Patient-specific stent for hemolytic anemia due to a kinked ascending aortic graft.
Lee, June; Lim, Hyun Ah; Hong, Seok Beom; Kim, Yong Han; Kim, Hwan Wook; Kim, Do Yeon.
Afiliação
  • Lee J; Department of Thoracic and Cardiovascular Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
  • Lim HA; Department of Thoracic and Cardiovascular Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
  • Hong SB; Department of Thoracic and Cardiovascular Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
  • Kim YH; Department of Thoracic and Cardiovascular Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
  • Kim HW; Department of Thoracic and Cardiovascular Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
  • Kim DY; Department of Thoracic and Cardiovascular Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
J Thorac Dis ; 15(8): 4497-4502, 2023 Aug 31.
Article em En | MEDLINE | ID: mdl-37691686
ABSTRACT
Hemolytic anemia after thoracic aortic surgery is rare. A 69-year-old female patient, who underwent ascending aorta replacement for acute type A aortic dissection 7 years ago, was transferred from another hospital due to hemolytic anemia necessitating regular blood transfusions. After engaging in interdisciplinary discussions, we ruled out alternative medical diagnoses and treatments, ultimately identifying the kinked aortic graft as the primary cause of hemolysis. Due to the patient's comorbidity, it was deemed high-risk to perform a redo thoracic aorta surgery for the replacement of the kinked aortic graft. Therefore, we decided to insert a patient-specific bare-metal stent into the kinked aortic graft to alleviate the kinking and promote improvement in hemolysis. We decided to perform the stent procedure under local anesthesia. The bare-metal stent was designed to be approximately 15% larger than the size of the pre-existing kinked aortic graft. A stent measuring 32 mm × 80 mm was successfully inserted and ballooned within the patient's ascending aortic graft. As a result, the patient's hemolytic condition showed gradual improvement over time. In this way, we aim to share our experience of inserting a patient-specific bare-metal stent under local anesthesia in patients who exhibit mechanical hemolysis due to a kinked aortic graft, especially when redo surgery is considered high-risk.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Idioma: En Revista: J Thorac Dis Ano de publicação: 2023 Tipo de documento: Article