road, Oxford, UK.
  • Grammatopoulos G; Dept of Orthopaedic Surgery, Nara medical University, Kashihara-City, 634-8522, Nara, Japan.
  • Tanaka Y; Nuffield Department Orthopaedics Rheumatology & Musculoskeletal Sciences (NDORMS), University of Oxford, Windmill Road, Oxford, UK. george.grammatopoulos@ndorms.ox.ac.uk.
  • Gibbons M; Nuffield Orthopaedic Centre, Windmill Road, Headington, Oxford, OX3 7LD, UK. george.grammatopoulos@ndorms.ox.ac.uk.
  • Athanasou NA; Dept of Orthopaedic Surgery, Nara medical University, Kashihara-City, 634-8522, Nara, Japan.
J Mater Sci Mater Med ; 28(5): 66, 2017 May.
Article em En | MEDLINE | ID: mdl-28332156
ABSTRACT
The McKee-Farrar (MF) prosthesis was the first widely used total hip replacement (THR) to employ a metal-on-metal (MoM) articulation. These implants had a high rate of early aseptic loosening but a number achieved good long-term implant survival, stimulating the reintroduction of second and third generation implants of this type. In this study we analysed archival histopathology of periprosthetic tissues in twenty cases of MF aseptic implant failure to determine if there was evidence of an innate and adaptive immune response similar to that seen in modern MoM implants. The presence of macrophages, the extent of necrosis and the ALVAL response were graded semi-quantitatively. Variable but in most cases extensive tissue necrosis was associated with a heavy macrophage response to Cobalt-Chrome (Co-Cr) wear particles in periprosthetic tissues; most cases also contained evidence of a predominantly lymphocyte response which in eight cases was moderate or heavy (Oxford Grade 2/3). Our findings show that inflammatory and necrotic changes to deposition of Co-Cr wear particles are found in periprosthetic tissues of failed MF implants, indicating that there is an innate and adaptive response similar to that noted in second/third generation MoM implants; they also suggest that the pathobiological response to metal wear particles is likely to have contributed to MF implant failure in these cases.
Assuntos

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Falha de Prótese / Próteses Articulares Metal-Metal / Prótese de Quadril Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Mater Sci Mater Med Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Falha de Prótese / Próteses Articulares Metal-Metal / Prótese de Quadril Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Mater Sci Mater Med Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Reino Unido