Your browser doesn't support javascript.
loading
Accuracy of a time-of-flight (ToF) imaging system for monitoring deep-inspiration breath-hold radiotherapy (DIBH-RT) for left breast cancer patients.
Abubakar, Auwal; Shaukat, Shazril Imran; Karim, Noor Khairiah A; Kassim, Mohammed Zakir; Lim, Siew Yong; Appalanaido, Gokula Kumar; Zin, Hafiz Mohd.
Afiliação
  • Abubakar A; Biomedical Imaging Department/Oncology and Radiotherapy Unit, Advanced Medical & Dental Institute, Universiti Sains Malaysia, Bertam, 13200, Kepala Batas, Pulau Pinang, Malaysia. a.abubakar@unimaid.edu.ng.
  • Shaukat SI; Department of Medical Radiography, Faculty of Allied Health Sciences, College of Medical Sciences, University of Maiduguri, Maiduguri, Nigeria. a.abubakar@unimaid.edu.ng.
  • Karim NKA; Biomedical Imaging Department/Oncology and Radiotherapy Unit, Advanced Medical & Dental Institute, Universiti Sains Malaysia, Bertam, 13200, Kepala Batas, Pulau Pinang, Malaysia.
  • Kassim MZ; Oncology Unit, Pantai Hospital Sungai Petani, Bandar Amanjaya, 08000, Sungai Petani, Kedah, Malaysia.
  • Lim SY; Biomedical Imaging Department/Oncology and Radiotherapy Unit, Advanced Medical & Dental Institute, Universiti Sains Malaysia, Bertam, 13200, Kepala Batas, Pulau Pinang, Malaysia.
  • Appalanaido GK; Breast Cancer Translational Research Programme (BCTRP), Advanced Medical & Dental Institute, Universiti Sains Malaysia, Bertam, 13200, Kepala Batas, Pulau Pinang, Malaysia.
  • Zin HM; Biomedical Imaging Department/Oncology and Radiotherapy Unit, Advanced Medical & Dental Institute, Universiti Sains Malaysia, Bertam, 13200, Kepala Batas, Pulau Pinang, Malaysia.
Phys Eng Sci Med ; 46(1): 339-352, 2023 Mar.
Article em En | MEDLINE | ID: mdl-36847965
ABSTRACT
Deep inspiration breath-hold radiotherapy (DIBH-RT) reduces cardiac dose by over 50%. However, poor breath-hold reproducibility could result in target miss which compromises the treatment success. This study aimed to benchmark the accuracy of a Time-of-Flight (ToF) imaging system for monitoring breath-hold during DIBH-RT. The accuracy of an Argos P330 3D ToF camera (Bluetechnix, Austria) was evaluated for patient setup verification and intra-fraction monitoring among 13 DIBH-RT left breast cancer patients. The ToF imaging was performed simultaneously with in-room cone beam computed tomography (CBCT) and electronic portal imaging device (EPID) imaging systems during patient setup and treatment delivery, respectively. Patient surface depths (PSD) during setup were extracted from the ToF and the CBCT images during free breathing and DIBH using MATLAB (MathWorks, Natick, MA) and the chest surface displacement were compared. The mean difference ± standard deviation, correlation coefficient, and limit of agreement between the CBCT and ToF were 2.88 ± 5.89 mm, 0.92, and - 7.36, 1.60 mm, respectively. The breath-hold stability and reproducibility were estimated using the central lung depth extracted from the EPID images during treatment and compared with the PSD from the ToF. The average correlation between ToF and EPID was - 0.84. The average intra-field reproducibility for all the fields was within 2.70 mm. The average intra-fraction reproducibility and stability were 3.74 mm, and 0.80 mm, respectively. The study demonstrated the feasibility of using ToF camera for monitoring breath-hold during DIBH-RT and shows good breath-hold reproducibility and stability during the treatment delivery.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Temas: Geral / Tipos_de_cancer / Outros_tipos / Tratamento / Radioterapia Base de dados: MEDLINE Assunto principal: Neoplasias da Mama Limite: Female / Humans Idioma: En Revista: Phys Eng Sci Med Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Malásia

Texto completo: 1 Coleções: 01-internacional Temas: Geral / Tipos_de_cancer / Outros_tipos / Tratamento / Radioterapia Base de dados: MEDLINE Assunto principal: Neoplasias da Mama Limite: Female / Humans Idioma: En Revista: Phys Eng Sci Med Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Malásia