Bench Validation of a Handcrafted Prototype Catheter for Intra-gastric Pressure Monitoring.
García, Alberto Federico; Chica Yantén, Julián; Sánchez, Álvaro Ignacio; Aldana, Jose Luis; Mejía, Jorge Humberto; Burbano, Daniela; Salazar, Camilo.
Afiliação
  • García AF; Center for Clinical Research (CIC), Fundación Valle del Lili, Carrera 98 No. 18-49, Torre 7, Piso 3 (CIC), Cali, Colombia. alberto.garcia@fvl.org.co.
  • Chica Yantén J; Department of Surgery, Fundación Valle del Lili, Carrera 98 No. 18-49, Cali, Colombia. alberto.garcia@fvl.org.co.
  • Sánchez ÁI; School of Medicine, Universidad ICESI, Calle 18 No. 122-135, Cali, Colombia. alberto.garcia@fvl.org.co.
  • Aldana JL; Center for Clinical Research (CIC), Fundación Valle del Lili, Carrera 98 No. 18-49, Torre 7, Piso 3 (CIC), Cali, Colombia.
  • Mejía JH; School of Medicine, Universidad ICESI, Calle 18 No. 122-135, Cali, Colombia.
  • Burbano D; Center for Clinical Research (CIC), Fundación Valle del Lili, Carrera 98 No. 18-49, Torre 7, Piso 3 (CIC), Cali, Colombia.
  • Salazar C; Department of Surgery, Fundación Valle del Lili, Carrera 98 No. 18-49, Cali, Colombia.
World J Surg ; 44(6): 1706-1711, 2020 06.
Article em En | MEDLINE | ID: mdl-32016541
ABSTRACT

BACKGROUND:

Abdominal compartment syndrome is a sustained intra-abdominal pressure (IAP) >20 mm Hg associated with new organ dysfunction. In order to prevent its development and related complications, IAP monitoring should be performed in patients with risk factors. Although techniques for its monitoring have been developed, they are of high cost and not always available in low- and lower-middle-income countries. Therefore, we aim to develop and validate in a bench model a handcrafted catheter to be used as an alternative method to measure the intra-gastric pressure (IGP) as a surrogate of the IAP.

METHODS:

We used an acrylic water container as a model of the abdomen and four handcrafted catheters made of a 16 Fr Levin tube with a globe finger tied with silk in the distal end, inflated with 1 cm of air. They were placed on the bottom of the container where the water pressure was directly measured as a gold standard. The agreement between the two measures was assessed with the Bland-Altman method.

RESULTS:

We performed 120 simultaneous measures. The mean pressure difference was 0.218 (95% CI 0.074 to 0.363).

CONCLUSIONS:

The handcrafted prototype catheter and the direct measure were highly correlated. The new catheter is a reliable and reproducible tool for pressure monitoring. However, before it can be used in the clinical setting for IAP monitoring, validation in human models in a real clinical setting needs to be performed.
Assuntos

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Abdome / Catéteres / Hipertensão Intra-Abdominal / Manometria / Monitorização Fisiológica Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: World J Surg Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Colômbia

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Abdome / Catéteres / Hipertensão Intra-Abdominal / Manometria / Monitorização Fisiológica Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: World J Surg Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Colômbia